Literature DB >> 8059273

Pedicle screws and bone mineral density.

K Kumano1, S Hirabayashi, Y Ogawa, Y Aota.   

Abstract

STUDY
DESIGN: Thirty-five consecutive patients who underwent surgeries for decompression, and one-level fusions with Cotrel-Dubousset pedicle screws for spinal canal stenosis were included in this study. The relationship between bone mineral density of the lumbar spine and the rate of successful fusion and screw problems was studied.
OBJECTIVES: Bone mineral density of the third lumbar vertebral body was measured by quantitative computed tomography preoperatively, and the degree of osteoporosis was graded by radiographic films of lumbar spines. The films were followed with anteroposterior and lateral dynamic radiographic examinations of lumbar spines. These examinations were evaluated clinically after surgery. Follow-up duration was at least 19 months, with a mean of 28 months. SUMMARY OF BACKGROUND DATA: The mean bone mineral density of all the patients was 117.1 mg/mL (SD = 42.5). The rate of successful fusion was 88.6%, the non-union rate was 5.7%, and the rate of fusions that were undetermined was 5.7%. The incidence of screw loosening occurred in 5.7% of the patients, screw breakage occurred in 5.7% of patients, and symptomatic screws related to non-union were present in 5.7% of patients. The degree of bone mineral density or radiographic gradings of osteoporosis and the successful fusion or the incidence of symptomatic screws, which were defined as screws related to non-union, were not related.
METHODS: The patients were divided into two groups. One group consisted of 12 patients who had less than 100 mg/mL of bone mineral density, with a mean of 72.0 mg/mL (SD = 21.2). The other group consisted of 23 patients who had more than 100 mg/mL of bone mineral density, with a mean of 140.6 mg/mL (SD = 30.0). The rate of successful fusion or non-union and the rate of screw problems such as screw loosening or screw breakage were compared in the two groups. The presence of bony trabeculation in grafted bone is the most important criterion for successful fusion.
RESULTS: A statistical difference did not occur between the two groups. Non-union, screw breakage, and screw loosening were observed in patients with both increased and decreased bone mineral density and those with increased and decreased grade osteoporosis. Radiographic grading of osteoporosis was not correlated to bone mineral density, but patients with unexpected bone mineral density were frequent in this series.
CONCLUSIONS: The use of decompression and pedicle screws for spinal canal stenosis due to degenerative lumbar disorders can be done safely with one-level fusion in patients with decreased bone mineral density if patients with grade III osteoporosis are excluded. Bone mineral density is more reliable than radiographic grading to evaluate the degree of osteoporosis and should be included in the preoperative evaluation of patients with osteoporosis.

Entities:  

Mesh:

Year:  1994        PMID: 8059273     DOI: 10.1097/00007632-199405001-00012

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  13 in total

1.  Pullout strength of anterior spinal instrumentation: a product comparison of seven screws in calf vertebral bodies.

Authors:  Konrad Seller; Dieter Wahl; Alexander Wild; Rüdiger Krauspe; Erich Schneider; Berend Linke
Journal:  Eur Spine J       Date:  2007-02-02       Impact factor: 3.134

Review 2.  Surgical treatment options in patients with impaired bone quality.

Authors:  Norman A Johanson; Jody Litrenta; Jay M Zampini; Frederic Kleinbart; Haviva M Goldman
Journal:  Clin Orthop Relat Res       Date:  2011-08       Impact factor: 4.176

3.  Influence of alendronate and endplate degeneration to single level posterior lumbar spinal interbody fusion.

Authors:  Sang-Mok Kim; Wootack Rhee; Seongil Ha; Jae Hyeon Lim; Il Tae Jang
Journal:  Korean J Spine       Date:  2014-12-31

4.  Polymethylmethacrylate augmentation of the pedicle screw: the cement distribution in the vertebral body.

Authors:  Ming-Hsien Hu; Hung Ta H Wu; Ming-Chau Chang; Wing-Kuang Yu; Shih-Tien Wang; Chien-Lin Liu
Journal:  Eur Spine J       Date:  2011-05-01       Impact factor: 3.134

5.  Risk factors associated with the halo phenomenon after lumbar fusion surgery and its clinical significance.

Authors:  Ho-Joong Kim; Sul-Gee Kim; Hwan-Mo Lee; Hak-Sun Kim; Eun-Su Moon; Jin-Oh Park; Nam-Heon Seol; Seong-Hwan Moon
Journal:  Asian Spine J       Date:  2008-06-30

Review 6.  Pedicle screw loosening: a clinically relevant complication?

Authors:  Fabio Galbusera; David Volkheimer; Sandra Reitmaier; Nikolaus Berger-Roscher; Annette Kienle; Hans-Joachim Wilke
Journal:  Eur Spine J       Date:  2015-01-24       Impact factor: 3.134

7.  Insufficient stability of pedicle screws in osteoporotic vertebrae: biomechanical correlation of bone mineral density and pedicle screw fixation strength.

Authors:  Lukas Weiser; Gerd Huber; Kay Sellenschloh; Lennart Viezens; Klaus Püschel; Michael M Morlock; Wolfgang Lehmann
Journal:  Eur Spine J       Date:  2017-04-08       Impact factor: 3.134

8.  The role of bone mineral density in adult spinal deformity patients undergoing corrective surgery: a matched analysis.

Authors:  Syed I Khalid; Ravi S Nunna; Jennifer S Smith; Rachyl M Shanker; Alecia A Cherney; Kyle B Thomson; Sai Chilakapati; Ankit I Mehta; Owoicho Adogwa
Journal:  Acta Neurochir (Wien)       Date:  2022-08-04       Impact factor: 2.816

9.  Contribution of the xenograft bone plate-screw system in lumbar transpedicular stabilization of dogs: an in-vitro study.

Authors:  Hakan Salci; Sani Sarigul; Seref Dogan; Huseyin Lekesiz; Resat Ozcan; O Sacit Gorgul; Kaya Aksoy
Journal:  J Vet Sci       Date:  2008-06       Impact factor: 1.672

10.  Axial pullout strength comparison of different screw designs: fenestrated screw, dual outer diameter screw and standard pedicle screw.

Authors:  Evangelos Christodoulou; Suresh Chinthakunta; Divya Reddy; Saif Khalil; Thomas Apostolou; Philipp Drees; Konstantinos Kafchitsas
Journal:  Scoliosis       Date:  2015-05-01
View more