Accuracy of Goldmann, ocular response analyser, Pascal and TonoPen XL tonometry in keratoconic and normal eyes.

S P Mollan1, J S Wolffsohn, M Nessim, M Laiquzzaman, S Sivakumar, S Hartley, S Shah.   

Abstract

AIM: The aim of this study was to evaluate the practicality and accuracy of tonometers used in routine clinical practice for established keratoconus (KC).
METHODS: This was a prospective study of 118 normal and 76 keratoconic eyes where intraocular pressure (IOP) was measured in random order using the Goldman applanation tonometer (GAT), Pascal dynamic contour tonometer (DCT), Reichert ocular response analyser (ORA) and TonoPen XL tonometer. Corneal hysteresis (CH) and corneal resistance factor (CRF), as calculated by the ORA, were recorded. Central corneal thickness (CCT) was measured using an ultrasound pachymeter.
RESULTS: The difference in IOP values between instruments was highly significant in both study groups (p<0.001). All other IOP measures were significantly higher than those for GAT, except for the Goldmann-correlated IOP (average of the two applanation pressure points) (IOPg) as measured by ORA in the control group and the CH-corrected IOP (corneal-compensated IOP value) (IOPcc) measures in the KC group. CCT, CH and CRF were significantly less in the KC group (p<0.001). Apart from the DCT, all techniques tended to measure IOP higher in eyes with thicker corneas.
CONCLUSION: The DCT and the ORA are currently the most appropriate tonometers to use in KC for the measurement of IOPcc. Corneal factors such as CH and CRT may be of more importance than CCT in causing inaccuracies in applanation tonometry techniques.

Entities:  

Mesh:

Year:  2008        PMID: 18757471     DOI: 10.1136/bjo.2007.136473

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  20 in total

1.  Repeatability of intra-ocular pressure and central corneal thickness measurements provided by a non-contact method of tonometry and pachymetry.

Authors:  Domenico Schiano Lomoriello; Marco Lombardo; Laura Tranchina; Francesco Oddone; Sebastiano Serrao; Pietro Ducoli
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-10-28       Impact factor: 3.117

2.  Comparation of the new rebound tonometer IOPen and the Goldmann tonometer, and their relationship to corneal properties.

Authors:  J Moreno-Montañés; I Gosende; J Caire; M García-Granero
Journal:  Eye (Lond)       Date:  2010-10-01       Impact factor: 3.775

3.  Agreement between diurnal variations of intraocular pressure by Tono-Pen and Goldmann applanation tonometer in patients on topical anti-glaucoma medication.

Authors:  Shikha Gupta; Gautam Sinha; Reetika Sharma; Bhagabat Nayak; Bharat Patil; Bibhuti Kashyap; Abdul Shameer; Tanuj Dada
Journal:  Int Ophthalmol       Date:  2015-03-28       Impact factor: 2.031

Review 4.  Tonometers-which one should I use?

Authors:  Kanza Aziz; David S Friedman
Journal:  Eye (Lond)       Date:  2018-02-19       Impact factor: 3.775

5.  The influence of corneal geometrical and biomechanical properties on tonometry readings in keratoconic eyes.

Authors:  Mustafa Değer Bilgeç; Eray Atalay; Ömer Sözer; Hüseyin Gürsoy; Muzaffer Bilgin; Nilgün Yıldırım
Journal:  Int Ophthalmol       Date:  2019-12-02       Impact factor: 2.031

6.  Utility of the Tono-Pen in Measuring Intraocular Pressure in Trinidad: A Cross-sectional Study.

Authors:  A Billy; P E David; A K Mahabir; C P Seerattan; J M Street; V D Walcott; R J Yarna; D C Murray; R G Maharaj
Journal:  West Indian Med J       Date:  2015-05-12       Impact factor: 0.171

7.  Comparison of Tono-Pen and Goldmann applanation tonometers for measurement of intraocular pressure in healthy children.

Authors:  Yasmin S Bradfield; Brett M Kaminski; Michael X Repka; Michele Melia; Bradley V Davitt; David A Johnson; Raymond T Kraker; Ruth E Manny; Noelle S Matta; Susan Schloff; Katherine K Weise
Journal:  J AAPOS       Date:  2012-03-28       Impact factor: 1.220

8.  Comparison among Ocular Response Analyzer, Corvis ST and Goldmann applanation tonometry in healthy children.

Authors:  Ramin Salouti; Ali Agha Alishiri; Reza Gharebaghi; Mostafa Naderi; Khosrow Jadidi; Ahmad Shojaei-Baghini; Mohammadreza Talebnejad; Zahra Nasiri; Seyedmorteza Hosseini; Fatemeh Heidary
Journal:  Int J Ophthalmol       Date:  2018-08-18       Impact factor: 1.779

9.  Evaluation of a contact lens-embedded sensor for intraocular pressure measurement.

Authors:  Michael D Twa; Cynthia J Roberts; Huikai J Karol; Ashraf M Mahmoud; Paul A Weber; Robert H Small
Journal:  J Glaucoma       Date:  2010-08       Impact factor: 2.503

10.  Which tonometry in eyes with keratoconus?

Authors:  H Altinkaynak; C Kocasarac; H Dundar; N Sayin; N Kara; E Bozkurt; N Duru
Journal:  Eye (Lond)       Date:  2015-12-04       Impact factor: 3.775

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.