Louise M Jennings1, Mazen Al-Hajjar, Claire L Brockett, Sophie Williams, Joanne L Tipper, Eileen Ingham, John Fisher.
Abstract
A new Stratified Approach For Enhanced Reliability (SAFER) pre-clinical simulation testing of joint prostheses is presented in this article. The aim of this approach is preclinical systematic testing of wear performance in the much wider envelope of conditions found clinically rather than relying only on the standard testing conditions that are currently used. The approach includes variations in surgical delivery, variations in kinematics, variations in the patient population and degradation of the biomaterial properties. Clinical experience of existing prostheses has been used to validate the new in vitro methods.Entities:
Year: 2012 PMID: 23335949 PMCID: PMC3546063 DOI: 10.1016/j.mporth.2012.05.006
Source DB: PubMed Journal: Orthop Trauma ISSN: 1877-1327
Wear rates (mm3/million cycles) for hip replacements under standard conditions
| Type | Metal on polyethylene | Metal-on-metal | Ceramic-on-metal | Ceramic-on-ceramic | ||||
|---|---|---|---|---|---|---|---|---|
| Conventional | Cross-linked 7.5–10 MRad | Alumina | Delta | |||||
| Head size (mm) | 28 | 28–36 | 28 | 36 | 39–55 | 28 to 36 | 28 | 28 |
| Wear rate (mm3/million cycles) | 25 to 40 | 5 to 10 | 0.1 to 1 | 0.4 to 0.8 | 0.1 to 0.4 | 0.02 to 0.1 | 0.02 to 0.1 | 0.02 to 0.1 |
Wear rates (mm3/million cycles) for knee replacements under standard conditions
| Type of material | Conventional polyethylene | Cross-linked polyethylene 5 MRad |
|---|---|---|
| Wear rate (mm3/million cycles) | 6 to 12 | 3 to 6 |
Effect of deterioration in femoral head roughness in the hip on wear rates
| Material | Conventional polyethylene | |
|---|---|---|
| Condition | Smooth | Scratched head |
| Wear rate (mm3/million cycles) | 25 to 40 | 100 to 140 |
Effect of different surgical delivery and component positioning on wear in the hip
| Type | Metal-on-metal | Ceramic-on-metal | Ceramic-on-ceramic | |||
|---|---|---|---|---|---|---|
| Alumina | Delta | |||||
| Head size (mm) | 28 | 39 | 39 | 28 | 28 | 28 |
| Adverse condition | Translational mal-position | Rotational mal-position | Translational rotational mal-position | Translational mal-position | Translational mal-position | Translational mal-position |
| Wear rate (mm3/million cycles) | 0.5 to 3 | 1 to 10 | 8 to 14 | 0.1 to 0.5 | 0.5 to 1.8 | 0.5 to 0.25 |
Variation in kinematic conditions in the knee
| Type of material | Conventional polyethylene | Cross-linked polyethylene 5 MRad | Conventional polyethylene |
|---|---|---|---|
| Condition | Increased anterior posterior translation | Increased anterior posterior translation | Lift off |
| Wear rate (mm3/million cycles) | 9 to 15 | 4.5 to 8.5 | 12 to 19 |
Friction factors for size 28 mm hip replacements
| Type of hip | Metal on polyethylene | Metal-on-metal | Ceramic-on-metal | Ceramic-on-ceramic |
|---|---|---|---|---|
| 100% serum | 0.07 | 0.10 | 0.07 | 0.05 |
| 25% serum | 0.06 | 0.12 | 0.05 | 0.04 |
| Water | 0.05 | 0.17 | 0.02 | 0.02 |