Bagher Farhood1, Mohammad Taghi Bahreyni Toossi2, Shokouhozaman Soleymanifard2, Tohid Mortezazadeh1.
Abstract
AIM OF THE STUDY: Our objective was to quantify the accuracy of dose calculation in the build-up region of the tangential field of the breast for a TiGRT treatment planning system (TPS).Entities:
Keywords: TPS; TiGRT; build up region; dose calculation accuracy; radiotherapy; surface dose
Year: 2017 PMID: 29180932 PMCID: PMC5701585 DOI: 10.5114/wo.2017.70114
Source DB: PubMed Journal: Contemp Oncol (Pozn) ISSN: 1428-2526
Fig. 1Therapeutic range for the breast region on the RANDO phantom
Fig. 2Slice no. 18 of the RANDO phantom and arrangement of the thermoluminescent dosimeter chips at depths of 5 mm and 15 mm
Differences (%) between the doses calculated by the TiGRT treatment planning system and those measured by the TLD-100 for the open field with gantry angles of 15°, 30°, and 60° at depths of 5 mm and 15 mm
| Gantry angle | 15˚ | 30˚ | 60˚ | ||||
|---|---|---|---|---|---|---|---|
| The dose differences (%) between TiGRT TPS and TLD-100 | Superior slice | –17.38 | 0.84 | –11.85 | 9.1 | 10.2 | 12.17 |
| 4.32 | –10.01 | 6.18 | 4.91 | 6.44 | 14.08 | ||
| 8.97 | 3.09 | 5.35 | –2.19 | 25.22 | 5.86 | ||
| 2.02 | 15.65 | 8.2 | 12.45 | 42.16 | 10.85 | ||
| –14.08 | –10.4 | –1.13 | –8.99 | 25.68 | 13.78 | ||
| Middle slice | 6.86 | 10.52 | 6.67 | 6.62 | 10.95 | 7.84 | |
| –5.96 | 12.67 | –8.04 | 8.29 | 14.45 | 8.59 | ||
| 5.07 | 15.28 | –0.34 | 15.46 | 19.67 | 16.06 | ||
| 9.97 | –2.84 | 13.22 | –0.33 | 31.92 | 13.86 | ||
| 16.87 | 14.31 | 6.6 | 13.54 | 32.08 | 13.02 | ||
| Inferior slice | 15.15 | 13.9 | 25 | –7.66 | 6.92 | 9.69 | |
| –14.89 | 11.56 | –0.7 | –5.63 | 16.74 | 9.84 | ||
| –4.22 | 17.47 | 14.32 | 15.51 | 13.45 | 3.66 | ||
| 6.41 | 14.15 | 16.54 | 12.71 | 12.66 | 9.33 | ||
| 5.73 | 24.06 | –8.39 | 15.07 | 6.62 | 11.86 | ||
Dose differences (%) between the doses calculated by the TiGRT treatment planning system and the doses measured by the TLD-100 for the wedge field with gantry angles of 15°, 30°, and 60° at depths of 5 mm and 15 mm
| Gantry angle | 15˚ | 30˚ | 60˚ | ||||
|---|---|---|---|---|---|---|---|
| The dose differences (%) between TiGRT TPS and TLD-100 | Superior slice | 17.55 | 2.66 | 16.65 | 9.19 | 20.86 | 17.89 |
| 5.81 | –3.14 | 23.38 | 18.89 | 26.42 | 14.22 | ||
| 11.24 | 7.88 | –2.57 | –4.11 | 25.78 | 11.38 | ||
| 5.82 | 31.84 | 9.48 | 7.78 | 17.46 | 12.28 | ||
| 18.07 | 4.69 | 32.39 | –1.66 | –10.27 | 14.46 | ||
| Middle slice | 13.99 | 18.75 | 15.09 | 12.97 | 19.77 | 11.27 | |
| 15.96 | 15.66 | 10.01 | 13.23 | 17.42 | 9.66 | ||
| 5.28 | 21.63 | 7.68 | 6.75 | 15.66 | 16.76 | ||
| 7.21 | –3.25 | 11.84 | 7.87 | 25.44 | 18.28 | ||
| 20.8 | 7 | 9.35 | 13.11 | 24.93 | 18.85 | ||
| Inferior slice | 15.15 | 16.55 | 16.7 | 5.28 | 25.44 | 17.71 | |
| 10.09 | 23.75 | 8.54 | 21.22 | 15.81 | 19.31 | ||
| 24.35 | 13.53 | 24.98 | 19.73 | 31.11 | 13.48 | ||
| 10.98 | –2.32 | 10.08 | 13.56 | 28.73 | 18.04 | ||
| 4.66 | 20.74 | –2.3 | 18.57 | 19.67 | 16.13 | ||
Fig. 3Dose differences (%) between the calculated doses by TiGRT TPS and the measured doses by TLD-100 thermoluminescent dosimeter for the open field with gantry angles of 15°, 30°, and 60° at depths of 5 mm (A) and 15 mm (B), respectively
Fig. 4Dose differences (%) between the calculated doses by TiGRT TPS and the measured doses by TLD-100 thermoluminescent dosimeter for the wedge field with gantry angles of 15°, 30°, and 60° at depths of 5 mm (A) and 15 mm (B), respectively
Confidence limit values obtained for the open and wedge fields in gantry angles of 15°, 30°, and 60°
| Confidence limit (∆) | ||||
|---|---|---|---|---|
| Depth | 15˚ | 30˚ | 60˚ | |
| Open | 5 mm | 17.68 | 19.97 | 34.62 |
| 15 mm | 24.01 | 19.07 | 15.74 | |
| Wedge | 5 mm | 21.67 | 26.80 | 34.87 |
| 15 mm | 27.92 | 22.04 | 20.03 | |