Literature DB >> 28018941

Modularized compact positron emission tomography detector for rapid system development.

Daoming Xi1, Xiang Liu2, Chen Zeng3, Wei Liu1, Yanzhao Li2, Yuexuan Hua3, Xiongze Mei3, Heejong Kim4, Peng Xiao5, Chien-Min Kao4, Qingguo Xie5.   

Abstract

We report the development of a modularized compact positron emission tomography (PET) detector that outputs serial streams of digital samples of PET event pulses via an Ethernet interface using the UDP/IP protocol to enable rapid configuration of a PET system by connecting multiple such detectors via a network switch to a computer. Presently, the detector is [Formula: see text] in extent (excluding I/O connectors) and contains an [Formula: see text] array of [Formula: see text] one-to-one coupled lutetium-yttrium oxyorthosilicate/silicon photomultiplier pixels. It employs cross-wire and stripline readouts to merge the outputs of the 216 detector pixels to 24 channels. Signals at these channels are sampled using a built-in 24-ch, 4-level field programmable gate arrays-only multivoltage threshold digitizer. In the computer, software programs are implemented to analyze the digital samples to extract event information and to perform energy qualification and coincidence filtering. We have developed two such detectors. We show that all their pixels can be accurately discriminated and measure a crystal-level energy resolution of 14.4% to 19.4% and a detector-level coincidence time resolution of 1.67 ns FWHM. Preliminary imaging results suggests that a PET system based on the detectors can achieve an image resolution of [Formula: see text].

Entities:  

Keywords:  PET detector; cross-wire; multivoltage threshold; rapid system development; silicon photomultiplier; stripline

Year:  2016        PMID: 28018941      PMCID: PMC5169362          DOI: 10.1117/1.JMI.4.1.011006

Source DB:  PubMed          Journal:  J Med Imaging (Bellingham)        ISSN: 2329-4302


  13 in total

1.  Maximum likelihood reconstruction for emission tomography.

Authors:  L A Shepp; Y Vardi
Journal:  IEEE Trans Med Imaging       Date:  1982       Impact factor: 10.048

2.  A Multi-Threshold Sampling Method for TOF PET Signal Processing.

Authors:  H Kim; C M Kao; Q Xie; C T Chen; L Zhou; F Tang; H Frisch; W W Moses; W S Choong
Journal:  Nucl Instrum Methods Phys Res A       Date:  2009-04-21       Impact factor: 1.455

3.  Simultaneous PET-MRI: a new approach for functional and morphological imaging.

Authors:  Martin S Judenhofer; Hans F Wehrl; Danny F Newport; Ciprian Catana; Stefan B Siegel; Markus Becker; Axel Thielscher; Manfred Kneilling; Matthias P Lichy; Martin Eichner; Karin Klingel; Gerald Reischl; Stefan Widmaier; Martin Röcken; Robert E Nutt; Hans-Jürgen Machulla; Kamil Uludag; Simon R Cherry; Claus D Claussen; Bernd J Pichler
Journal:  Nat Med       Date:  2008-03-23       Impact factor: 53.440

4.  A feasibility study of a PET/MRI insert detector using strip-line and waveform sampling data acquisition.

Authors:  H Kim; C-T Chen; N Eclov; A Ronzhin; P Murat; E Ramberg; S Los; Alice M Wyrwicz; Limin Li; C-M Kao
Journal:  Nucl Instrum Methods Phys Res A       Date:  2015-06-01       Impact factor: 1.455

Review 5.  PET-MRI: a review of challenges and solutions in the development of integrated multimodality imaging.

Authors:  Stefaan Vandenberghe; Paul K Marsden
Journal:  Phys Med Biol       Date:  2015-02-04       Impact factor: 3.609

6.  Sparse signal recovery methods for multiplexing PET detector readout.

Authors:  Garry Chinn; Peter D Olcott; Craig S Levin
Journal:  IEEE Trans Med Imaging       Date:  2013-02-26       Impact factor: 10.048

7.  First characterization of a digital SiPM based time-of-flight PET detector with 1 mm spatial resolution.

Authors:  Stefan Seifert; Gerben van der Lei; Herman T van Dam; Dennis R Schaart
Journal:  Phys Med Biol       Date:  2013-04-15       Impact factor: 3.609

8.  Multiplexing strategies for monolithic crystal PET detector modules.

Authors:  L A Pierce; W C J Hunter; D R Haynor; L R MacDonald; P E Kinahan; R S Miyaoka
Journal:  Phys Med Biol       Date:  2014-08-22       Impact factor: 3.609

9.  A sub-millimeter resolution PET detector module using a multi-pixel photon counter array.

Authors:  Tae Yong Song; Heyu Wu; Sergey Komarov; Stefan B Siegel; Yuan-Chuan Tai
Journal:  Phys Med Biol       Date:  2010-04-14       Impact factor: 3.609

10.  Photo-detectors for time of flight positron emission tomography (ToF-PET).

Authors:  Virginia Ch Spanoudaki; Craig S Levin
Journal:  Sensors (Basel)       Date:  2010-11-18       Impact factor: 3.576

View more
  2 in total

1.  Multiplexing Readout for Time-of-Flight (TOF) PET Detectors Using Striplines.

Authors:  Heejong Kim; Chien-Min Kao; Yuexuan Hua; Qingguo Xie; Chin-Tu Chen
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2021-01-13

2.  New Digital Plug and Imaging Sensor for a Proton Therapy Monitoring System Based on Positron Emission Tomography.

Authors:  Nicola D'Ascenzo; Min Gao; Emanuele Antonecchia; Paolo Gnudi; Hsien-Hsin Chen; Fang-Hsin Chen; Ji-Hong Hong; Ing-Tsung Hsiao; Tzu-Chen Yen; Weidong Wang; Daoming Xi; Bo Zhang; Qingguo Xie
Journal:  Sensors (Basel)       Date:  2018-09-08       Impact factor: 3.576

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.