Samuel Kyei1,2, Cynthia Pakyennu Gboglu1, Michael Agyemang Kwarteng2,3, Frank Assiamah1.
Abstract
OBJECTIVES: The objective of the study was to explore the usage of the Goldmann applanation tonometry and noncontact tonometry interchangeably in the measurement of intraocular pressure (IOP) in glaucoma patients.Entities:
Keywords: Ghana; Goldmann applanation tonometer; glaucoma; intraocular pressure; noncontact tonometer
Year: 2020 PMID: 33888929 PMCID: PMC8040943 DOI: 10.4103/nmj.NMJ_177_20
Source DB: PubMed Journal: Niger Med J ISSN: 0300-1652
Mean intraocular pressure findings in Goldmann applanation tonometer and noncontact tonometer
| IOP measures | Mean | SD | Median | |
|---|---|---|---|---|
| OD GAT | 17.40 | 441 | 7.48 | 16.00 |
| OD NCT | 20.15 | 441 | 8.30 | 18.00 |
| OS GAT | 16.80 | 441 | 7.49 | 14.00 |
| OS NCT | 19.74 | 441 | 8.31 | 17.00 |
IOP – Intraocular pressure; GAT – Goldmann applanation tonometer; NCT – Noncontact tonometer; SD – Standard deviation; OS – Left eye; OD – Right eye
Correlation test between Goldmann applanation tonometer and noncontact tonometer readings
| IOP measures | Correlation | Significant | ||
|---|---|---|---|---|
| Pair 1 | OD GAT and OD NCT | 441 | 0.871 | 0.000 |
| Pair 2 | OS GAT and OS NCT | 441 | 0.887 | 0.000 |
IOP – Intraocular pressure; GAT – Goldmann applanation tonometer; NCT – Noncontact tonometer; OS – Left eye; OD – Right eye
Figure 1A correlation between intraocular pressure readings with GAT and NCT by participants who in the right eye. The x-axis and y-axis are the intraocular pressure readings. The line represents the equivalence line
Mean differences between Goldmann applanation tonometer and noncontact tonometer readings
| IOP measures | Mean±SD | Significant (two-tailed) |
|---|---|---|
| OD GAT – OD NCT | −2.75±4.09 | 0.000 |
| OS GAT – OD NCT | −2.95±3.84 | 0.000 |
IOP – Intraocular pressure; GAT – Goldmann applanation tonometer; NCT – Noncontact tonometer; SD – Standard deviation; OS – Left eye; OD – Right eye
Figure 2A correlation between intraocular pressure readings with GAT and NCT by participants in the left eye. The x-axis and y-axis are the intraocular pressure readings. The line represents the equivalence line
Distribution of intraocular pressure variations of noncontact tonometer to Goldmann applanation tonometer
| IOP measurement by noncontact | Right eye (%) | Left eye (%) | Total (%) |
|---|---|---|---|
| Equal to GAT measurement | 37 (8.4) | 37 (8.4) | 74 (8.4) |
| Higher than GAT measurement | 333 (74.8) | 350 (79.4) | 683 (77.4) |
| Lower than GAT measurement | 74 (16.8) | 54 (12.2) | 124 (14.2) |
| Total | 441 (100) | 441 (100) | 882 (100) |
IOP – Intraocular pressure; GAT – Goldmann applanation tonometer
Figure 3Bland–Altman test for correspondence between intraocular pressure measurements (mmHg) taken with NCT and GAT in the right eye
Figure 4Bland–Altman test for correspondence between intraocular pressure measurements (mmHg) taken with NCT and GAT in the left eye