Insufficient stability of pedicle screws in osteoporotic vertebrae: biomechanical correlation of bone mineral density and pedicle screw fixation strength.

Lukas Weiser1, Gerd Huber2, Kay Sellenschloh2, Lennart Viezens3, Klaus Püschel4, Michael M Morlock2, Wolfgang Lehmann3.   

Abstract

PURPOSE: Loosening of pedicle screws is one major complication of posterior spinal stabilisation, especially in the patients with osteoporosis. Augmentation of pedicle screws with cement or lengthening of the instrumentation is widely used to improve implant stability in these patients. However, it is still unclear from which value of bone mineral density (BMD) the stability of pedicle screws is insufficient and an additional stabilisation should be performed. The aim of this study was to investigate the correlation of bone mineral density and pedicle screw fatigue strength as well as to define a threshold value for BMD below which an additional stabilisation is recommended.
METHODS: Twenty-one human T12 vertebral bodies were collected from donors between 19 and 96 years of age and the BMD was measured using quantitative computed tomography. Each vertebral body was instrumented with one pedicle screw and mounted in a servo-hydraulic testing machine. Fatigue testing was performed by implementing a cranio-caudal sinusoidal, cyclic (0.5 Hz) load with stepwise increasing peak force.
RESULTS: A significant correlation between BMD and cycles to failure (r = 0.862, r 2 = 0.743, p < 0.001) as well as for the linearly related fatigue load was found. Specimens with BMD below 80 mg/cm3 only reached 45% of the cycles to failure and only 60% of the fatigue load compared to the specimens with adequate bone quality (BMD > 120 mg/cm3).
CONCLUSIONS: There is a close correlation between BMD and pedicle screw stability. If the BMD of the thoracolumbar spine is less than 80 mg/cm3, stability of pedicle screws might be insufficient and an additional stabilisation should be considered.

Entities:  

Keywords:  BMD; Bone mineral density; Osteoporosis; Pedicle screw; Spine

Mesh:

Year:  2017        PMID: 28391382     DOI: 10.1007/s00586-017-5091-x

Source DB:  PubMed          Journal:  Eur Spine J        ISSN: 0940-6719            Impact factor:   3.134


  35 in total

1.  Pullout performance comparison of novel expandable pedicle screw with expandable poly-ether-ether-ketone shells and cement-augmented pedicle screws.

Authors:  Mehmet Fatih Aycan; Tolga Tolunay; Teyfik Demir; Mesut Emre Yaman; Yusuf Usta
Journal:  Proc Inst Mech Eng H       Date:  2017-01-18       Impact factor: 1.617

2.  Incidence and economic burden of osteoporosis-related fractures in the United States, 2005-2025.

Authors:  Russel Burge; Bess Dawson-Hughes; Daniel H Solomon; John B Wong; Alison King; Anna Tosteson
Journal:  J Bone Miner Res       Date:  2007-03       Impact factor: 6.741

3.  Radiofrequency-activated PMMA-augmentation through cannulated pedicle screws: A cadaver study to determine the biomechanical benefits in the osteoporotic spine.

Authors:  T Karius; C Deborre; D C Wirtz; C Burger; A Prescher; A Fölsch; K Kabir; R Pflugmacher; H Goost
Journal:  Technol Health Care       Date:  2017       Impact factor: 1.285

4.  A Historical Cohort Study of Pedicle Screw Fixation in Thoracic, Lumbar, and Sacral Spinal Fusions.

Authors:  H A Yuan; S R Garfin; C A Dickman; S M Mardjetko
Journal:  Spine (Phila Pa 1976)       Date:  1994-10-15       Impact factor: 3.468

5.  Pedicle screws and bone mineral density.

Authors:  K Kumano; S Hirabayashi; Y Ogawa; Y Aota
Journal:  Spine (Phila Pa 1976)       Date:  1994-05-15       Impact factor: 3.468

6.  Biomechanical analysis of pedicle screws in osteoporotic bone with bioactive cement augmentation using simulated in vivo multicomponent loading.

Authors:  Theodore J Choma; Wesley F Frevert; William L Carson; Nicole P Waters; Ferris M Pfeiffer
Journal:  Spine (Phila Pa 1976)       Date:  2011-03-15       Impact factor: 3.468

Review 7.  Pedicle screw augmentation in osteoporotic spine: indications, limitations and technical aspects.

Authors:  S Hoppe; M J B Keel
Journal:  Eur J Trauma Emerg Surg       Date:  2016-12-19       Impact factor: 3.693

8.  Stability of transpedicle screwing for the osteoporotic spine. An in vitro study of the mechanical stability.

Authors:  K Okuyama; K Sato; E Abe; H Inaba; Y Shimada; H Murai
Journal:  Spine (Phila Pa 1976)       Date:  1993-11       Impact factor: 3.468

9.  Influence of the screw augmentation technique and a diameter increase on pedicle screw fixation in the osteoporotic spine: pullout versus fatigue testing.

Authors:  Rebecca A Kueny; Jan P Kolb; Wolfgang Lehmann; Klaus Püschel; Michael M Morlock; Gerd Huber
Journal:  Eur Spine J       Date:  2014-08-01       Impact factor: 3.134

10.  Biomechanical characteristics of pedicle screws in osteoporotic vertebrae-comparing a new cadaver corpectomy model and pure pull-out testing.

Authors:  Martin Schulze; Dominic Gehweiler; Oliver Riesenbeck; Dirk Wähnert; Michael J Raschke; René Hartensuer; Thomas Vordemvenne
Journal:  J Orthop Res       Date:  2016-04-06       Impact factor: 3.494

View more
  29 in total

1.  Pedicle screw loosening is correlated to chronic subclinical deep implant infection: a retrospective database analysis.

Authors:  Lukas Leitner; Isabella Malaj; Patrick Sadoghi; Florian Amerstorfer; Mathias Glehr; Klaus Vander; Andreas Leithner; Roman Radl
Journal:  Eur Spine J       Date:  2018-04-13       Impact factor: 3.134

2.  Hounsfield units as predictor for cage subsidence and loss of reduction: following posterior-anterior stabilization in thoracolumbar spine fractures.

Authors:  Bernhard Wilhelm Ullrich; Philipp Schenk; Ulrich J Spiegl; Thomas Mendel; Gunther Olaf Hofmann
Journal:  Eur Spine J       Date:  2018-10-19       Impact factor: 3.134

3.  Time to augment?! Impact of cement augmentation on pedicle screw fixation strength depending on bone mineral density.

Authors:  Lukas Weiser; Gerd Huber; Kay Sellenschloh; Lennart Viezens; Klaus Püschel; Michael M Morlock; Wolfgang Lehmann
Journal:  Eur Spine J       Date:  2018-06-09       Impact factor: 3.134

4.  Experimental validation of adaptive pedicle screws-a novel implant concept using shape memory alloys.

Authors:  Michael Werner; Niels Hammer; Christian Rotsch; Isabell Berthold; Mario Leimert
Journal:  Med Biol Eng Comput       Date:  2019-11-18       Impact factor: 2.602

5.  Pedicle screw anchorage of carbon fiber-reinforced PEEK screws under cyclic loading.

Authors:  Richard A Lindtner; Rene Schmid; Thomas Nydegger; Marko Konschake; Werner Schmoelz
Journal:  Eur Spine J       Date:  2018-03-01       Impact factor: 3.134

6.  The sacral screw placement depending on morphological and anatomical peculiarities.

Authors:  Carolin Meyer; Peter Pfannebecker; Jan Siewe; David Grevenstein; Jan Bredow; Peer Eysel; Max Joseph Scheyerer
Journal:  Surg Radiol Anat       Date:  2019-11-23       Impact factor: 1.246

7.  Cortical threaded pedicle screw improves fatigue strength in decreased bone quality.

Authors:  Lukas Weiser; Kay Sellenschloh; Klaus Püschel; Michael M Morlock; Lennart Viezens; Wolfgang Lehmann; Gerd Huber
Journal:  Eur Spine J       Date:  2020-09-17       Impact factor: 3.134

8.  Pulmonary cement embolism following cement-augmented fenestrated pedicle screw fixation in adult spinal deformity patients with severe osteoporosis (analysis of 2978 fenestrated screws).

Authors:  Onur Levent Ulusoy; Sinan Kahraman; Isik Karalok; Emel Kaya; Meric Enercan; Cem Sever; Burak Abay; Selhan Karadereler; Azmi Hamzaoglu
Journal:  Eur Spine J       Date:  2018-04-18       Impact factor: 3.134

9.  Effect of zoledronic acid on lumbar spinal fusion in osteoporotic patients.

Authors:  Qirui Ding; Jian Chen; Jin Fan; Qingqing Li; Guoyong Yin; Lipeng Yu
Journal:  Eur Spine J       Date:  2017-09-01       Impact factor: 3.134

10.  Vertebral body changes after continuous spinal distraction in scoliotic children.

Authors:  Sebastian Lippross; Antonia Grages; Katja A Lueders; Lena Braunschweig; Friederike Austein; Konstantinos Tsaknakis; Heiko M Lorenz; Anna K Hell
Journal:  Eur Spine J       Date:  2021-02-22       Impact factor: 3.134

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.