Agreement analysis of Lenstar with other four techniques of biometry before cataract surgery.

Jinling Zhang1, Yong Wang2.   

Abstract

PURPOSE: To test whether some biometry measurements provided by the Lenstar LS900 compared well with the AL-Scan, Pentacam rotating Scheimpflug camera, Ultrasound Biomicroscopy (UBM) and Tomey EM-3000.
METHODS: Two hundred and one patients having routine cataract surgery had standard preoperative assessment. In this clinical study, the axis length (AL) and lens thickness (LT) were taken by Lenstar LS900 and AL-Scan; anterior chamber depth (ACD) was taken by Lenstar LS900, A-Scan, Pentacam and UBM; central corneal thickness (CCT) was taken by Lenstar LS900, Pentacam and Tomey EM-3000. The results were compared using a Wilcoxon-Mann-Whitney U test and Pearson correlation calculations. Agreement was assessed through intraclass correlation coefficients and Bland-Altman plots.
RESULTS: The highest correlation was found between Lenstar and AL-Scan for AL (r = 0.975; P < 0.001). For LT measurements, the correlation between these two devices was also good (r = 0.699; P < 0.001). Excellent correlations were showed between Lenstar and Pentacam or UBM for ACD (r = 0.948, 0.704, respectively, both P < 0.001), but not between Lenstar and AL-Scan (r = 0.453, P < 0.001). The correlations of CCT between Lenstar and Pentacam or Tomey EM-3000 were both excellent (r = 0.817, 0.882, respectively, both P < 0.001).
CONCLUSIONS: In phakic eyes of cataract patients, measurements of AL, LT, ACD and CCT from Lenstar LS900 yielded results that correlated very well with other clinical instruments.
© 2022. The Author(s), under exclusive licence to Springer Nature B.V.

Entities:  

Keywords:  Biometry; Cataract surgery; Lenstar LS900

Year:  2022        PMID: 35543852     DOI: 10.1007/s10792-022-02352-z

Source DB:  PubMed          Journal:  Int Ophthalmol        ISSN: 0165-5701            Impact factor:   2.029


  4 in total

1.  [Comparison of swept source anterior segment optical coherence tomography and ultrasound biomicroscopy in measurement of anterior chamber depth and anterior chamber angle data in age-related cataract patients].

Authors:  L Shen; X N Wang; D J Li; Z Y Wang; W Chen; Q Zhao; Y F Li; R Cui; W L Yang
Journal:  Zhonghua Yan Ke Za Zhi       Date:  2018-09-11

Review 2.  Device interchangeability on anterior chamber depth and white-to-white measurements: a thorough literature review.

Authors:  Alberto Domínguez-Vicent; Cari Pérez-Vives; Teresa Ferrer-Blasco; Santiago García-Lázaro; Robert Montés-Micó
Journal:  Int J Ophthalmol       Date:  2016-07-18       Impact factor: 1.779

3.  Comparison of anterior segment measurements with LenStar and Pentacam in patients with newly diagnosed glaucoma.

Authors:  Emine Sen; Merve Inanc; Ufuk Elgin; Pelin Yilmazbas
Journal:  Int Ophthalmol       Date:  2017-01-21       Impact factor: 2.031

4.  [A comparison study of anterior chamber depth using A sonography and biomicroscopy in patient with cataract].

Authors:  Tang Xu; Yi-zhuang Li; Si Zhang
Journal:  Zhonghua Yan Ke Za Zhi       Date:  2013-02
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.