Instruments for self-monitoring of blood glucose: comparisons of testing quality achieved by patients and a technician.

Svein Skeie1, Geir Thue, Kari Nerhus, Sverre Sandberg.   

Abstract

BACKGROUND: Instruments for self-monitoring of blood glucose (SMBG) are increasingly used by patients with diabetes. The analytical quality of meters in routine use is poorly characterized.
METHODS: We compared SMBG performance achieved by patients and by a medical laboratory technician. Imprecision was calculated from duplicate measurements, and deviation as the difference between the first measurement and the mean of duplicate laboratory-method results (calibrated with NIST material). Analytical quality for five groups of SMBG instruments was compared with quality specifications for BG measurements. All participants completed a questionnaire assessing both SMBG training and use of the meters.
RESULTS: We recruited 159 SMBG users from a hospital outpatient clinic and 263 others from 65 randomly selected general practices (total of 422). Most (two thirds) used insulin. CVs for the five meter types were 7%, 11%, 18%, 18%, and 20% in the hands of patients and 2.5-5.9% for the technician. For three of five meter types, patients' BG measurements had larger deviations from the laboratory results than did the technician's results. The technician's performance could not predict the patients'. No instrument when used by patients (but two operated by the technician) met published quality specifications. The analytical quality of patients' results was not related to whether they had chosen the instruments on advice from healthcare personnel (one-third of patients), were only self-educated in SMBG (50%), or performed SMBG fewer than seven times/week (62%).
CONCLUSIONS: The analytical quality of SMBG among patients was poorer than, and could not be predicted from, the performance of the meters in the hands of a technician. We suggest that new instruments be tested in the hands of patients who are trained on meter use in a routine way.

Entities:  

Mesh:

Year:  2002        PMID: 12089166

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  29 in total

1.  The new glucose standard POCT12-A3 misses the mark.

Authors:  Jan S Krouwer
Journal:  J Diabetes Sci Technol       Date:  2013-09-01

2.  Sources of glycemic variability--what type of technology is needed?

Authors:  Jonas Kildegaard; Toke Folke Christensen; Ole Kristian Hejlesen
Journal:  J Diabetes Sci Technol       Date:  2009-07-01

3.  Why the Details of Glucose Meter Evaluations Matters.

Authors:  Jan S Krouwer
Journal:  J Diabetes Sci Technol       Date:  2018-10-17

4.  Integrated self-monitoring of blood glucose system: handling step analysis.

Authors:  Guido Freckmann; Christina Schmid; Katharina Ruhland; Annette Baumstark; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2012-07-01

5.  Lot-to-lot variability of test strips and accuracy assessment of systems for self-monitoring of blood glucose according to ISO 15197.

Authors:  Annette Baumstark; Stefan Pleus; Christina Schmid; Manuela Link; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2012-09-01

6.  Clinical evaluation of a novel on-strip calibration method for blood glucose measurement.

Authors:  Michael Noble; John Rippeth; David Edington; Gerry Rayman; Sarah Brandon-Jones; Zoe Hollowood; Simon Kew
Journal:  J Diabetes Sci Technol       Date:  2014-05-12

7.  Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus.

Authors:  David B Sacks; Mark Arnold; George L Bakris; David E Bruns; Andrea Rita Horvath; M Sue Kirkman; Ake Lernmark; Boyd E Metzger; David M Nathan
Journal:  Diabetes Care       Date:  2011-06       Impact factor: 19.112

8.  Quality control of self-monitoring of blood glucose: why and how?

Authors:  Bogdan Solnica; Jerzy W Naskalski
Journal:  J Diabetes Sci Technol       Date:  2007-03

9.  Significant insulin dose errors may occur if blood glucose results are obtained from miscoded meters.

Authors:  Charles H Raine; Linda E Schrock; Steven V Edelman; Sunder Raj D Mudaliar; Weiping Zhong; Lois J Proud; Joan Lee Parkes
Journal:  J Diabetes Sci Technol       Date:  2007-03

10.  Glucose measurement: time for a gold standard.

Authors:  Joakim Hagvik
Journal:  J Diabetes Sci Technol       Date:  2007-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.