Noise power spectra of images from digital mammography detectors.

M B Williams1, P A Mangiafico, P U Simoni.   

Abstract

Noise characterization through estimation of the noise power spectrum (NPS) is a central component of the evaluation of digital x-ray systems. We begin with a brief review of the fundamentals of NPS theory and measurement, derive explicit expressions for calculation of the one- and two-dimensional (1D and 2D) NPS, and discuss some of the considerations and tradeoffs when these concepts are applied to digital systems. Measurements of the NPS of two detectors for digital mammography are presented to illustrate some of the implications of the choices available. For both systems, two-dimensional noise power spectra obtained over a range of input fluence exhibit pronounced asymmetry between the orthogonal frequency dimensions. The 2D spectra of both systems also demonstrate dominant structures both on and off the primary frequency axes indicative of periodic noise components. Although the two systems share many common noise characteristics, there are significant differences, including markedly different dark-noise magnitudes, differences in NPS shape as a function of both spatial frequency and exposure, and differences in the natures of the residual fixed pattern noise following flat fielding corrections. For low x-ray exposures, quantum noise-limited operation may be possible only at low spatial frequency. Depending on the method of obtaining the 1D NPS (i.e., synthetic slit scanning or slice extraction from the 2D NPS), on-axis periodic structures can be misleadingly smoothed or missed entirely. Our measurements indicate that for these systems, 1D spectra useful for the purpose of detective quantum efficiency calculation may be obtained from thin cuts through the central portion of the calculated 2D NPS. On the other hand, low-frequency spectral values do not converge to an asymptotic value with increasing slit length when 1D spectra are generated using the scanned synthetic slit method. Aliasing can contribute significantly to the digital NPS, especially near the Nyquist frequency. Calculation of the theoretical presampling NPS and explicit inclusion of aliased noise power shows good agreement with measured values.

Mesh:

Year:  1999        PMID: 10435530     DOI: 10.1118/1.598623

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  15 in total

1.  Mammographic imaging with a small format CCD-based digital cassette: physical characteristics of a clinical system.

Authors:  S Vedantham; A Karellas; S Suryanarayanan; I Levis; M Sayag; R Kleehammer; R Heidsieck; C J D'Orsi
Journal:  Med Phys       Date:  2000-08       Impact factor: 4.071

2.  Integrated model for estimating phosphor signal and noise transfer characteristics on medical images: application to CdPO3Cl:Mn phosphor screens.

Authors:  D Cavouras; I Kandarakis; G S Panayiotakis; C D Nomicos
Journal:  Med Biol Eng Comput       Date:  2002-05       Impact factor: 2.602

3.  Solid-state fluoroscopic imager for high-resolution angiography: physical characteristics of an 8 cm x 8 cm experimental prototype.

Authors:  Srinivasan Vedantham; Andrew Karellas; Sankararaman Suryanarayanan; Steven K Onishi
Journal:  Med Phys       Date:  2004-06       Impact factor: 4.071

4.  A comparative study of limited-angle cone-beam reconstruction methods for breast tomosynthesis.

Authors:  Yiheng Zhang; Heang-Ping Chan; Berkman Sahiner; Jun Wei; Mitchell M Goodsitt; Lubomir M Hadjiiski; Jun Ge; Chuan Zhou
Journal:  Med Phys       Date:  2006-10       Impact factor: 4.071

5.  Accurate MTF measurement in digital radiography using noise response.

Authors:  Andrew Kuhls-Gilcrist; Amit Jain; Daniel R Bednarek; Kenneth R Hoffmann; Stephen Rudin
Journal:  Med Phys       Date:  2010-02       Impact factor: 4.071

6.  Investigation of noise sources for digital radiography systems.

Authors:  Lutfi Ergun; Turan Olgar
Journal:  Radiol Phys Technol       Date:  2016-10-01

7.  On the performance of the noise power spectrum from the gain-corrected radiography images.

Authors:  Dong Sik Kim; Eunae Lee
Journal:  J Med Imaging (Bellingham)       Date:  2018-03-30

8.  Design and evaluation of a grid reciprocation scheme for use in digital breast tomosynthesis.

Authors:  Tushita Patel; Helen Sporkin; Heather Peppard; Mark B Williams
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2016-03-29

9.  Novel method to determine recursive filtration and noise reduction in fluoroscopic imaging - a comparison of four different vendors.

Authors:  Bente Konst; Jacob Nøtthellen; Stine Nalum Naess; Magnus Båth
Journal:  J Appl Clin Med Phys       Date:  2020-12-14       Impact factor: 2.102

10.  The effects of x-ray beam hardening on detective quantum efficiency and radiation dose.

Authors:  Molly Donovan Wong; Xizeng Wu; Hong Liu
Journal:  J Xray Sci Technol       Date:  2011       Impact factor: 1.535

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.