Assessment of positional reproducibility in the head and neck on a 1.5-T MR simulator for an offline MR-guided radiotherapy solution.

Yihang Zhou1, Jing Yuan1, Oi Lei Wong1, Winky Wing Ki Fung2, Ka Fai Cheng2, Kin Yin Cheung1, Siu Ki Yu1.   

Abstract

BACKGROUND: Recently, a shuttle-based offline magnetic resonance-guided radiotherapy (MRgRT) approach was proposed. This study aims to evaluate the positional reproducibility in the immobilized head and neck using a 1.5-T MR-simulator (MR-sim) on healthy volunteers.
METHODS: A total of 159 scans of 14 healthy volunteers were conducted on a 1.5-T MR-sim with thermoplastic mask immobilization. MR images with isotropic 1.053 mm3 voxel size were rigidly registered to the first scan based on fiducial, anatomical and gross positions. Mean and standard deviation of positional displacements in translation and rotation were assessed. Systematic error and random errors of positioning in the head and neck on the MR-sim were determined in the translation of, and in the rotation of roll, pitch and yaw.
RESULTS: The systematic error (Σ) of translation in left-right (LR), anterior-posterior (AP) and superior-inferior (SI) direction was 0.57, 0.22 and 0.26 mm for fiducial displacement, 0.28, 0.10 and 0.52 mm for anatomical displacement, and 0.53, 0.22 and 0.49 mm for gross displacement, respectively. The random error (σ) in corresponding translation direction was 2.07, 0.54 and 1.32 mm for fiducial displacement, 1.34, 0.73 and 2.04 mm for anatomical displacement, and 2.24, 0.86 and 2.61 mm for gross displacement. The systematic error and random error of rotation were generally smaller than 1°.
CONCLUSIONS: Our results suggested that high gross positional reproducibility (<1 mm translational and <1° rotational systematic error) could be achieved on an MR-sim for the proposed offline MRgRT.

Entities:  

Keywords:  Positional reproducibility; head and neck; immobilization; magnetic resonance-guided radiotherapy (MRgRT); magnetic resonance-simulator (MR-sim)

Year:  2018        PMID: 30505721      PMCID: PMC6218207          DOI: 10.21037/qims.2018.10.03

Source DB:  PubMed          Journal:  Quant Imaging Med Surg        ISSN: 2223-4306


  32 in total

1.  MRI/linac integration.

Authors:  Jan J W Lagendijk; Bas W Raaymakers; Alexander J E Raaijmakers; Johan Overweg; Kevin J Brown; Ellen M Kerkhof; Richard W van der Put; Björn Hårdemark; Marco van Vulpen; Uulke A van der Heide
Journal:  Radiother Oncol       Date:  2007-11-26       Impact factor: 6.280

2.  Image quality assessment of a 1.5T dedicated magnetic resonance-simulator for radiotherapy with a flexible radio frequency coil setting using the standard American College of Radiology magnetic resonance imaging phantom test.

Authors:  Oi Lei Wong; Jing Yuan; Siu Ki Yu; Kin Yin Cheung
Journal:  Quant Imaging Med Surg       Date:  2017-04

3.  The ViewRay system: magnetic resonance-guided and controlled radiotherapy.

Authors:  Sasa Mutic; James F Dempsey
Journal:  Semin Radiat Oncol       Date:  2014-07       Impact factor: 5.934

Review 4.  The magnetic resonance imaging-linac system.

Authors:  Jan J W Lagendijk; Bas W Raaymakers; Marco van Vulpen
Journal:  Semin Radiat Oncol       Date:  2014-07       Impact factor: 5.934

5.  Radiotherapy treatment planning of prostate cancer using magnetic resonance imaging alone.

Authors:  Young K Lee; Marc Bollet; Geoffrey Charles-Edwards; Maggie A Flower; Martin O Leach; Helen McNair; Elizabeth Moore; Carl Rowbottom; Steve Webb
Journal:  Radiother Oncol       Date:  2003-02       Impact factor: 6.280

6.  [Interfractional and intrafractional setup errors in radiotherapy for tumors analyzed by cone-beam computed tomography].

Authors:  Feng Xu; Jin Wang; Sen Bai; Qing-Feng Xu; Ya-Li Shen; Ren-Ming Zhong
Journal:  Ai Zheng       Date:  2008-10

7.  Comparison of 2D radiographic images and 3D cone beam computed tomography for positioning head-and-neck radiotherapy patients.

Authors:  Heng Li; X Ronald Zhu; Lifei Zhang; Lei Dong; Sam Tung; Anesa Ahamad; K S Clifford Chao; William H Morrison; David I Rosenthal; David L Schwartz; Radhe Mohan; Adam S Garden
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008-04-18       Impact factor: 7.038

8.  MR-guidance--a clinical study to evaluate a shuttle- based MR-linac connection to provide MR-guided radiotherapy.

Authors:  Tilman Bostel; Nils H Nicolay; Jörg G Grossmann; Angela Mohr; Stefan Delorme; Gernot Echner; Peter Häring; Jürgen Debus; Florian Sterzing
Journal:  Radiat Oncol       Date:  2014-01-09       Impact factor: 3.481

9.  MRI-based Assessment of 3D Intrafractional Motion of Head and Neck Cancer for Radiation Therapy.

Authors:  Oliver J Gurney-Champion; Dualta McQuaid; Alex Dunlop; Kee H Wong; Liam C Welsh; Angela M Riddell; Dow-Mu Koh; Uwe Oelfke; Martin O Leach; Christopher M Nutting; Shreerang A Bhide; Kevin J Harrington; Rafal Panek; Kate L Newbold
Journal:  Int J Radiat Oncol Biol Phys       Date:  2017-10-16       Impact factor: 7.038

10.  Assessment of three-dimensional set-up errors in conventional head and neck radiotherapy using electronic portal imaging device.

Authors:  Tejpal Gupta; Supriya Chopra; Avinash Kadam; Jai Prakash Agarwal; P Reena Devi; Sarbani Ghosh-Laskar; Ketayun Ardeshir Dinshaw
Journal:  Radiat Oncol       Date:  2007-12-14       Impact factor: 3.481

View more
  3 in total

1.  Thermoplastic Mask-Induced Contact Dermatitis: A Case Report.

Authors:  Louis Cappelli; Spencer Poiset; Benjamin Greenberger; Voichita Bar-Ad
Journal:  Cureus       Date:  2022-04-04

2.  A pilot study of highly accelerated 3D MRI in the head and neck position verification for MR-guided radiotherapy.

Authors:  Yihang Zhou; Oi Lei Wong; Kin Yin Cheung; Siu Ki Yu; Jing Yuan
Journal:  Quant Imaging Med Surg       Date:  2019-07

3.  Acquisition repeatability of MRI radiomics features in the head and neck: a dual-3D-sequence multi-scan study.

Authors:  Cindy Xue; Jing Yuan; Yihang Zhou; Oi Lei Wong; Kin Yin Cheung; Siu Ki Yu
Journal:  Vis Comput Ind Biomed Art       Date:  2022-04-01
  3 in total