Distortion-free imaging: A double encoding method (DIADEM) combined with multiband imaging for rapid distortion-free high-resolution diffusion imaging on a compact 3T with high-performance gradients.

Myung-Ho In1, Ek Tsoon Tan2, Joshua D Trzasko1, Yunhong Shu1, Daehun Kang1, Uten Yarach1, Shengzhen Tao1, Erin M Gray1, John Huston1, Matt A Bernstein1.   

Abstract

BACKGROUND: Distortion-free, high-resolution diffusion imaging using DIADEM (Distortion-free Imaging: A Double Encoding Method), proposed recently, has great potential for clinical applications. However, it can suffer from prolonged scan times and its reliability for quantitative diffusion imaging has not been evaluated.
PURPOSE: To investigate the clinical feasibility of DIADEM-based high-resolution diffusion imaging on a novel compact 3T (C3T) by evaluating the reliability of quantitative diffusion measurements and utilizing both the high-performance gradients (80 mT/m, 700 T/m/s) and the sequence optimization with the navigator acquisition window reduction and simultaneous multislice (multiband) imaging. STUDY TYPE: Prospective feasibility study. PHANTOM/
SUBJECTS: Diffusion quality control phantom scans to evaluate the reliability of quantitative diffusion measurements; 36 normal control scans for B0 -field mapping; six healthy and two patient subject scans with a brain tumor for comparisons of diffusion and anatomical imaging. FIELD STRENGTH/SEQUENCE: 3T; the standard single-shot echo-planar-imaging (EPI), multishot DIADEM diffusion, and anatomical (2D-FSE [fast-spin-echo], 2D-FLAIR [fluid-attenuated-inversion-recovery], and 3D-MPRAGE [magnetization prepared rapid acquisition gradient echo]) imaging. ASSESSMENT: The scan time reduction, the reliability of quantitative diffusion measurements, and the clinical efficacy for high-resolution diffusion imaging in healthy control and brain tumor volunteers. STATISTICAL TEST: Bland-Altman analysis.
RESULTS: The scan time for high in-plane (0.86 mm2 ) resolution, distortion-free, and whole brain diffusion imaging were reduced from 10 to 5 minutes with the sequence optimizations. All of the mean apparent diffusion coefficient (ADC) values in phantom were within the 95% confidence interval in the Bland-Altman plot. The proposed acquisition with a total off-resonance coverage of 597.2 Hz wider than the expected bandwidth of 500 Hz in human brain could yield a distortion-free image without foldover artifacts. Compared with EPI, therefore, this approach allowed direct image matching with the anatomical images and enabled improved delineation of the tumor boundaries. DATA
CONCLUSION: The proposed high-resolution diffusion imaging approach is clinically feasible on C3T due to a combination of hardware and sequence improvements. LEVEL OF EVIDENCE: 3 TECHNICAL EFFICACY: Stage 1 J. Magn. Reson. Imaging 2020;51:296-310.
© 2019 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  DIADEM; compact 3T; diffusion imaging; distortion-free; geometric distortion; multishot EPI

Mesh:

Year:  2019        PMID: 31111581      PMCID: PMC6864220          DOI: 10.1002/jmri.26792

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  39 in total

1.  Diffusion imaging in brain tumors.

Authors:  Andrei I Holodny; Martin Ollenschlager
Journal:  Neuroimaging Clin N Am       Date:  2002-02       Impact factor: 2.264

2.  Dynamic field map estimation using a spiral-in/spiral-out acquisition.

Authors:  Bradley P Sutton; Douglas C Noll; Jeffrey A Fessler
Journal:  Magn Reson Med       Date:  2004-06       Impact factor: 4.668

3.  Readout-segmented EPI for rapid high resolution diffusion imaging at 3 T.

Authors:  Samantha J Holdsworth; Stefan Skare; Rexford D Newbould; Raphael Guzmann; Nikolas H Blevins; Roland Bammer
Journal:  Eur J Radiol       Date:  2007-11-05       Impact factor: 3.528

4.  Correction for geometric distortion in echo planar images from B0 field variations.

Authors:  P Jezzard; R S Balaban
Journal:  Magn Reson Med       Date:  1995-07       Impact factor: 4.668

5.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

6.  Toward Precision and Reproducibility of Diffusion Tensor Imaging: A Multicenter Diffusion Phantom and Traveling Volunteer Study.

Authors:  E M Palacios; A J Martin; M A Boss; F Ezekiel; Y S Chang; E L Yuh; M J Vassar; D M Schnyer; C L MacDonald; K L Crawford; A Irimia; A W Toga; P Mukherjee
Journal:  AJNR Am J Neuroradiol       Date:  2016-12-22       Impact factor: 3.825

7.  Peripheral nerve stimulation characteristics of an asymmetric head-only gradient coil compatible with a high-channel-count receiver array.

Authors:  Seung-Kyun Lee; Jean-Baptiste Mathieu; Dominic Graziani; Joseph Piel; Eric Budesheim; Eric Fiveland; Christopher J Hardy; Ek Tsoon Tan; Bruce Amm; Thomas K-F Foo; Matt A Bernstein; John Huston; Yunhong Shu; John F Schenck
Journal:  Magn Reson Med       Date:  2015-12-02       Impact factor: 4.668

8.  Model-based iterative reconstruction for single-shot EPI at 7T.

Authors:  Uten Yarach; Myung-Ho In; Itthi Chatnuntawech; Berkin Bilgic; Frank Godenschweger; Hendrik Mattern; Alessandro Sciarra; Oliver Speck
Journal:  Magn Reson Med       Date:  2017-02-10       Impact factor: 4.668

9.  A robust multi-shot scan strategy for high-resolution diffusion weighted MRI enabled by multiplexed sensitivity-encoding (MUSE).

Authors:  Nan-Kuei Chen; Arnaud Guidon; Hing-Chiu Chang; Allen W Song
Journal:  Neuroimage       Date:  2013-01-28       Impact factor: 6.556

10.  Template-based field map prediction for rapid whole brain B0 shimming.

Authors:  Yuhang Shi; S Johanna Vannesjo; Karla L Miller; Stuart Clare
Journal:  Magn Reson Med       Date:  2017-11-28       Impact factor: 4.668

View more
  2 in total

1.  The effect of spiral trajectory correction on pseudo-continuous arterial spin labeling with high-performance gradients on a compact 3T scanner.

Authors:  Daehun Kang; Uten Yarach; Myung-Ho In; Erin M Gray; Joshua D Trzasko; Hang Joon Jo; Yunhong Shu; John Huston; Matt A Bernstein
Journal:  Magn Reson Med       Date:  2019-12-04       Impact factor: 4.668

2.  The benefit of high-performance gradients on echo planar imaging for BOLD-based resting-state functional MRI.

Authors:  Daehun Kang; Hang Joon Jo; Myung-Ho In; Uten Yarach; Nolan K Meyer; Lydia J Bardwell Speltz; Erin M Gray; Joshua D Trzasko; John Huston Iii; Matt A Bernstein; Yunhong Shu
Journal:  Phys Med Biol       Date:  2020-11-27       Impact factor: 3.609

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.