| Literature DB >> 28058533 |
Edward Castillo1,2, Richard Castillo3, Yevgeniy Vinogradskiy4, Thomas Guerrero5.
Abstract
Computed tomography (CT)-derived ventilation imaging utilizes deformable image registration (DIR) to recover respiratory-induced tissue volume changes from inhale/exhale 4DCT phases. While current strategies for validating CT ventilation rely on analyzing its correlation with existing functional imaging modalities, the numerical stability of the CT ventilation calculation has not been characterized.Entities:
Keywords: Computed tomography; Deformable image registration; Functional image analysis; Ventilation
Mesh:
Year: 2017 PMID: 28058533 PMCID: PMC5362676 DOI: 10.1007/s11548-016-1509-x
Source DB: PubMed Journal: Int J Comput Assist Radiol Surg ISSN: 1861-6410 Impact factor: 2.924
Fig. 1Undeformed voxel (green) is mapped by the displacement vectors to create the deformed voxel (purple). Equation (10) describes how the ventilation metric varies when is perturbed by a vector with magnitude . The perturbation implies that the mapped position lies within the red ball of radius of centered on the unperturbed position, denoted
Properties of the inhale/exhale CT image pair (Case 6 from www.dir-lab.com) used for the constrained Jacobian experiments are given in the first column
| Image properties | Spatial transformation | Average mm error (Std.) | Max mm error | Average Jacobian value (std.) |
|---|---|---|---|---|
| Size (voxels): 512 | No DIR | 11.10 (6.98) | 27.59 | |
| Voxel dimensions (MM): 0.97 | Unconstrained DIR | 0.99 (0.99) | 5.37 | 0.79 (0.11) |
| Number of landmarks: 419 | Contraction constraint: LB | 1.08 (1.02) | 5.00 | 0.79 (0.08) |
| Contraction constraint: LB | 1.66 (1.16) | 5.71 | 0.73 (0.03) | |
| Forced spatial correlation with SPECT Perfusion | 1.47 (1.16) | 5.46 | 0.79 (0.03) |
The spatial accuracy results for the unconstrained DIR solution and each of three constrained DIR examples are given in the remaining columns. For all experiments, the average millimeter (MM) error stays below the axial slice spacing
Fig. 2Coronal slice 293 from the four CT ventilation Jacobian images corresponding to the experiments described in Table 1 are shown in the left column. The intensity values indicate the Jacobian measured volume change for each voxel. In the right column, a histogram plot of all Jacobian values for each image ventilation image is shown, with the unconstrained histogram superimposed for reference. The results indicate that for the same inhale/exhale CT image pair, it is possible to compute transformation-based CT images with significantly different physical characteristics, despite the fact that the corresponding DIR solutions maintain subvoxel average mm accuracy
Fig. 3A plot of the unconstrained DIR Jacobian values (x-axis) versus the spatially corresponding SPECT Perfusion values (y-axis) is given in blue. The correlation between the two datasets is . The constrained Jacobian values are also plotted (red). The spatial correlation between the constrained values and the SPECT values is . Despite these seemingly significant differences, the average mm spatial error of the corresponding DIR solutions is 0.99 and 1.47 mm, respectively, both of which are well below the axial slice spacing of the CT image pairs (2.5 mm)