Explant analysis of the Biomet Magnum/ReCap metal-on-metal hip joint.

S C Scholes1, B J Hunt2, V M Richardson2, D J Langton3, E Smith4, T J Joyce2.   

Abstract

OBJECTIVES: The high revision rates of the DePuy Articular Surface Replacement (ASR) and the DePuy ASR XL (the total hip arthroplasty (THA) version) have led to questions over the viability of metal-on-metal (MoM) hip joints. Some designs of MoM hip joint do, however, have reasonable mid-term performance when implanted in appropriate patients. Investigations into the reasons for implant failure are important to offer help with the choice of implants and direction for future implant designs. One way to assess the performance of explanted hip prostheses is to measure the wear (in terms of material loss) on the joint surfaces.
METHODS: In this study, a coordinate measuring machine (CMM) was used to measure the wear on five failed cementless Biomet Magnum/ReCap/ Taperloc large head MoM THAs, along with one Biomet ReCap resurfacing joint. Surface roughness measurements were also taken. The reason for revision of these implants was pain and/or adverse reaction to metal debris (ARMD) and/or elevated blood metal ion levels.
RESULTS: The mean wear rate of the articulating surfaces of the heads and acetabular components of all six joints tested was found to be 6.1 mm3/year (4.1 to 7.6). The mean wear rate of the femoral head tapers of the five THAs was 0.054 mm3/year (0.021 to 0.128) with a mean maximum wear depth of 5.7 µm (4.3 to 8.5).
CONCLUSION: Although the taper wear was relatively low, the wear from the articulating surfaces was sufficient to provide concern and was potentially large enough to have been the cause of failure of these joints. The authors believe that patients implanted with the ReCap system, whether the resurfacing prosthesis or the THA, should be closely monitored.Cite this article: S. C. Scholes, B. J. Hunt, V. M. Richardson, D. J. Langton, E. Smith, T. J. Joyce. Explant analysis of the Biomet Magnum/ReCap metal-on-metal hip joint. Bone Joint Res 2017;6:113-122. DOI: 10.1302/2046-3758.62.BJR-2016-0130.R2.
© 2017 Scholes et al.

Entities:  

Keywords:  Arthroplasty; Film Thickness; Hip; Lubrication; Metal-on-Metal; Taper; Wear

Year:  2017        PMID: 28246095      PMCID: PMC5331178          DOI: 10.1302/2046-3758.62.BJR-2016-0130.R2

Source DB:  PubMed          Journal:  Bone Joint Res        ISSN: 2046-3758            Impact factor:   5.853


How does the ex vivo wear of failed Biomet ReCap joints compare with that of similar MoM hip joint designs? The wear rates of each of these THAs and surface replacement ReCap joints (4.1 to 7.6 mm3/year) were higher than those considered to be able to cause wear-related failures in similar large diameter monoblock MoM hips. The wear of the joints measured in this study was mainly found to be from the articulating surfaces. The volumetric wear of the ReCap joint has been measured and is reported for the first time using a viable technique for both the articulating surfaces (of the THAs and the resurfacing) and the taper junctions of the THA. This paper reports on a relatively low sample size.

Introduction

In an attempt to minimise the occurrence of failure of hip prostheses due to wear particle-induced osteolysis, there was a resurgence of metal-on-metal (MoM) hip joints between the mid 1990s and mid 2000s.[1-3] These all-metal hip joints had a larger diameter than those that had been available in previous years in an attempt to optimise lubrication and the range of motion available for use, as well as reduce the incidence of dislocation. Larger diameter all-metal hips were often targeted at the younger, more active patient. There were many designs of these large diameter (⩾ 36 mm) MoM hip joints and there was often the option to have either a surface replacement (hip resurfacing) joint or a total hip prosthesis. Some designs of MoM hip joint, such as the DePuy Articular Surface Replacement (ASR) and the ASR XL total hip arthroplasty (THA), have shown high rates of wear and revision.[4-7] Many failures of these prostheses were attributed to adverse reaction to metal debris (ARMD).[4] It has been suggested that the relatively shallow acetabular component used in both the ASR surface replacement and ASR XL predisposed the implant to edge wear.[4] In addition to this, reports have been published discussing the taper wear of the ASR XL and the possibility of this taper wear being a contributing factor in the failure of MoM THAs.[6,8-10] The high rates of failure of the ASR and ASR XL are not alone among MoM hips. High revision rates have also been reported with the Zimmer Durom,[11] the DePuy Pinnacle,[12] and for the smaller diameter (⩽ 46 mm) Smith & Nephew Birmingham Hip Resurfacing (BHR).[13,14] This view has led to a reduction in the use of MoM hip prostheses, but there are still a variety of designs available. Some of these had acceptable mid-term performance in appropriate patients.[15,16] Recently, in the United Kingdom, the acceptable survivorship for hip replacements given by the National Institute for Health and Care Excellence (NICE) has been changed from 90% at ten years to 95% at ten years.[17] It is important to monitor the in vivo performance of all hip prostheses in order to protect the patients from further surgery by becoming more knowledgeable regarding those implants that are performing well, and those that are not. One way to assess the performance of explanted hip prostheses is to measure the material loss from the bearing surfaces.[18] A previous study has shown that wear volumes of a contemporary design of MoM hip joint higher than 2.3 mm3 per year can result in a wear-related failure of the prosthesis.[6] Another paper stated “Wear rates above 1 mm3/million cycles were consistent with metal ion levels above 10 ppm (sic), which may produce adverse reactions clinically”.[19] While many MoM hips have high revision rates, one study on the Biomet ReCap resurfacing MoM hip reported that “the Biomet ReCap and Magnum components are not susceptible to the same design problems resulting in implant recall specific to the Durom and ASR prostheses”.[20] The aim of the study reported here was to use a coordinate measuring machine (CMM) to measure the wear (in terms of material loss) on failed Biomet Magnum/ ReCap/ Taperloc large head MoM THAs, along with a Biomet ReCap resurfacing joint. The articulating surface roughness of these explanted prostheses was also measured. The relationship between wear volume, roughness and failure was then explored.

Patients and Methods

Joints investigated

Between July 2006 and September 2011, 96 patients (110 hips) received the Biomet Magnum/ ReCap/ Taperloc large head MoM THA or the Biomet ReCap resurfacing hip joint as part of the “Magnum Total Hip System International Prospective Data Collection” study at North Bristol NHS Trust (Musculoskeletal Clinical Study 1939). Five patients (six hips) died but the cause of death was unrelated to the hip replacement. Ten patients (11 hips) were lost to follow-up, leaving 93 hips for analysis. A total of 19 patients (20 hips) received the resurfacing joint and 62 patients (73 hips) received the Magnum/ ReCap/ Taperloc THA. There were 38 females (46 hips) and 43 males (47 hips) in the cohort. The mean age at surgery was 57.9 years (26.8 to 73.8) and mean follow-up was 7.1 years (3.7 to 9.2). Eight hips (seven patients) were revised. Two early revisions were performed within the first month, one for a fractured neck of the femur in a resurfacing patient and one for early acetabular component rotation. Six revisions were performed between 31 and 92 months after index surgery; primary diagnosis was osteoarthritis. Written informed consent for retrieval of implants at time of revision was requested from the patient prior to any surgery. This complied with the European Community Regulations. The implant revised at 31 months was not available for investigation in this study. Five cementless Biomet Magnum/ ReCap/ Taperloc large head MoM THAs and one Biomet ReCap resurfacing hip joint were investigated in this study (examples shown in Fig. 1). The resurfacing joint and four of the five THAs (THA Magnum 1 to 4: Table I) formed part of the prospective Bristol ReCap study. THA Magnum 1 and THA Magnum 3 were sequential bilateral revisions from the same patient. For the resurfacing joint and THA Magnum 1 to 4, index and revision surgeries were performed at the Avon Orthopaedic Centre, North Bristol NHS Trust with revision undertaken between September 2012 and February 2014. For the joint labelled THA Magnum 5 in Table I, both the index and revision surgeries were performed at the Royal United Hospital Bath but by different surgeons. This implant was part of the retrieval study reported here, but is not part of the prospective Bristol ReCap study. Mean time in vivo for all of the implants was 6.7 years (5.3 to 7.6) (Table I). The mean age of the patients was 60.3 years (56.2 to 68.1) at primary surgery. The nominal implant diameters ranged from 44 mm to 52 mm. Five of the implants were retrieved from female patients, one was from a male (THA Magnum 4).

a) Biomet Magnum/ ReCap/ Taperloc large head MoM THA and (b) Biomet ReCap resurfacing hip joint.

Table I.

Clinical data for the explants measured in this study

ImplantImplant diameter (mm)Patient age at primary surgery (yrs)Time in vivo (yrs)Acetabular component inclination angleCobalt (µg/l)Chromium (µg/l)Histology (Yes/No)
Resurfacing4456.25.3NANANANo
THA Magnum 14460.37.443°4.754.57Yes
THA Magnum 24658.17.537°13.816.5Yes
THA Magnum 34659.37.645°4.754.57Yes
THA Magnum 45259.85.842°1.962.48Yes
THA Magnum 54868.16.8NA18.113.8No

THA, total hip arthroplasty; NA, data not available

a) Biomet Magnum/ ReCap/ Taperloc large head MoM THA and (b) Biomet ReCap resurfacing hip joint. Clinical data for the explants measured in this study THA, total hip arthroplasty; NA, data not available All patients were revised for hip pain and/or investigations indicative of an ARMD and/or elevated blood metal ion levels. Prior to retrieval, blood was sampled from four patients (five hips) using a stainless steel 21-gauge needle (Becton Dickinson UK Ltd, Oxford, United Kingdom) and collected in a trace element tube that contained sodium ethylenediaminetetraacetic acid (EDTA). The samples were measured by inductively coupled plasma mass spectrometry (ICPMS) for whole blood. Cobalt (Co) and chromium (Cr) levels are represented in µg/l (Table I) and ranged from 1.96 µg/l to 18.1 µg/l for Co, and 2.48 µg/l to 16.5 µg/l for Cr. In accordance with medicines and healthcare products regulatory agency guidelines, the blood metal ion levels in two patients were higher than 7 µg/l and identified as being at risk for metal-on-metal bearings.[13] Further investigations were performed on three patients (four hips) using metal artefact reduction sequence (MARS) for magnetic resonance imaging (MRI). Evidence of a cystic collection or a soft-tissue mass further supported the presence of an ARMD in all four hips. The MARS MRI was not available in the other two patients. Finally, a positive histological analysis (examination of retrieved cellular tissue) of ARMD was confirmed as the diagnosis in four hips and the reason for failure. Histology was not available in two patients (the resurfacing joint and THA Magnum 5). The bearing surfaces of the ReCap resurfacing are identical to those of the THA design. Both the resurfacing joint and THA were made from high carbon (> 0.2%) as-cast cobalt-chromium-molybdenum (CoCrMo). The radial clearance is specified as being between 75 µm and 150 µm with sphericity of less than 5 µm. The femoral head extends approximately 23° beyond a full hemisphere. The outside geometry of the acetabular component is hemispherical. The CoCr THA femoral head is designed to accept a tapered titanium (Ti) alloy sleeve. This Ti sleeve has an internal female taper and is used with a forged Ti Taperloc stem to give a Ti/Ti taper/trunnion junction. These data are taken from the ReCap design rationale booklet.[21] The acetabular component internal arc of cover ranges from 154.6° for a 38 mm diameter implant to 163.6° for the 60 mm diameter implant.[20] All explanted joints were sterilised in 10% formaldehyde solution and cleaned thoroughly before being measured to determine wear volumes.

Equipment

The volumetric wear from the articulating surfaces (femoral head and acetabular component) of the six retrieved implants was measured on a CMM (Legex 322; Mitutoyo UK Ltd, Andover, United Kingdom) with an accuracy of 0.8 µm.[8,18] The volumetric wear of the tapers of the five femoral heads from the Magnum/ ReCap/ Taperloc THAs was also measured on this CMM. No femoral stems were available for analysis. The CMM is a contacting instrument where a stylus is used to map the geometrical profile of the specimen surface.[8,18] A custom-designed Matlab programme was used to provide the volumetric wear of the surfaces. The CMM and Matlab method have been shown to have an accuracy of approximately 0.5 mm3 [18,22] for the wear measurements of the articulating surfaces, and 0.2 mm3 for the wear measurements of the tapers.[8] In addition to the wear measurements, surface roughness measurements were taken on the articulating surfaces using a Zygo NewView 5000 non-contacting white light interferometric profilometer (Zygo Corporation, Middlefield, Connecticut). This has previously been used to measure the roughness of explanted MoM hips.[23-25] The ×10 lens was used with a x2 zoom, giving an area of view of 317 x 238 µm. The manufacturers of the NewView 5000 state that the vertical resolution of this profilometer is better than 1 nm with a 0.99 µm lateral (x, y) resolution for the magnification used. Measurements of S (mean surface roughness over the 3D area of view), S (root mean square surface roughness over the 3D area of view), and S (surface skewness over the 3D area of view) were taken. The skewness of a surface is a measure of symmetry of the profile about the mean line. A positively skewed surface has a predominance of peaks, and a negatively skewed surface has more pronounced valleys. All six heads and all six acetabular components were measured in the unworn region and the worn region. In order to determine the areas of the surfaces that were unworn and worn, the surfaces of these joints were visually inspected – this was then confirmed by images taken using the CMM. Ten measurements were taken within each region and the mean calculated. Surface roughness measurements were also taken on the five femoral head tapers using a Mitutoyo Surftest SJ-210 (Mitutoyo Corporation, Kanagawa, Japan). Four 5 mm line profiles were taken at 90° angles to each other. These profiles were taken in the region where the taper was engaged with the trunnion (as shown by the CMM measurements). The mean R (mean surface roughness of the 2D line profile) and R (root mean square surface roughness of the 2D line profile) values of these four roughness profiles were calculated for each of the five tapers.

Theoretical analysis

Theoretical studies were also performed. The predicted minimum film thickness using the theory by Hamrock and Dowson,[26] along with the surface roughness measurements of the articulating surfaces, was used to calculate the theoretical lubrication regimes acting within the joints.[27] These equations have been used previously in the calculation of the theoretical lubrication regimes of MoM joints,[28] and the equations are given below. The minimum film thickness equation:[26] h is the minimum film thickness, R is the equivalent radius for a ball on plane model (equation shown below), η is the viscosity of the lubricant (taken as 0.010 Pa s, a similar viscosity to pathological synovial fluid under the shear-thinning encountered during the stance phase of the walking cycle[29]), u is the entraining velocity (0.012 ms−1, again, during the stance phase of walking), E′ is the equivalent elastic modulus of the material pairing (equation shown below; 2.3×1011 Pa for CoCrMo) and L is the applied load (taken as 2000 N). The equivalent radius equation: R is the equivalent radius, R is the radius of the femoral head and R is the radius of the acetabular component. The equation for the equivalent elastic modulus: υ and E are the Poisson’s ratio and elastic modulus of the head (0.3 and 2.1×1011 Pa) and υ and E are the Poisson’s ratio and elastic modulus of the component (0.3 and 2.1×1011 Pa). Finally, the calculated minimum film thicknesses, together with the recorded mean values of average root mean square S for the femoral head and acetabular component (S and S) for each joint, were used to determine the dimensionless parameter λ and therefore the theoretical lubrication regimes acting within the joints.[27] Previous work has shown that if the ratio of the minimum film thickness to the combined surface roughness is greater than one but less than three, mixed lubrication is likely.[27] If λ > 3 then a full-fluid film-lubricating regime is predicted,[27] i.e. the asperities of the bearing surfaces are completely separated by the lubricant film. A λ value of less than one indicates boundary lubrication.[27]

Results

Wear of the articulating surfaces

Wear data for the articulating surfaces of the femoral head and acetabular component of each joint are shown in Table II, along with the radial clearances of the unworn sections (as measured on the CMM). These wear values are given as means because the data were found to be normally distributed using the Anderson-Darling test for normality (p = 0.918). The total wear volume for the head and the acetabular component of the THAs and resurfacing joint was between 30.0 mm3 and 57.3 mm3, and the volumetric wear rate was between 4.1 mm3/year and 7.6 mm3/year (mean: 6.1 mm3/year). The mean volumetric wear rate measured for the five THAs was also 6.1 mm3/year, with a range of 4.1 to 7.6 mm3/year. The ReCap resurfacing hip joint provided a wear rate of 5.9 mm3/year. In all cases, the femoral head wear volume was larger than the wear volume from the acetabular component. The mean femoral head:acetabular component wear ratio was 62:38 (54:46 to 69:31). Examples of the images obtained showing the wear profiles for the articulating surfaces of the femoral head and acetabular component are shown in Figure 2. The femoral head wear can be seen as being located close to the pole and the acetabular component wear was mainly at the rim. The measured radial clearances were within the range given in the manufacturer’s specifications.[21]
Table II.

Wear data for each hip joint

JointRadial clearance (µm)Wear (mm3)
Total wear (mm3)Wear rate (mm3/yr)
Femoral headAcetabular component
Resurfacing8421.110.031.15.9
THA Magnum 19217.712.330.04.1
THA Magnum 212031.026.357.37.6
THA Magnum 310825.015.640.65.3
THA Magnum 49826.314.640.97.1
THA Magnum 511029.813.343.16.3
Fig. 2

Examples of worn images (blue/green) of areas on femoral head (left) and acetabular component (right) indicative of all joints measured in this study.

Wear data for each hip joint Examples of worn images (blue/green) of areas on femoral head (left) and acetabular component (right) indicative of all joints measured in this study.

Surface roughness measurements of the articulating surfaces

The surface roughness measurements performed on the unworn and worn regions of the femoral heads and acetabular components are shown in Tables III and IV. It was not possible to distinguish an unworn area on the acetabular component of THA Magnum 3, therefore, no unworn measurements were available for this acetabular component. Figures 3 and 4 show examples of images acquired by the non-contacting profilometer on the unworn (a) and worn (b) regions of the femoral heads and acetabular components, respectively.
Table III.

Surface roughness (S and S) results for the six femoral heads in the unworn and worn regions

JointSa (µm)
Sqh (µm)
Ssk
UnwornWornUnwornWornUnwornWorn
Resurfacing0.048 (0.006)0.022 (0.011)0.064 (0.007)0.043 (0.022)1.112 (0.613)0.277 (1.411)
THA Magnum 10.006 (0.000)0.019 (0.010)0.011 (0.003)0.037 (0.021)1.231 (6.205)-0.326 (5.741)
THA Magnum 20.007 (0.001)0.046 (0.026)0.012 (0.002)0.070 (0.036)2.776 (4.635)0.128 (2.342)
THA Magnum 30.019 (0.003)0.025 (0.032)0.027 (0.004)0.045 (0.040)1.321 (0.359)3.225 (4.458)
THA Magnum 40.008 (0.002)0.068 (0.102)0.011 (0.003)0.092 (0.117)0.719 (1.681)-2.305 (2.537)
THA Magnum 50.033 (0.003)0.038 (0.051)0.047 (0.005)0.062 (0.066)1.655 (1.066)0.382 (2.343)

Mean values with standard deviations shown in brackets

Table IV.

Surface roughness (S and S) results for the six acetabular components in the unworn and worn regions

JointSa (µm)
Sqc (µm)
Ssk
UnwornWornUnwornWornUnwornWorn
Resurfacing0.026 (0.003)0.028 (0.010)0.037 (0.004)0.049 (0.028)1.361 (0.955)-0.712 (2.080)
THA Magnum 10.016 (0.002)0.021 (0.005)0.021 (0.003)0.040 (0.012)0.691 (0.301)-3.642 (1.418)
THA Magnum 20.024 (0.004)0.034 (0.017)0.030 (0.005)0.061 (0.031)0.539 (0.316)0.629 (2.789)
THA Magnum 3NA0.034 (0.014)NA0.059 (0.021)NA-1.257 (2.160)
THA Magnum 40.007 (0.001)0.036 (0.021)0.013 (0.004)0.065 (0.031)3.035 (8.560)-2.416 (2.240)
THA Magnum 50.008 (0.001)0.049 (0.032)0.013 (0.003)0.087 (0.054)1.456 (4.308)-0.466 (1.276)

THA, total hip arthroplasty; NA, data not available

Mean values with standard deviations shown in brackets

Surface profilometry images of a femoral head (a) unworn region showing carbides (S: 0.047 µm; S: 0.929), and (b) worn region showing scratching (S: 0.075 µm; S: 0.473).

Surface profilometry images of an acetabular component (a) unworn region showing carbides (S: 0.035 µm; S: 1.032) and (b) worn region showing scratching (S: 0.054 µm; S: -0.112).

Surface roughness (S and S) results for the six femoral heads in the unworn and worn regions Mean values with standard deviations shown in brackets Surface roughness (S and S) results for the six acetabular components in the unworn and worn regions THA, total hip arthroplasty; NA, data not available Mean values with standard deviations shown in brackets Surface profilometry images of a femoral head (a) unworn region showing carbides (S: 0.047 µm; S: 0.929), and (b) worn region showing scratching (S: 0.075 µm; S: 0.473). Surface profilometry images of an acetabular component (a) unworn region showing carbides (S: 0.035 µm; S: 1.032) and (b) worn region showing scratching (S: 0.054 µm; S: -0.112).

Theoretical lubrication regimes

Using the equations for predicted minimum film thickness, h, and the dimensionless parameter, λ, the theoretical lubrication regime acting within each joint during the stance phase of the walking cycle was calculated. Table V shows the predicted lubrication regimes for each joint using the surface roughness values obtained in the unworn region and the worn region of each component.
Table V.

Predicted minimum film thicknesses (h) and lubrication regimes

JointImplant diameter (mm)hmin (µm)λ
UnwornWorn
Resurfacing440.050.68 Boundary0.78 Boundary
THA Magnum 1440.062.61 Mixed1.14 Mixed
THA Magnum 2460.061.91 Mixed0.67 Boundary
THA Magnum 3460.05NA0.68 Boundary
THA Magnum 4520.073.89 Full-fluid film0.59 Boundary
THA Magnum 5480.061.29 Mixed0.59 Boundary

THA, total hip arthroplasty; NA, data not available

Predicted minimum film thicknesses (h) and lubrication regimes THA, total hip arthroplasty; NA, data not available

Taper wear

The taper angles and wear volumes measured on the female tapers of the five THA femoral heads are shown in Table VI. The volumetric wear for each of the five Biomet Magnum THA tapers ranged from 0.16 to 0.96 mm3, with a mean total volumetric wear of 0.38 mm3 and a mean wear rate of 0.054 mm3/year. The maximum wear depths ranged from 4.6 µm to 8.5 µm, with a mean of 5.7 µm. Indicative images displaying the positioning of the wear on the tapers are shown in Figure 5.
Table VI.

Taper angle and wear data for each femoral head taper

JointTaper angleWear (mm3)Wear rate (mm3/yr)Maximum wear depth (µm)
THA Magnum 14.06°0.190.0264.6
THA Magnum 24.09°0.960.1288.5
THA Magnum 34.00°0.160.0214.3
THA Magnum 44.01°0.240.0416.6
THA Magnum 54.02°0.360.0534.7

THA, total hip arthroplasty

Images of worn (blue/green) areas on the femoral head tapers of the THAs.

Taper angle and wear data for each femoral head taper THA, total hip arthroplasty Images of worn (blue/green) areas on the femoral head tapers of the THAs. Table VII provides the surface roughness results for the tapers of each THA Magnum femoral head. These were measured in the regions that were engaged with the trunnion when implanted.
Table VII.

Taper surface roughness measurements (R and R) (mean values with standard deviations shown in brackets below)

JointRa (µm)Rq (µm)
THA Magnum 10.286 (0.016)0.349 (0.019)
THA Magnum 20.707 (0.124)0.868 (0.138)
THA Magnum 31.177 (0.129)1.331 (0.136)
THA Magnum 40.803 (0.087)0.957 (0.090)
THA Magnum 50.319 (0.063)0.398 (0.094)

THA, total hip arthroplasty

Taper surface roughness measurements (R and R) (mean values with standard deviations shown in brackets below) THA, total hip arthroplasty

Discussion

For all six explants, wear on the femoral head was localised in a region towards the pole (as illustrated in Fig. 2). The wear area of the acetabular component was positioned mainly at the rim. Variable magnitudes of this edge wear was found on all of the acetabular components measured. The acetabular component inclination angles ranged from 37° to 45° (Table I). Surgical positioning has, therefore, been excluded as the reason for this edge wear. It must be noted, however, that the explant with the lowest acetabular component inclination angle (37°) gave the highest wear, along with high blood metal ion levels; this explant also had the highest radial clearance. Acetabular component edge wear is a common feature with failed MoM hip joints.[4,30] Previous work performed on 57 failed DePuy ASR MoM hip joints[6] reported that a volumetric wear rate of 2.3 mm3/year from the articulating surfaces was sufficient to cause revision due to ARMD (the range of wear rates was 2.3 mm3/year to 95.5 mm3/year). The wear rates of each of these THA and surface replacement ReCap joints (4.1 to 7.6 mm3/year) fall within this range. Published work posing a view on wear rates of MoM hips sufficient to lead to revision has been available for many years. In 2003, 22 explanted MoM hips of 28 mm diameter were measured to have a mean wear rate of 2.02 mm3/year (0.55 to 3.74).[31] It was reported that these joints were revised for early aseptic loosening. Therefore, it could have been argued in 2003 that a wear rate of more than 0.55 mm3/year was sufficient to cause failure of a MoM hip. The cumulative percentage probability of revision for the Biomet ReCap resurfacing replacement joint is shown in the 2016 National Joint Registry (NJR) as 7.79% at seven years. This compares with 5.47% for the BHR and 20.88% for the ASR.[7] In the Australian Orthopaedic Association (AOA) National Joint Registry Annual Report 2015, the Biomet ReCap resurfacing has a revision rate of 12.2% at seven years.[32] It is clear that the Biomet ReCap resurfacing replacement has a lower revision rate than the DePuy ASR, and this may be because they are ‘not susceptible to the same design problems’ as the ASR.[20] This may be due to the larger arc of cover provided by the acetabular components. It has been shown[4,8] that the ASR has a lower arc of cover (151° for a 52 mm diameter joint) than the BHR (162° for a 52 mm joint). The arc of cover of the Biomet ReCap resurfacing joint ranges from 154.6° for a 38 mm diameter implant to 163.6° for the 60 mm diameter implant.[20] With regard to the Magnum/ ReCap/ Taperloc THA, the AOA Registry showed the cumulative percentage revision as 8.6% at seven years.[32] No data were available on the Magnum/ ReCap/ Taperloc THA in the NJR.[7] Table VIII shows the cumulative percentage revision of MoM primary THAs at seven years from the Australian Joint Registry. It can be seen that, in comparison with alternative designs, the Magnum/ ReCap/ Taperloc produced only slightly higher cumulative percentage rates of revision at seven years than the BHR.
Table VIII.

Revision rates for metal-on-metal total hip arthroplastys[32]

ManufacturerFemoral componentAcetabular componentCumulative percentage revision at 7 yrs
Smith & NephewSynergyBHR7.4
BiometTaperlocReCap8.6
ZimmerAlloclassicDurom10.4
DePuySummitASR32.4
DePuyCorailASR38.7

BHR, Birmingham Hip Resurfacing; ASR, articular surface replacement

Revision rates for metal-on-metal total hip arthroplastys[32] BHR, Birmingham Hip Resurfacing; ASR, articular surface replacement The revision rates stated in the NJR for MoM THAs are higher than those for the MoM hip resurfacings.[7] The reasons for this were investigated[6,8] and it was concluded that wear at the modular junction between the head taper and trunnion of the stem in the DePuy ASR and the DePuy 36 mm diameter MoM Pinnacle Articuleze joints can result in ARMD. Any taper wear will add to the wear produced at the articulating surfaces and lead to greater failure rates of THA.[6] The taper wear measurements for the Ti taper of the Biomet Magnum THAs (mean: 0.05 mm3/year) were low compared with those measured for the CoCr tapers of the DePuy ASR XL (mean: 2.60 mm3/year) and DePuy Pinnacle Articuleze prostheses (mean 2.80 mm3/year).[8] The wear results displayed in Table V show the low taper wear for this Ti/Ti taper junction in this study. In fact, the wear measurements on the tapers of three of the joints (THA Magnum 1, 3 and 5) were close to or below the 0.2 mm3 accuracy limit of the CMM used for these taper wear measurements. So why did these Biomet joints provide such low taper wear? This may be due to the material combination used at the taper junction; Ti/Ti junctions used with the THA Magnums have been shown to provide lower wear than the Ti stem/CoCr head taper used with the ASR XL and Pinnacle.[33] However, this should not be seen as a panacea as similar materials can ‘cold weld’ and show adhesive wear. Other factors may also be involved. It appears that the taper design used with the Biomet Magnum joints has reduced the wear at the taper junction compared with other designs. It is appreciated that any material released at this Ti/Ti taper junction will not have contributed to the Co and Cr metal ion levels measured. Thus, for the Biomet ReCap explants analysed in this study, failure was due to ARMD likely caused by wear from the articulating surfaces, not the femoral head tapers. In summary, the wear rates for the five retrieved Biomet Magnum/ ReCap/ Taperloc THAs and one Biomet ReCap resurfacing hip joint were higher than those considered to be able to cause wear-related failures in similar large diameter monoblock MoM hips (2.3 mm3/year). The wear of the joints measured in this study was mainly found to be from the articulating surfaces. The authors believe that patients implanted with the Biomet ReCap system should be monitored.
  23 in total

1.  Hip resurfacing with the Biomet Hybrid ReCap-Magnum system: 7-year results.

Authors:  Thomas P Gross; Fei Liu
Journal:  J Arthroplasty       Date:  2012-05-15       Impact factor: 4.757

2.  Modes of implant failure after hip resurfacing: morphological and wear analysis of 267 retrieval specimens.

Authors:  Michael M Morlock; Nick Bishop; Jozef Zustin; Michael Hahn; Wolfgang Rüther; Michael Amling
Journal:  J Bone Joint Surg Am       Date:  2008-08       Impact factor: 5.284

3.  Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: A consequence of excess wear.

Authors:  D J Langton; S S Jameson; T J Joyce; N J Hallab; S Natu; A V F Nargol
Journal:  J Bone Joint Surg Br       Date:  2010-01

4.  Clinical outcome of the metal-on-metal hybrid Corin Cormet 2000 hip resurfacing system: an up to 11-year follow-up study.

Authors:  Thomas P Gross; Fei Liu; Lee A Webb
Journal:  J Arthroplasty       Date:  2011-09-09       Impact factor: 4.757

5.  Metal on metal surface replacement of the hip. Experience of the McMinn prothesis.

Authors:  D McMinn; R Treacy; K Lin; P Pynsent
Journal:  Clin Orthop Relat Res       Date:  1996-08       Impact factor: 4.176

6.  Shorter, rough trunnion surfaces are associated with higher taper wear rates than longer, smooth trunnion surfaces in a contemporary large head metal-on-metal total hip arthroplasty system.

Authors:  Timothy M Brock; Raghavendra Sidaginamale; Steven Rushton; Antoni V F Nargol; John G Bowsher; Christina Savisaar; Tom J Joyce; David J Deehan; James K Lord; David J Langton
Journal:  J Orthop Res       Date:  2015-07-25       Impact factor: 3.494

Review 7.  The future role of metal-on-metal hip resurfacing.

Authors:  Gulraj S Matharu; Hemant G Pandit; David W Murray; Ronan B C Treacy
Journal:  Int Orthop       Date:  2015-02-24       Impact factor: 3.075

8.  The clinical implications of metal debris release from the taper junctions and bearing surfaces of metal-on-metal hip arthroplasty: joint fluid and blood metal ion concentrations.

Authors:  R P Sidaginamale; T J Joyce; J G Bowsher; J K Lord; P J Avery; S Natu; A V F Nargol; D J Langton
Journal:  Bone Joint J       Date:  2016-07       Impact factor: 5.082

9.  Outcomes of a metal-on-metal total hip replacement system.

Authors:  G S Matharu; K Theivendran; P B Pynsent; L Jeys; A M Pearson; D J Dunlop
Journal:  Ann R Coll Surg Engl       Date:  2014-10       Impact factor: 1.891

10.  (iv) Enhancing the safety and reliability of joint replacement implants.

Authors:  Louise M Jennings; Mazen Al-Hajjar; Claire L Brockett; Sophie Williams; Joanne L Tipper; Eileen Ingham; John Fisher
Journal:  Orthop Trauma       Date:  2012-08