Ali Sadeghi-Naini1,2,3,4, Harini Suraweera2,3, William Tyler Tran3,5, Farnoosh Hadizad2,3, Giancarlo Bruni6, Rashin Fallah Rastegar6, Belinda Curpen6, Gregory J Czarnota7,8,9,10.
Abstract
This study evaluated, for the first time, the efficacy of quantitative ultrasound (QUS) spectral parametric maps in conjunction with texture-analysis techniques to differentiate non-invasively benign versus malignant breast lesions. Ultrasound B-mode images and radiofrequency data were acquired from 78 patients with suspicious breast lesions. QUS spectral-analysis techniques were performed on radiofrequency data to generate parametric maps of mid-band fit, spectral slope, spectral intercept, spacing among scatterers, average scatterer diameter, and average acoustic concentration. Texture-analysis techniques were applied to determine imaging biomarkers consisting of mean, contrast, correlation, energy and homogeneity features of parametric maps. These biomarkers were utilized to classify benign versus malignant lesions with leave-one-patient-out cross-validation. Results were compared to histopathology findings from biopsy specimens and radiology reports on MR images to evaluate the accuracy of technique. Among the biomarkers investigated, one mean-value parameter and 14 textural features demonstrated statistically significant differences (p < 0.05) between the two lesion types. A hybrid biomarker developed using a stepwise feature selection method could classify the legions with a sensitivity of 96%, a specificity of 84%, and an AUC of 0.97. Findings from this study pave the way towards adapting novel QUS-based frameworks for breast cancer screening and rapid diagnosis in clinic.Entities:
Year: 2017 PMID: 29057899 PMCID: PMC5651882 DOI: 10.1038/s41598-017-13977-x
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Figure 1Representative MR and ultrasound B-mode images with QUS parametric overlays of MBF, SS, SI, SAS, ESD, EAC obtained from a benign and a malignant lesion. The color bar represents a scale encompassing 20 dBr for MBF, 5 dBr/MHz for SS, 30 dBr for SI, 3 mm for SAS, 160 µm for ESD, and 50 dB/cm3 for EAC. The scale bar represents 2 cm and 5 mm in MR and ultrasound images, respectively.
Figure 2Scatter plots of the QUS mean-value and textural parameters for benign versus malignant lesions. Statistically significant (p < 0.5), highly significant (p < 0.01), and extremely significant (p < 0.001) differences are shown with *, **, and ***, respectively.
The parameters with significant contribution to the hybrid QUS biomarker, identified through a stepwise linear discriminant analysis.
| Parameter | Standardized Coefficient |
|---|---|
| EAC Homogeneity | −2.30 |
| EAC Energy | 2.21 |
| MBF Homogeneity | 1.42 |
| SS Correlation | 1.26 |
| SAS Homogeneity | −0.79 |
| MBF | 0.77 |
| ESD Energy | −0.73 |
Figure 3Scatter plot of the hybrid QUS biomarker for benign versus malignant lesions, demonstrating a statistically extremely significant difference between the lesion types (p < 0.001).
Cross-validated results of breast-lesion classification using different QUS biomarkers.
| Parameter | Sensitivity (%) | Specificity (%) | Accuracy (%) | AUC |
|---|---|---|---|---|
| MBF | 78.3 | 68.7 | 74.4 | 0.86 |
| MBF Contrast | 84.8 | 59.4 | 74.4 | 0.87 |
| MBF Correlation | 82.6 | 71.9 | 78.2 | 0.91 |
| MBF Homogeneity | 78.2 | 75.0 | 76.9 | 0.92 |
| MBF Energy | 87.0 | 71.9 | 80.8 | 0.93 |
| SS | 78.3 | 62.5 | 71.8 | 0.86 |
| SS Contrast | 80.4 | 53.1 | 69.2 | 0.84 |
| SS Correlation | 78.3 | 68.7 | 74.4 | 0.86 |
| SS Homogeneity | 82.6 | 56.3 | 71.8 | 0.85 |
| SS Energy | 84.8 | 68.8 | 78.2 | 0.90 |
| SI | 87.0 | 65.6 | 78.2 | 0.85 |
| SI Contrast | 80.4 | 75.0 | 78.2 | 0.89 |
| SI Correlation | 82.6 | 75.0 | 79.5 | 0.94 |
| SI Homogeneity | 82.6 | 81.3 | 82.1 | 0.93 |
| SI Energy | 87.0 | 65.6 | 78.2 | 0.94 |
| SAS | 87.0 | 78.1 | 83.3 | 0.95 |
| SAS Contrast | 82.6 | 68.8 | 76.9 | 0.89 |
| SAS Correlation | 87.0 | 78.1 | 83.3 | 0.93 |
| SAS Homogeneity | 80.4 | 71.9 | 76.9 | 0.88 |
| SAS Energy | 87.0 | 59.4 | 75.6 | 0.88 |
| ESD | 89.1 | 59.4 | 76.9 | 0.84 |
| ESD Contrast | 87.0 | 68.8 | 79.5 | 0.92 |
| ESD Correlation | 78.3 | 75.0 | 76.9 | 0.88 |
| ESD Homogeneity | 87.0 | 71.9 | 80.8 | 0.90 |
| ESD Energy | 89.1 | 65.6 | 79.5 | 0.92 |
| EAC | 82.6 | 78.1 | 80.8 | 0.90 |
| EAC Contrast | 87.0 | 59.4 | 75.6 | 0.88 |
| EAC Correlation | 82.6 | 71.9 | 78.2 | 0.95 |
| EAC Homogeneity | 93.5 | 65.6 | 82.1 | 0.94 |
| EAC Energy | 84.8 | 78.1 | 82.1 | 0.94 |
| Hybrid Vector | 95.7 | 84.4 | 91.0 | 0.97 |