Effect of adverse storage conditions on performance of glucometer test strips.

Richard Bamberg1, Kathleen Schulman, Melissa MacKenzie, Joanna Moore, Sarah Olchesky.   

Abstract

OBJECTIVE: A study was conducted to assess the impact of adverse storage environments, i.e., not manufacturer recommended, on the performance of reagent test strips used with a point of care testing (POCT) glucometer to measure whole blood glucose levels. DESIGN/
SETTING: Glucose reagent test strips were placed in open, i.e., uncapped, and closed, i.e., capped vials. These vials were those used by the manufacturer to package and store the reagent test strips. One of each type of vial was placed in the manufacturer-recommended storage environment at room temperature and the adverse environments (incubator, direct light to mimic sunlight exposure, humidity, and refrigerated). The Accu-Chek Easy glucometer and reagent test strips as well as Accu-Chek Easy high and low glucose control solutions, manufactured by Roche, were used for this study. MAIN OUTCOME MEASURES: On day-3, day-7, and then once every 7 days, one strip from each vial in each environment was tested with the same glucometer using both a high and a low glucose control. The strip was considered failed for a type of vial and storage environment when either control was out of the reference range on a regular testing day and still out of range when tested the subsequent day. Testing continued up to 50 days.
RESULTS: For the tested environments it was found that, overall, test strip stability lasted longer for closed vials than open vials. For open vials in adverse storage conditions, the refrigerator environment offered the longest stability at 35 to 50 days and direct light and humidity offered the shortest periods of stability at 3 to 14 days.
CONCLUSIONS: The results of this study support the manufacturer's recommendations to store POCT glucose test strips in their original vial, capped, and at room temperature, though refrigeration may offer an alternative storage environment with acceptable stability. As compliance with testing, quality control, and storage instructions is often an issue with POCT, the manufacturers of these systems for blood glucose measurement should design storage systems that allow the patient to store the glucose meter and the reagent strips in the same location. Manufacturers may also need to consider designing storage systems that are more portable, knowing that patients must take the glucose meters and test strips with them when they travel. Roche's Accu-Chek Compact system is an example of such a design. The glucose test strips are incorporated into a drum that is stored in the Accu-Chek meter itself. When a patient performs a fingerstick blood glucose measurement, the drum advances to move a test strip outside the meter. When the test is complete, the test strip is ejected for disposal. Future studies to clarify the effect of adverse storage conditions, particularly refrigeration, on the integrity of POCT test systems and reagent strips is warranted with currently marketed brands.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16315736

Source DB:  PubMed          Journal:  Clin Lab Sci        ISSN: 0894-959X


  15 in total

1.  Diabetes in the desert: what do patients know about the heat?

Authors:  Adrienne A Nassar; Raymond D Childs; Mary E Boyle; Kimberly A Jameson; Margaret Fowke; Ken R Waters; Michael J Hovan; Curtiss B Cook
Journal:  J Diabetes Sci Technol       Date:  2010-09-01

Review 2.  Factors affecting blood glucose monitoring: sources of errors in measurement.

Authors:  Barry H Ginsberg
Journal:  J Diabetes Sci Technol       Date:  2009-07-01

3.  Integrated self-monitoring of blood glucose system: handling step analysis.

Authors:  Guido Freckmann; Christina Schmid; Katharina Ruhland; Annette Baumstark; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2012-07-01

4.  Self-monitoring of blood glucose: impact of a time delay between capillary blood sampling and glucose measurement.

Authors:  Guido Freckmann; Stefan Pleus; Annette Baumstark; Christina Schmid; Manuela Link; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2014-08-21

5.  System Accuracy and User Performance Evaluation of an Improved System for Self-Monitoring of Blood Glucose.

Authors:  Annette Baumstark; Nina Jendrike; Christina Liebing; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2018-01-01

Review 6.  Glucose meters: a review of technical challenges to obtaining accurate results.

Authors:  Ksenia Tonyushkina; James H Nichols
Journal:  J Diabetes Sci Technol       Date:  2009-07-01

7.  Consensus report: the current role of self-monitoring of blood glucose in non-insulin-treated type 2 diabetes.

Authors:  David C Klonoff; Lawrence Blonde; George Cembrowski; Antonio Roberto Chacra; Guillaume Charpentier; Stephen Colagiuri; George Dailey; Robert A Gabbay; Lutz Heinemann; David Kerr; Antonio Nicolucci; William Polonsky; Oliver Schnell; Robert Vigersky; Jean-François Yale
Journal:  J Diabetes Sci Technol       Date:  2011-11-01

8.  The food and drug administration is now preparing to establish tighter performance requirements for blood glucose monitors.

Authors:  David C Klonoff
Journal:  J Diabetes Sci Technol       Date:  2010-05-01

9.  Thermal stress and point-of-care testing performance: suitability of glucose test strips and blood gas cartridges for disaster response.

Authors:  Richard F Louie; Stephanie L Sumner; Shaunyé Belcher; Ron Mathew; Nam K Tran; Gerald J Kost
Journal:  Disaster Med Public Health Prep       Date:  2009-03       Impact factor: 1.385

10.  Evaluation of Hematocrit Influence on Self-Monitoring of Blood Glucose Based on ISO 15197:2013: Comparison of a Novel System With Five Systems With Different Hematocrit Ranges.

Authors:  Andrew Hattemer; Sami Wardat
Journal:  J Diabetes Sci Technol       Date:  2018-01-29
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.