Evaluation of a pulsed xenon ultraviolet disinfection system to decrease bacterial contamination in operating rooms.

Lynn El Haddad1, Shashank S Ghantoji1, Mark Stibich1,2, Jason B Fleming3, Cindy Segal4, Kathy M Ware1, Roy F Chemaly5.   

Abstract

BACKGROUND: Environmental cleanliness is one of the contributing factors for surgical site infections in the operating rooms (ORs). To decrease environmental contamination, pulsed xenon ultraviolet (PX-UV), an easy and safe no-touch disinfection system, is employed in several hospital environments. The positive effect of this technology on environmental decontamination has been observed in patient rooms and ORs during the end-of-day cleaning but so far, no study explored its feasibility between surgical cases in the OR.
METHODS: In this study, 5 high-touch surfaces in 30 ORs were sampled after manual cleaning and after PX-UV intervention mimicking between-case cleaning to avoid the disruption of the ORs' normal flow. The efficacy of a 1-min, 2-min, and 8-min cycle were tested by measuring the surfaces' contaminants by quantitative cultures using Tryptic Soy Agar contact plates.
RESULTS: We showed that combining standard between-case manual cleaning of surfaces with a 2-min cycle of disinfection using a portable xenon pulsed ultraviolet light germicidal device eliminated at least 70% more bacterial load after manual cleaning.
CONCLUSIONS: This study showed the proof of efficacy of a 2-min cycle of PX-UV in ORs in eliminating bacterial contaminants. This method will allow a short time for room turnover and a potential reduction of pathogen transmission to patients and possibly surgical site infections.

Entities:  

Keywords:  Between-cases mimicking; Environment cleanliness; Operating rooms; Pulsed xenon ultraviolet

Mesh:

Substances:

Year:  2017        PMID: 29017457      PMCID: PMC5635568          DOI: 10.1186/s12879-017-2792-z

Source DB:  PubMed          Journal:  BMC Infect Dis        ISSN: 1471-2334            Impact factor:   3.090


Background

About 400,000 surgical site infections (SSIs) are documented annually in the United States, with associated costs of around $21,000 per case [1, 2]. Prevalent organisms associated with SSIs, such as Staphylococcus aureus, Enterococcus species, Klebsiella spp., Pseudomonas aeruginosa, and Escherichia coli, can persist on surfaces from 1.5 h to more than 30 months [3]. Standard manual cleaning alone is not sufficient to eliminate these pathogens; only around 47% of surfaces are appropriately disinfected during between-case and end-of-day terminal manual cleaning [4]. Implementation of efficient environmental disinfection methods as a supplement to manual cleaning may aid in reducing the risk of wound contamination and subsequent infection, thus eliminating the possible transmission of pathogens to patients [5, 6]. The portable ultraviolet light germicidal device employing pulsed xenon lamps (PX-UV) has been shown to be a safe, easy-to-operate, and effective system in decreasing the number of pathogens [7]. PX-UV uses a xenon flash lamp to generate broad-spectrum, high-intensity ultraviolet light to deactivate and kill bacteria, spores, and viruses on high-touch surfaces in 5 min or less [7]. Two studies have shown that the use of PX-UV in addition to standard end-of-day manual cleaning helped reduce bacterial contamination levels on surfaces in the operating rooms (ORs) by 62% and 81% [8, 9]. Furthermore, it was shown that contamination in the OR increases with sequential cases, leading to a more contaminated environment for each subsequent patient during operative hours [9]. Hence, rapid and effective between-case cleaning could reduce environmental contamination, protecting subsequent patients during the same day of operation. While improved patient outcomes have been observed after PX-UV during nightly terminal cleaning practices [8, 9], no data are available on the impact of this technology when applied between surgical cases. In this study, we aimed to determine the sufficient time required by the PX-UV device to reach environmental cleanliness.

Methods

This environmental sampling study was conducted at The University of Texas MD Anderson Cancer Center. The sampling occurred between the last end-of-case cleaning and the nightly standard terminal cleaning practices. Cleaning efficacy was assessed after 1, 2, and 8 min of PX-UV cycles using a PX-UV device (Xenex Disinfection Services). These cycle times were chosen based on proof-of-concept experiments conducted in the laboratory setting (data not shown). For each OR, high-touch surfaces were sampled at two distinct time points: after standard end-of-case cleaning and after PX-UV disinfection. At the conclusion of surgical cases each day, the room was cleaned by OR staff according to standard end-of-case protocols (manual cleaning with ready-to-use germicidal wipes or diluted solution). Following this cleaning by not more than 1.5 h, samples from 5 high-touch surfaces (computer monitor, electrocautery unit, anesthesia cart, chair, and bed table controls) were collected for quantitative culturing using Tryptic Soy Agar contact plates. For non-flat surfaces, the plates were rolled so that their entire area came in contact with the high-touch surface. The ORs were then disinfected with a PX-UV device for 1, 2, or 8 min (10 rooms for each cycle time) at the head of the table, ensuring direct line of sight for the UV light for high-touch surfaces (Fig. 1). Following PX-UV disinfection, the same 5 high-touch surfaces were sampled at sites adjacent to the first sites using Tryptic Soy Agar contact plates. After 48-h incubation at 37 °C of the plates, colony counts were recorded. We sampled 30 ORs, generating 150 samples before PX-UV use and 150 samples after PX-UV. Table 1 gives a detailed description of all cases in each sampled OR (Table 1).
Fig. 1

Schematic design of an operating room showing the accurate position of the PX-UV device (purple) to ensure direct line of sight of the UV light to the high-touch surfaces (identified as a to h)

Table 1

Description of the case flow and case types of the 30 sampled operating rooms

ORPX-UV cycle time (minutes)Mean c.f.u. before PX-UVMean c.f.u. after PX-UVTotal number of cases during the dayConsecutive case typeLast case type of the day before PX-UV
110.60.83Laparotomy; Incision and Drainage; MastectomyMastectomy
214.21.4 ↓2Wide Local Excision; Neck DissectionNeck Dissection
311.21.63Laparoscopic Nephrectomy; Laparoscopic AdrenalectomyPartial Nephrectomy
412.31.4 ↓3Mastectomy, Wide Local ExcisionExcision of lesion on back
513.21.4 ↓1Laminectomy with stabilizationLaminectomy with stabilization
612.80.4 ↓2NephrectomyDiagnostic Laparostomy
716.22.2 ↓2Thyroid with Neck DissectionClosure of enterostomy
812.22.62ParotidectomyExamination under Anaesthesia with biopsies
914.23.6 ↓1Removal of tibial nail hardware and complex closureRemoval of tibial nail hardware & complex closure
1011.01.63Insertion of Port-A-CathInsertion of Port-A-Cath
1120.00.0 ↓1Partial Excision of genitalia with reconstructionPartial Excision of genitalia with reconstruction
1221.01.22closure of enterostomyPartial colectomy
1321.82.21Laparoscopic GastrectomyLaparoscopic Gastrectomy
1421.40.6 ↓2Tandem and Ovoid insertionTandem and Ovoid insertion
1521.00.6 ↓1Partial Lobectomy Liver and HysterectomyPartial Lobectomy Liver and Hysterectomy
1621.40.8 ↓2Diagnostic Lap with Central Venous Catheter insertionDiagnostic Laparotomy with CVC insertion
1722.20.4 ↓2Experimental Laparoscopy with bowel anastomosisExcision of groin lymph node
1822.61.6 ↓1Proctectomy, Hysterectomy and reconstructionProctectomy, Hysterectomy and reconstruction
19211.80.8 ↓2Lap HysterectomyLap Hysterectomy
2021.40.4 ↓3Insertion of central lineThoracoscopy with segmental lung resection
2181.20.6 ↓1CraniotomyCraniotomy
2283.21.0 ↓2Laparoscopic liver resectionClosure of enterostomy and Laparoscopic liver
2383.20.8 ↓2Port-A-Cath insertionDiagnostic Laparotomy with liver biopsies
2484.00.6 ↓2Mastectomy with reconstructionSegmental Mastectomy
25810.81.6 ↓1Mastectomy with reconstructionMastectomy with reconstruction
2681.40.0 ↓3Diagnostic Laparostomy with Hysterectomy; Incision and Drainage cystCystoscopy with biopsies
2780.80.2 ↓2Laparoscopic Salpingo-OophorectomyDiagnostic Laparotomy with biopsies
2882.20.6 ↓2ThyroidectomyMastectomy
2987.82.6 ↓1Partial colectomy with nephrectomyPartial colectomy with nephrectomy
3087.60.5 ↓1Thoracotomy with Lobectomy and Pulmonary Arterial reconstructionThoracotomy with Lobectomy and Pulmonary Arterial reconstruction

Abbreviations: OR, Operating Room; PX-UV, pulsed xenon ultraviolet; c.f.u., Colony-forming units. The symbol “↓” indicates a decrease in c.f.u. after PX-UV

Schematic design of an operating room showing the accurate position of the PX-UV device (purple) to ensure direct line of sight of the UV light to the high-touch surfaces (identified as a to h) Description of the case flow and case types of the 30 sampled operating rooms Abbreviations: OR, Operating Room; PX-UV, pulsed xenon ultraviolet; c.f.u., Colony-forming units. The symbol “↓” indicates a decrease in c.f.u. after PX-UV The pre-PX-UV samples were combined for analysis to remove any variance issue. Means, medians, and ranges of colony counts were recorded at each sampling period for statistical analysis. As the data were nonparametric, a Wilcoxon rank sum test was used to examine the differences between groups.

Results

A total of 147 pre-PX-UV samples and 148 post-PX-UV samples measuring bacterial load obtained for the 5 high-touch surfaces were included in the analysis. Five plates (3 in the pre-PX-UV group and 2 in post-PX-UV groups) were discarded from the analysis as outliers because of counts that were too numerous to count (TNTC) and attributed to lab error, such as a dislodged cover plate. If included in the analysis, the outliers would have had undue leverage on the data for the intervention group that had no outliers (the 1-min group), and therefore the removal of the outliers was deemed conservative. Table 2 depicts the changes in the colony-forming units (c.f.u.) between pre- and post-PX-UV use at different cycle times. A 1-min cycle of PX-UV did not generate a significant reduction in the level of contamination on the high-touch surfaces (P = 0.594). However, 2- and 8-min cycles showed significant reduction in the level of environmental contamination by decreasing the mean colony counts by 72.5% (P = 0.0328) and 73.1% (P = 0.0075), respectively (Table 2). A 2-min PX-UV cycle was as effective in eliminating an equal load of bacterial contamination when compared to an 8-min cycle.
Table 2

Efficacy of 1-, 2-, and 8-min PX-UV disinfection cycle times in reducing operating room contamination

Timing of samplingSamples taken (n)Colony count (c.f.u.)Reductiona (%) P-value
MeanMinMaxIQR
Pre PX-UV (all cycles combined)1473.190553
Post 1-min PX-UV501.70014246.70.5940
Post 2-min PX-UV490.8809172.50.0328
Post 8-min PX-UV490.8607173.10.0075

aReduction of mean colony count after PX-UV in comparison with pre-PX-UV mean colony count. PX-UV, pulsed xenon ultraviolet; IQR, Interquartile range; c.f.u., Colony-forming units; min, minimum; max, maximum

Efficacy of 1-, 2-, and 8-min PX-UV disinfection cycle times in reducing operating room contamination aReduction of mean colony count after PX-UV in comparison with pre-PX-UV mean colony count. PX-UV, pulsed xenon ultraviolet; IQR, Interquartile range; c.f.u., Colony-forming units; min, minimum; max, maximum

Discussion

We found PX-UV disinfection effective in reducing colony counts when performed after standard cleaning. The 2- and 8-min PX-UV cycles produced equivalent and significant reduction of level of contamination when compared to standard OR cleaning alone and were more effective than the 1-min PX-UV cycle. We conclude that a 2-min cycle optimizes efficacy and efficiency. A recent meta-analysis of financial impact on the United States healthcare system showed that SSIs contribute to 33.7% of the overall annual cost ($9.8 billion) of healthcare-associated infections [2]. By implementing this SSI prevention approach in the OR setting, contamination in the OR could be controlled during sequential cases, leading to a decontaminated environment for subsequent patients and may have positive impact on the rate of SSIs and associated costs. PX-UV has been successfully used to reduce or eliminate pathogens such as vancomycin-resistant enterococci, Methicillin-resistant Staphylococcus aureus (MRSA), as well as Clostridium difficile on high-touch surfaces in patient rooms [10, 11]. In fact, PX-UV combined with quaternary ammonium removed 95% of C. difficile spores compared to a 70% of spores reduction when disinfecting patient rooms with bleach [10]. Moreover, PX-UV was 16 times more effective than manual cleaning in eliminating MRSA [12] and 100% effective against VRE [11]. The efficacy of this method has also been confirmed against fungi, Bacillus anthracis, and viruses such as Ebola virus [13]. In addition, PX-UV does not damage materials in hospital settings and is not transmitted through glass windows [10]. Another method for decontaminating OR rooms between cases is the use of improved hydrogen peroxide products (IHP) such as Activated Hydrogen Peroxide (Clorox Healthcare). Even though this disinfectant is effective in reducing the contamination level to around 84% of the baseline, it presents a major limitation, i.e., manual cleaning for about 2 to 4.5 h [14]. Manual cleaning is not predictable nor optimal being dependent upon the education of the cleaning personnel and the nurses [15, 16]. In fact, when cleaning, tools such as buckets, mop heads, and wipes can rapidly become contaminated and potentially transfer pathogens to other cleaned surfaces [7]. Also, the continual and recurrent use of the same chemical disinfectant can lead to the emergence of resistant microorganisms [7]. Moreover, the time spent on manually cleaning constitutes an important drawback in ORs where rapid bed turnaround time is crucial and entails operational costs for training specialized personnel. Finally, IHP costs around $175 per room, whereas the PX-UV device costs approximately $3 per room to operate, excluding labor costs in both cases [10]. The present study was limited to 5 high-touch surfaces. Other high-touch surfaces such as floors, light switches, cabinet handles, and doorknobs could be added to future studies. Additional limitations are the somewhat small sample size used in this study and the lack of bacterial identification to the species level by our use of TSA sampling plates, which are limited in detection to aerobic bacteria only. Moreover, the impact of PX-UV use between cases on SSIs and identification of bacteria at the species level on ORs surfaces still need to be determined in future studies. Finally, an operational study that investigate the impact of the between-case use of PX-UV on OR case flow would be necessary.

Conclusions

In summary, our results suggest that supplementing standard cleaning procedures using a portable no-touch PX-UV system could be done routinely and rapidly between cases in the OR. A cycle of 2 min was sufficient in eliminating 70% or more of the bacterial load on inanimate high-touch surfaces, thus allowing short time for room turnover and potentially reducing pathogen transmission to patients and possibly SSI rates.
  12 in total

1.  Analysis of risk factors for sternal surgical site infection: emphasizing the appropriate ventilation of the operating theaters.

Authors:  Serap Simsek Yavuz; Yesim Bicer; Nihan Yapici; Sibel Kalaca; Osman Ozcan Aydin; Gercek Camur; Funda Kocak; Zuhal Aykac
Journal:  Infect Control Hosp Epidemiol       Date:  2006-08-14       Impact factor: 3.254

Review 2.  Health care-associated infections: a meta-analysis of costs and financial impact on the US health care system.

Authors:  Eyal Zimlichman; Daniel Henderson; Orly Tamir; Calvin Franz; Peter Song; Cyrus K Yamin; Carol Keohane; Charles R Denham; David W Bates
Journal:  JAMA Intern Med       Date:  2013 Dec 9-23       Impact factor: 21.873

Review 3.  Role of ultraviolet (UV) disinfection in infection control and environmental cleaning.

Authors:  Zubair Qureshi; Mohamed H Yassin
Journal:  Infect Disord Drug Targets       Date:  2013-06

4.  Cleanliness of portable medical equipment disinfected by nursing staff.

Authors:  Nancy L Havill; Heather L Havill; Elise Mangione; Diane G Dumigan; John M Boyce
Journal:  Am J Infect Control       Date:  2011-04-15       Impact factor: 2.918

5.  Evaluation of a pulsed-xenon ultraviolet room disinfection device for impact on hospital operations and microbial reduction.

Authors:  Mark Stibich; Julie Stachowiak; Benjamin Tanner; Matthew Berkheiser; Linette Moore; Issam Raad; Roy F Chemaly
Journal:  Infect Control Hosp Epidemiol       Date:  2011-03       Impact factor: 3.254

Review 6.  The role of the healthcare environment in the spread of multidrug-resistant organisms: update on current best practices for containment.

Authors:  Roy F Chemaly; Sarah Simmons; Charles Dale; Shashank S Ghantoji; Maria Rodriguez; Julie Gubb; Julie Stachowiak; Mark Stibich
Journal:  Ther Adv Infect Dis       Date:  2014-06

7.  Evaluation of the effectiveness of improved hydrogen peroxide in the operating room.

Authors:  Timothy L Wiemken; Daniel R Curran; Robert R Kelley; Emily B Pacholski; Ruth M Carrico; Paula Peyrani; M S S Khan; Julio A Ramirez
Journal:  Am J Infect Control       Date:  2014-06-14       Impact factor: 2.918

8.  Financial impact of surgical site infections on hospitals: the hospital management perspective.

Authors:  John Shepard; William Ward; Aaron Milstone; Taylor Carlson; John Frederick; Eric Hadhazy; Trish Perl
Journal:  JAMA Surg       Date:  2013-10       Impact factor: 14.766

Review 9.  How long do nosocomial pathogens persist on inanimate surfaces? A systematic review.

Authors:  Axel Kramer; Ingeborg Schwebke; Günter Kampf
Journal:  BMC Infect Dis       Date:  2006-08-16       Impact factor: 3.090

10.  Evaluation of a pulsed-xenon ultraviolet room disinfection device for impact on contamination levels of methicillin-resistant Staphylococcus aureus.

Authors:  Chetan Jinadatha; Ricardo Quezada; Thomas W Huber; Jason B Williams; John E Zeber; Laurel A Copeland
Journal:  BMC Infect Dis       Date:  2014-04-07       Impact factor: 3.090

View more
  12 in total

1.  Emergence and Transmission of Daptomycin and Vancomycin-Resistant Enterococci Between Patients and Hospital Rooms.

Authors:  Lynn El Haddad; Blake M Hanson; Cesar A Arias; Shashank S Ghantoji; Cynthia P Harb; Mark Stibich; Roy F Chemaly
Journal:  Clin Infect Dis       Date:  2021-12-16       Impact factor: 9.079

Review 2.  The efficacy of ultraviolet light-emitting technology against coronaviruses: a systematic review.

Authors:  F Chiappa; B Frascella; G P Vigezzi; M Moro; L Diamanti; L Gentile; P Lago; N Clementi; C Signorelli; N Mancini; A Odone
Journal:  J Hosp Infect       Date:  2021-05-21       Impact factor: 3.926

3.  Environmental disinfection with photocatalyst as an adjunctive measure to control transmission of methicillin-resistant Staphylococcus aureus: a prospective cohort study in a high-incidence setting.

Authors:  Min Hyung Kim; Seong Gyu Lee; Ki Sook Kim; Yoon Ji Heo; Ji Eun Oh; Su Jin Jeong
Journal:  BMC Infect Dis       Date:  2018-12-03       Impact factor: 3.090

Review 4.  Update on infection control practices in cancer hospitals.

Authors:  Ella J Ariza-Heredia; Roy F Chemaly
Journal:  CA Cancer J Clin       Date:  2018-07-09       Impact factor: 508.702

5.  Pulsed xenon ultraviolet and non-thermal atmospheric plasma treatments are effective for the disinfection of air in hospital blood sampling rooms.

Authors:  Shan-Ni Wang; Jing-Jing Li; Ying-Xin Liu; Zhi Lin; Jiao-Jiao Qiao; Li-Hua Chen; Yu Li; Yong Wu; Mei-Mei Wang; Yun-Bo Liu; Chen Yan; Zhi-Heng Chen; Chang-Qing Gao
Journal:  Photodiagnosis Photodyn Ther       Date:  2019-05-29       Impact factor: 3.631

6.  Evaluation of an Ultraviolet C (UVC) Light-Emitting Device for Disinfection of High Touch Surfaces in Hospital Critical Areas.

Authors:  Beatrice Casini; Benedetta Tuvo; Maria Luisa Cristina; Anna Maria Spagnolo; Michele Totaro; Angelo Baggiani; Gaetano Pierpaolo Privitera
Journal:  Int J Environ Res Public Health       Date:  2019-09-24       Impact factor: 3.390

7.  Ultraviolet disinfection robots to improve hospital cleaning: Real promise or just a gimmick?

Authors:  Magda Diab-El Schahawi; Walter Zingg; Margreet Vos; Hilary Humphreys; Lorena Lopez-Cerero; Astrid Fueszl; Jean Ralph Zahar; Elisabeth Presterl
Journal:  Antimicrob Resist Infect Control       Date:  2021-02-12       Impact factor: 4.887

Review 8.  Potential of pulsed light technology for control of SARS-CoV-2 in hospital environments.

Authors:  Julie Jean; María Isabel Rodríguez-López; Eric Jubinville; Estrella Núñez-Delicado; Vicente M Gómez-López
Journal:  J Photochem Photobiol B       Date:  2020-12-28       Impact factor: 6.252

9.  Development of a Pulsed Xenon Ultraviolet Disinfection Device for Real-Time Air Disinfection in Ambulances.

Authors:  Li Song; Wei Li; Jian'an He; Lang Li; Tao Li; Dayong Gu; Huanwen Tang
Journal:  J Healthc Eng       Date:  2020-02-24       Impact factor: 2.682

Review 10.  Use of ultraviolet-C in environmental sterilization in hospitals: A systematic review on efficacy and safety.

Authors:  Clysly Celine R Ramos; Josemaria Lorenzo A Roque; Diane B Sarmiento; Luis Enrico G Suarez; Janela Tanya P Sunio; Kaezzy Ila B Tabungar; Geraldine Susan C Tengco; Phylis C Rio; Allan L Hilario
Journal:  Int J Health Sci (Qassim)       Date:  2020 Nov-Dec
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.