Continuous 24-hour measurement of intraocular pressure in millimeters of mercury (mmHg) using a novel contact lens sensor: Comparison with pneumatonometry.

Kevin Gillmann1, Robert Wasilewicz2, Kirsten Hoskens1, Sonja Simon-Zoula3, Kaweh Mansouri1,4.   

Abstract

PURPOSE: To address the unmet need of continuous IOP monitoring, a Pressure-Measuring Contact Lens (PMCL) was developed to measure IOP in millimeters of mercury (mmHg) continuously over 24 hours. The present study assessed the reliability of the novel PMCL.
METHODS: In this prospective open-label clinical study, healthy and open-angle glaucoma (OAG) subjects were fitted with the PMCL, and pneumatonometry was performed on study eyes (in absence of the PMCL) and on fellow eyes before, during, and after provocative tests. The primary outcome measures were (1) mean IOP difference between same-eye measurements, and (2) percentage of timepoints at which IOP measured by the PMCL was within 5 mmHg of that measured by pneumatonometry in the fellow eye.
RESULTS: Eight subjects were analysed (4 healthy, 4 OAG). The average difference in successive IOP measurements made by pneumatonometry and with the PMCL was 2.0±4.3mmHg at placement-time, and 6.5±15.2mmHg at removal time. During water drinking test, a significant increase in IOP was detected both by PMCL in the study eye (2.4±2.5mmHg, p = 0.03) and by pneumatonometry in the fellow eye (1.9±1.9mmHg, p = 0.02). Over the 24-hour recording, 88.0% of IOP variations measured by the PMCL were within 5mmHg of that measured with the pneumatonometer in the fellow eye. A transient corneal erosion of severe intensity was observed following removal of the PMCL on one single eye, and may have affected measurement accuracy in that eye.
CONCLUSIONS: This study is a proof-of-concept for this novel PMCL, and its results are encouraging, with a fair accuracy in IOP values measurement and good sensitivity to subtle IOP variations.

Entities:  

Year:  2021        PMID: 33755676      PMCID: PMC7987168          DOI: 10.1371/journal.pone.0248211

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


  31 in total

Review 1.  Primary open-angle glaucoma.

Authors:  Robert N Weinreb; Peng Tee Khaw
Journal:  Lancet       Date:  2004-05-22       Impact factor: 79.321

2.  The Clinical Interpretation of Changes in Intraocular Pressure Measurements Using Goldmann Applanation Tonometry: A Review.

Authors:  John G Pearce; Ted Maddess
Journal:  J Glaucoma       Date:  2019-04       Impact factor: 2.503

Review 3.  24-h monitoring devices and nyctohemeral rhythms of intraocular pressure.

Authors:  Florent Aptel; Robert N Weinreb; Christophe Chiquet; Kaweh Mansouri
Journal:  Prog Retin Eye Res       Date:  2016-07-28       Impact factor: 21.198

4.  Telemetric Measurement of Intraocular Pressure via an Implantable Pressure Sensor-12-Month Results from the ARGOS-02 Trial.

Authors:  Lars Choritz; Kaweh Mansouri; Jacqueline van den Bosch; Melanie Weigel; H Burkhard Dick; Markus Wagner; Hagen Thieme
Journal:  Am J Ophthalmol       Date:  2019-09-20       Impact factor: 5.258

5.  Tolerability of 24-hour intraocular pressure monitoring of a pressure-sensitive contact lens.

Authors:  Katrin Lorenz; Christina Korb; Nicola Herzog; Jan M Vetter; Heike Elflein; Munir M Keilani; Norbert Pfeiffer
Journal:  J Glaucoma       Date:  2013 Apr-May       Impact factor: 2.503

6.  Diurnal tension curves for assessing the development or progression of glaucoma: an evidence-based analysis.

Authors: 
Journal:  Ont Health Technol Assess Ser       Date:  2011-06-01

7.  Comparison of current tonometry techniques in measurement of intraocular pressure.

Authors:  Behrooz Kouchaki; Hassan Hashemi; Abbasali Yekta; Mehdi Khabazkhoob
Journal:  J Curr Ophthalmol       Date:  2016-11-28

8.  Goldmann tonometer error correcting prism: clinical evaluation.

Authors:  Sean McCafferty; Garrett Lim; William Duncan; Eniko T Enikov; Jim Schwiegerling; Jason Levine; Corin Kew
Journal:  Clin Ophthalmol       Date:  2017-05-03

9.  Investigation of intraocular pressure fluctuation as a risk factor of glaucoma progression.

Authors:  Juliane Matlach; Sandra Bender; Jochem König; Harald Binder; Norbert Pfeiffer; Esther M Hoffmann
Journal:  Clin Ophthalmol       Date:  2018-12-18

Review 10.  Diurnal and 24-h Intraocular Pressures in Glaucoma: Monitoring Strategies and Impact on Prognosis and Treatment.

Authors:  Anastasios G Konstas; Malik Y Kahook; Makoto Araie; Andreas Katsanos; Luciano Quaranta; Luca Rossetti; Gábor Holló; Efstathios T Detorakis; Francesco Oddone; Dimitrios G Mikropoulos; Gordon N Dutton
Journal:  Adv Ther       Date:  2018-10-20       Impact factor: 3.845

View more
  1 in total

1.  Intraocular Pressure Changes during Hemodiafiltration with Two different Concentrations of Sodium in the Dialysate.

Authors:  Claudia Lerma; Nadia Saavedra-Fuentes; Jasbeth Ledesma-Gil; Martín Calderón-Juárez; Héctor Pérez-Grovas; Salvador López-Gil
Journal:  Biology (Basel)       Date:  2021-12-23
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.