| Literature DB >> 28374079 |
Robbert W van Hamersvelt1, Arnold M R Schilham2, Klaus Engelke3, Annemarie M den Harder2, Bart de Keizer4, Harald J Verhaar5, Tim Leiner2, Pim A de Jong2, Martin J Willemink2.
Abstract
OBJECTIVES: To investigate the accuracy of bone mineral density (BMD) quantification using dual-layer spectral detector CT (SDCT) at various scan protocols.Entities:
Keywords: Bone mineral density; Dual energy X-Ray absorptiometry; Dual-energy CT; Dual-layer spectral detector CT; Material decomposition
Mesh:
Year: 2017 PMID: 28374079 PMCID: PMC5579207 DOI: 10.1007/s00330-017-4801-4
Source DB: PubMed Journal: Eur Radiol ISSN: 0938-7994 Impact factor: 5.315
Fig. 1Phantom setup. An anthropomorphic European spine phantom on top of a bone density calibration phantom
Scan protocols
| kVp | mAs | CTDIvol (mGy) | DLP (mGy · cm) | ED (mSv) |
|---|---|---|---|---|
| 140 | 200 | 26.0 | 319.8 | 5.0 |
| 120 | 200 | 18.1 | 222.6 | 3.4 |
| 140 | 100 | 13.0 | 159.9 | 2.5 |
| 120 | 100 | 9.0 | 110.7 | 1.7 |
| 140 | 50 | 6.5 | 80.0 | 1.2 |
| 120 | 50 | 4.5 | 55.4 | 0.8 |
CTDIvol volumetric CT dose index, cm centimetre, DLP dose length product, ED effective radiation dose, estimated as DLP · k (k 0.0153 for 120 kVp and k 0.0155 for 140 kVp), kVp kilovoltage peak, mAs milliampere second, mGy milligray, mSv millisievert
Fig. 2Image analyses and bone mineral density quantification. (A) Axial image with a ROI drawn in the ESP (blue) and BDC (pink and green). (B) Spectral plot for the corresponding ROIs: mean HU versus monochromatic energy level (keV). (C) Attenuation profiles were constructed between 90 and 200 keV in steps of 10 keV. Using in-house developed software, HA concentrations were calculated by fitting the constructed profiles to known attenuation profiles of pure HA and pure water. For this image this concerns ROI S3, an insert with a phantom design value of 200 mg/cm3 HA concentration. Av average, BDC bone density calibration phantom, ESP European spine phantom, HA calcium hydroxyapatite, HU Hounsfield units, keV kilo electron voltage, ROI region of interest, SD standard deviation
Mean measurement errors. Subdivided by nominal design concentrations, mAs and kVp
| True BMD concentration (mg/cm3) | mAs | 120 kVp | 140 kVp | ||
|---|---|---|---|---|---|
| Measurement error | Measurement error | ||||
| mg/cm3 | % | mg/cm3 | % | ||
| 50* | All | -0.9 ± 2.1 | -1.8 ± 4.0 | 1.6 ± 1.1 | 3.1 ± 2.1 |
| 200 | 0.3 ± 1.5 | 0.5 ± 2.9 | 0.9 ± 1.0 | 1.8 ± 1.9 | |
| 100 | -0.2 ± 0.6 | -0.5 ± 1.1 | 1.5 ± 0.6 | 2.9 ± 1.1 | |
| 50 | -2.8 ± 2.6 | -5.5 ± 5.1 | 2.3 ± 1.5 | 4.5 ± 2.9 | |
| 100* | All | -3.0 ± 2.7 | -2.9 ± 2.6 | -0.7 ± 1.8 | -0.7 ± 1.7 |
| 200 | -2.7 ± 1.8 | -2.6 ± 1.8 | -0.5 ± 2.0 | -0.5 ± 1.9 | |
| 100 | -2.1 ± 3.8 | -2.0 ± 3.7 | -0.5 ± 2.0 | -0.5 ± 1.9 | |
| 50 | -4.1 ± 2.2 | -4.0 ± 2.1 | -1.1 ± 1.6 | -1.1 ± 1.5 | |
| 200* | All | -10.6 ± 5.3 | -5.3 ± 2.7 | -6.0 ± 2.7 | -2.9 ± 1.3 |
| 200 | -9.6 ± 4.6 | -4.7 ± 2.3 | -6.4 ± 1.6 | -3.2 ± 0.8 | |
| 100 | -10.5 ± 5.2 | -5.2 ± 2.6 | -6.7 ± 3.1 | -3.3 ± 1.5 | |
| 50 | -11.8 ± 6.7 | -5.8 ± 3.4 | -4.8 ± 3.1 | -2.4 ± 1.6 | |
Data are given as mean ± standard deviation. For 50 mg/cm3 nine measurements and for 100 and 200 mg/cm3 18 measurement were performed at both 120 and 140 kVp (three and six,respectively, per mAs value)
BMD bone mineral density, kVp kilovoltage peak, mAs milliampere second
* Phantom specific concentrations (ESP-143; 51.2, 102.3 and 201.2 mg/cm3 HA and BDC-03-29; 104.4 and 206.2 mg/cm3 HA) were used to determine measurement error
Fig. 3Accuracy of dual-layer detector spectral CT for bone mineral density quantification. Mean relative errors (%) per protocol per nominal design HA concentration are shown. Symbols indicate mean measurement error (%) and error bars standard deviation. * = Phantom specific concentrations (ESP-143; 51.2, 102.3 and 201.2 mg/cm3 HA and BDC-03-29; 104.4 and 206.2 mg/cm3 HA) were used to determine deviation. CTDIvol volumetric CT dose index, ESP European spine phantom, HA calcium hydroxyapatite, kVp kilovoltage peak, mAs milliampere second, mGy milligray, SD standard deviation
Fig. 4Comparison of BMD measurements on SDCT and aBMD measurements on DXA. Scatter plots with linear fit show a strong correlation. aBMD areal bone mineral density, BMD bone mineral density, DXA dual energy X-ray absorptiometry, kVp kilovoltage peak, SDCT dual-layer spectral detector computed tomography