Objectives: We have developed a pulsed xenon ultraviolet light-based real-time air disinfection system with rapid and effective disinfection by using high-intensity pulse germicidal UV. Disinfection of the ambulance's environment is critical in the prevention of infectious cross contamination.Entities:
Mesh:
Substances:
Year: 2020 PMID: 32184979 PMCID: PMC7060415 DOI: 10.1155/2020/6053065
Source DB: PubMed Journal: J Healthc Eng ISSN: 2040-2295 Impact factor: 2.682
Figure 1Experimental configuration of the pulsed xenon ultraviolet disinfection device (the size of the device is 505 × 500 × 120 mm).
Effectiveness of the real-time air disinfection device against E. coli in the air.
| Time (min) | No. |
|
|
|
|
|
|
|---|---|---|---|---|---|---|---|
| 30 | 1 | 2.70 × 105 | 1.76 × 105 | 34.81 | 2.64 × 105 | 0 | 100 |
| 2 | 2.54 × 105 | 1.66 × 105 | 34.65 | 2.40 × 105 | 0 | 100 | |
| 3 | 2.10 × 105 | 1.40 × 105 | 33.33 | 2.10 × 104 | 0 | 100 |
N : the natural extinction rate of bacteria in the air; V0 and V: the amount of bacteria in the air at different times before and during the experiment was conducted; V′ and V′: the amount of bacteria in the air at different times before and during the disinfection process of the experimental group; K: the disinfection rate of bacteria in air.
Effectiveness of the real-time air disinfection device against Staphylococcus albus in the air.
| Time (min) | No. |
|
|
|
|
|
|
|---|---|---|---|---|---|---|---|
| 30 | 1 | 2.64 × 105 | 1.76 × 105 | 33.33 | 2.64 × 105 | 160 | 99.91 |
| 2 | 2.44 × 105 | 1.51 × 105 | 38.11 | 2.52 × 105 | 150 | 99.90 | |
| 3 | 2.31 × 105 | 1.40 × 105 | 39.39 | 2.40 × 105 | 130 | 99.91 |
N : the natural extinction rate of bacteria in the air; V0 and V: the amount of bacteria in the air at different times before and during the experiment was conducted; V′ and V′: the amount of bacteria in the air at different times before and during the disinfection process of the experimental group; K: the disinfection rate of bacteria in air.
Disinfection efficiency of the real-time PX-UV for natural bacteria in ambulances.
| Time (min) | No. | Before (Cfu/m3) | After (Cfu/m3) | Killing rate (%) |
|---|---|---|---|---|
| 60 | 1 | 480 | 30 | 93.75 |
| 2 | 490 | 40 | 91.84 | |
| 3 | 400 | 40 | 90.00 |