Standardized evaluation of nine instruments for self-monitoring of blood glucose.

Gunn B B Kristensen1, Grete Monsen, Svein Skeie, Sverre Sandberg.   

Abstract

BACKGROUND: Instruments for self-monitoring of blood glucose (SMBG) should undergo a standardized evaluation including a user-test before being marketed. In this study the results from standardized evaluations of nine different SMBG instruments are presented, and the standardized evaluation is discussed.
METHODS: Approximately 80 diabetes patients using three lots of test strips participated in each evaluation. Half of the patients were educated in how to use the meter, and the evaluations were carried out by both medical laboratory technologists (MLTs) and patients. Questionnaires were used to assess the user manual and the user-friendliness of the instrument.
RESULTS: The imprecision obtained by the patients (coefficients of variation [CVs] of 3.2-8.1%) were generally higher compared to that by the MLT (CVs of 2.3-5.9%). Three of the nine instruments did not achieve the quality goal based on the recommendation in the International Organization for Standardization's ISO 15197 guideline in the hands of diabetes patients. The bias from the comparison method ranged from -10.4% to +3.2%. There were significant lot-to-lot variations and hematocrit effects for some of the instruments. Temperature difference between the instruments and the test strip caused deterioration of the quality in one instrument. The user-friendliness was in general acceptable.
CONCLUSIONS: The quality of instruments for SMBG seems to have improved during recent years, although there are still analytical problems. A standardized evaluation protocol is necessary and should be regularly revised taking into account the development of new technology and the needs of the patients.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 19049376     DOI: 10.1089/dia.2008.0034

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  19 in total

1.  Usability Evaluation of a Blood Glucose Monitoring System With a Spill-Resistant Vial, Easier Strip Handling, and Connectivity to a Mobile App: Improvement of Patient Convenience and Satisfaction.

Authors:  Craig Harvey; Richard Koubek; Vanessa Bégat; Stephan Jacob
Journal:  J Diabetes Sci Technol       Date:  2016-08-22

2.  Accuracy in blood glucose measurement: what will a tightening of requirements yield?

Authors:  Lutz Heinemann; Volker Lodwig; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

3.  New Criteria for Assessing the Accuracy of Blood Glucose Monitors meeting, October 28, 2011.

Authors:  John Walsh; Ruth Roberts; Robert A Vigersky; Frank Schwartz
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

Review 4.  Analytical Performance Requirements for Systems for Self-Monitoring of Blood Glucose With Focus on System Accuracy: Relevant Differences Among ISO 15197:2003, ISO 15197:2013, and Current FDA Recommendations.

Authors:  Guido Freckmann; Christina Schmid; Annette Baumstark; Malte Rutschmann; Cornelia Haug; Lutz Heinemann
Journal:  J Diabetes Sci Technol       Date:  2015-04-14

5.  System Accuracy Evaluation of Four Systems for Self-Monitoring of Blood Glucose Following ISO 15197 Using a Glucose Oxidase and a Hexokinase-Based Comparison Method.

Authors:  Manuela Link; Christina Schmid; Stefan Pleus; Annette Baumstark; Delia Rittmeyer; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2015-04-14

6.  Lot-to-lot variability of test strips and accuracy assessment of systems for self-monitoring of blood glucose according to ISO 15197.

Authors:  Annette Baumstark; Stefan Pleus; Christina Schmid; Manuela Link; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2012-09-01

7.  System accuracy evaluation of 43 blood glucose monitoring systems for self-monitoring of blood glucose according to DIN EN ISO 15197.

Authors:  Guido Freckmann; Christina Schmid; Annette Baumstark; Stefan Pleus; Manuela Link; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2012-09-01

Review 8.  Assessing the quality of publications evaluating the accuracy of blood glucose monitoring systems.

Authors:  Gary H Thorpe
Journal:  Diabetes Technol Ther       Date:  2013-02-15       Impact factor: 6.118

9.  Considerations for an institution for evaluation of diabetes technology devices to improve their quality in the European Union.

Authors:  Lutz Heinemann; Guido Freckmann; Theodor Koschinsky
Journal:  J Diabetes Sci Technol       Date:  2013-03-01

10.  System Accuracy and User Performance Evaluation of an Improved System for Self-Monitoring of Blood Glucose.

Authors:  Annette Baumstark; Nina Jendrike; Christina Liebing; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2018-01-01