B0 concomitant field compensation for MRI systems employing asymmetric transverse gradient coils.

Paul T Weavers1, Shengzhen Tao1,2, Joshua D Trzasko1, Louis M Frigo3, Yunhong Shu1, Matthew A Frick1, Seung-Kyun Lee4,5, Thomas K-F Foo4, Matt A Bernstein1.   

Abstract

PURPOSE: Imaging gradients result in the generation of concomitant fields, or Maxwell fields, which are of increasing importance at higher gradient amplitudes. These time-varying fields cause additional phase accumulation, which must be compensated for to avoid image artifacts. In the case of gradient systems employing symmetric design, the concomitant fields are well described with second-order spatial variation. Gradient systems employing asymmetric design additionally generate concomitant fields with global (zeroth-order or B0 ) and linear (first-order) spatial dependence.
METHODS: This work demonstrates a general solution to eliminate the zeroth-order concomitant field by applying the correct B0 frequency shift in real time to counteract the concomitant fields. Results are demonstrated for phase contrast, spiral, echo-planar imaging (EPI), and fast spin-echo imaging.
RESULTS: A global phase offset is reduced in the phase-contrast exam, and blurring is virtually eliminated in spiral images. The bulk image shift in the phase-encode direction is compensated for in EPI, whereas signal loss, ghosting, and blurring are corrected in the fast-spin echo images.
CONCLUSION: A user-transparent method to compensate the zeroth-order concomitant field term by center frequency shifting is proposed and implemented. This solution allows all the existing pulse sequences-both product and research-to be retained without any modifications. Magn Reson Med 79:1538-1544, 2018.
© 2017 International Society for Magnetic Resonance in Medicine. © 2017 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  B0; asymmetric gradients; frequency shifting; real-time correction

Mesh:

Year:  2017        PMID: 28639370      PMCID: PMC5740018          DOI: 10.1002/mrm.26790

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  22 in total

1.  Concomitant gradient field effects in spiral scans.

Authors:  K F King; A Ganin; X J Zhou; M A Bernstein
Journal:  Magn Reson Med       Date:  1999-01       Impact factor: 4.668

2.  NonCartesian MR image reconstruction with integrated gradient nonlinearity correction.

Authors:  Shengzhen Tao; Joshua D Trzasko; Yunhong Shu; John Huston; Kevin M Johnson; Paul T Weavers; Erin M Gray; Matt A Bernstein
Journal:  Med Phys       Date:  2015-12       Impact factor: 4.071

3.  Concomitant magnetic-field-induced artifacts in axial echo planar imaging.

Authors:  X J Zhou; Y P Du; M A Bernstein; H G Reynolds; J K Maier; J A Polzin
Journal:  Magn Reson Med       Date:  1998-04       Impact factor: 4.668

4.  ACR guidance document on MR safe practices: 2013.

Authors:  Emanuel Kanal; A James Barkovich; Charlotte Bell; James P Borgstede; William G Bradley; Jerry W Froelich; J Rod Gimbel; John W Gosbee; Ellisa Kuhni-Kaminski; Paul A Larson; James W Lester; John Nyenhuis; Daniel Joe Schaefer; Elizabeth A Sebek; Jeffrey Weinreb; Bruce L Wilkoff; Terry O Woods; Leonard Lucey; Dina Hernandez
Journal:  J Magn Reson Imaging       Date:  2013-01-23       Impact factor: 4.813

5.  High slew-rate head-only gradient for improving distortion in echo planar imaging: Preliminary experience.

Authors:  Ek T Tan; Seung-Kyun Lee; Paul T Weavers; Dominic Graziani; Joseph E Piel; Yunhong Shu; John Huston; Matt A Bernstein; Thomas K F Foo
Journal:  J Magn Reson Imaging       Date:  2016-02-26       Impact factor: 4.813

6.  Peripheral nerve stimulation characteristics of an asymmetric head-only gradient coil compatible with a high-channel-count receiver array.

Authors:  Seung-Kyun Lee; Jean-Baptiste Mathieu; Dominic Graziani; Joseph Piel; Eric Budesheim; Eric Fiveland; Christopher J Hardy; Ek Tsoon Tan; Bruce Amm; Thomas K-F Foo; Matt A Bernstein; John Huston; Yunhong Shu; John F Schenck
Journal:  Magn Reson Med       Date:  2015-12-02       Impact factor: 4.668

7.  Integrated image reconstruction and gradient nonlinearity correction.

Authors:  Shengzhen Tao; Joshua D Trzasko; Yunhong Shu; John Huston; Matt A Bernstein
Journal:  Magn Reson Med       Date:  2014-10-08       Impact factor: 4.668

8.  The effect of concomitant fields in fast spin echo acquisition on asymmetric MRI gradient systems.

Authors:  Shengzhen Tao; Paul T Weavers; Joshua D Trzasko; John Huston; Yunhong Shu; Erin M Gray; Thomas K F Foo; Matt A Bernstein
Journal:  Magn Reson Med       Date:  2017-06-22       Impact factor: 4.668

9.  Regularized field map estimation in MRI.

Authors:  Amanda K Funai; Jeffrey A Fessler; Desmond T B Yeo; Valur T Olafsson; Douglas C Noll
Journal:  IEEE Trans Med Imaging       Date:  2008-10       Impact factor: 10.048

10.  Torque free asymmetric gradient coils for echo planar imaging.

Authors:  A M Abduljalil; A H Aletras; P M Robitaille
Journal:  Magn Reson Med       Date:  1994-04       Impact factor: 4.668

View more
  11 in total

1.  Distortion-free imaging: A double encoding method (DIADEM) combined with multiband imaging for rapid distortion-free high-resolution diffusion imaging on a compact 3T with high-performance gradients.

Authors:  Myung-Ho In; Ek Tsoon Tan; Joshua D Trzasko; Yunhong Shu; Daehun Kang; Uten Yarach; Shengzhen Tao; Erin M Gray; John Huston; Matt A Bernstein
Journal:  J Magn Reson Imaging       Date:  2019-05-20       Impact factor: 4.813

2.  Lightweight, compact, and high-performance 3T MR system for imaging the brain and extremities.

Authors:  Thomas K F Foo; Evangelos Laskaris; Mark Vermilyea; Minfeng Xu; Paul Thompson; Gene Conte; Christopher Van Epps; Christopher Immer; Seung-Kyun Lee; Ek T Tan; Dominic Graziani; Jean-Baptise Mathieu; Christopher J Hardy; John F Schenck; Eric Fiveland; Wolfgang Stautner; Justin Ricci; Joseph Piel; Keith Park; Yihe Hua; Ye Bai; Alex Kagan; David Stanley; Paul T Weavers; Erin Gray; Yunhong Shu; Matthew A Frick; Norbert G Campeau; Joshua Trzasko; John Huston; Matt A Bernstein
Journal:  Magn Reson Med       Date:  2018-03-13       Impact factor: 4.668

3.  Application of Adaptive Image Receive Coil Technology for Whole-Brain Imaging.

Authors:  Petrice M Cogswell; Joshua D Trzasko; Erin M Gray; Norbert G Campeau; Phillip J Rossman; Daehun Kang; Fraser Robb; Robert S Stormont; Scott A Lindsay; Matt A Bernstein; Kiaran P McGee; John Huston
Journal:  AJR Am J Roentgenol       Date:  2020-11-25       Impact factor: 3.959

4.  250 MHz Rapid Scan Cross Loop Resonator.

Authors:  Laura A Buchanan; Lukas B Woodcock; George A Rinard; Richard W Quine; Yilin Shi; Sandra S Eaton; Gareth R Eaton
Journal:  Appl Magn Reson       Date:  2018-10-03       Impact factor: 0.831

5.  The effect of concomitant fields in fast spin echo acquisition on asymmetric MRI gradient systems.

Authors:  Shengzhen Tao; Paul T Weavers; Joshua D Trzasko; John Huston; Yunhong Shu; Erin M Gray; Thomas K F Foo; Matt A Bernstein
Journal:  Magn Reson Med       Date:  2017-06-22       Impact factor: 4.668

6.  Optimization of MRI Gradient Coils With Explicit Peripheral Nerve Stimulation Constraints.

Authors:  Mathias Davids; Bastien Guerin; Valerie Klein; Lawrence L Wald
Journal:  IEEE Trans Med Imaging       Date:  2020-12-29       Impact factor: 10.048

7.  Improving apparent diffusion coefficient accuracy on a compact 3T MRI scanner using gradient nonlinearity correction.

Authors:  Ashley T Tao; Yunhong Shu; Ek T Tan; Joshua D Trzasko; Shengzhen Tao; Robert D Reid; Paul T Weavers; John Huston; Matt A Bernstein
Journal:  J Magn Reson Imaging       Date:  2018-09-26       Impact factor: 4.813

8.  The effect of spiral trajectory correction on pseudo-continuous arterial spin labeling with high-performance gradients on a compact 3T scanner.

Authors:  Daehun Kang; Uten Yarach; Myung-Ho In; Erin M Gray; Joshua D Trzasko; Hang Joon Jo; Yunhong Shu; John Huston; Matt A Bernstein
Journal:  Magn Reson Med       Date:  2019-12-04       Impact factor: 4.668

9.  Reduced acoustic noise in diffusion tensor imaging on a compact MRI system.

Authors:  Ek T Tan; Christopher J Hardy; Yunhong Shu; Myung-Ho In; Arnaud Guidon; John Huston; Matt A Bernstein; Thomas K F Foo
Journal:  Magn Reson Med       Date:  2017-10-02       Impact factor: 4.668

10.  The benefit of high-performance gradients on echo planar imaging for BOLD-based resting-state functional MRI.

Authors:  Daehun Kang; Hang Joon Jo; Myung-Ho In; Uten Yarach; Nolan K Meyer; Lydia J Bardwell Speltz; Erin M Gray; Joshua D Trzasko; John Huston Iii; Matt A Bernstein; Yunhong Shu
Journal:  Phys Med Biol       Date:  2020-11-27       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.