Literature DB >> 27980345

Task Equivalence for Model and Human-Observer Comparisons in SPECT Localization Studies.

Anando Sen1, Faraz Kalantari2, Howard C Gifford3.   

Abstract

While mathematical model observers are intended for efficient assessment of medical imaging systems, their findings should be relevant for human observers as the primary clinical end users. We have investigated whether pursuing equivalence between the model and human-observer tasks can help ensure this goal. A localization ROC (LROC) study tested prostate lesion detection in simulated In-111 SPECT imaging with anthropomorphic phantoms. The test images were 2D slices extracted from reconstructed volumes. The iterative OSEM reconstruction method was used with Gaussian postsmoothing. Variations in the number of iterations and the level of postfiltering defined the test strategies in the study. Human-observer performance was compared with that of a visual-search (VS) observer, a scanning channelized Hotelling observer, and a scanning nonprewhitening (CNPW) observer. These model observers were applied with precise information about the target regions of interest (ROIs). ROI knowledge was a study variable for the human observers. In one study format, the humans read the SPECT image alone. With a dual-modality format, the SPECT image was presented alongside an anatomical image slice extracted from the density map of the phantom. Performance was scored by area under the LROC curve. The human observers performed significantly better with the dual-modality format, and correlation with the model observers was also improved. Given the human-observer data from the SPECT study format, the Pearson correlation coefficients for the model observers were 0.58 (VS), -0.12 (CH), and -0.23 (CNPW). The respective coefficients based on the human-observer data from the dual-modality study were 0.72, 0.27, and -0.11. These results point towards the continued development of the VS observer for enhancing task equivalence in model-observer studies.

Entities:  

Keywords:  SPECT; dual-modality imaging; image quality assessment; lesion detection; model observers; visual search

Year:  2016        PMID: 27980345      PMCID: PMC5152772          DOI: 10.1109/TNS.2016.2542042

Source DB:  PubMed          Journal:  IEEE Trans Nucl Sci        ISSN: 0018-9499            Impact factor:   1.679


  21 in total

1.  Effect of filtering on the detection and localization of small Ga-67 lesions in thoracic single photon emission computed tomography images.

Authors:  R G Wells; P H Simkin; P F Judy; M A King; H Pretorius; H C Gifford
Journal:  Med Phys       Date:  1999-07       Impact factor: 4.071

2.  Evaluation of internal noise methods for Hotelling observer models.

Authors:  Yani Zhang; Binh T Pham; Miguel P Eckstein
Journal:  Med Phys       Date:  2007-08       Impact factor: 4.071

3.  Calculation and validation of the use of effective attenuation coefficient for attenuation correction in In-111 SPECT.

Authors:  Youngho Seo; Kenneth H Wong; Bruce H Hasegawa
Journal:  Med Phys       Date:  2005-12       Impact factor: 4.071

4.  The efficacy of diagnostic imaging.

Authors:  D G Fryback; J R Thornbury
Journal:  Med Decis Making       Date:  1991 Apr-Jun       Impact factor: 2.583

5.  The meaning and use of the area under a receiver operating characteristic (ROC) curve.

Authors:  J A Hanley; B J McNeil
Journal:  Radiology       Date:  1982-04       Impact factor: 11.105

6.  A random-sum Wilcoxon statistic and its application to analysis of ROC and LROC data.

Authors:  Liansheng Larry Tang; N Balakrishnan
Journal:  J Stat Plan Inference       Date:  2010-06-12       Impact factor: 1.111

7.  Progress in SPECT/CT imaging of prostate cancer.

Authors:  Youngho Seo; Benjamin L Franc; Randall A Hawkins; Kenneth H Wong; Bruce H Hasegawa
Journal:  Technol Cancer Res Treat       Date:  2006-08

8.  Towards Visual-Search Model Observers for Mass Detection in Breast Tomosynthesis.

Authors:  Beverly A Lau; Mini Das; Howard C Gifford
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2013-03-21

9.  Assessment of scatter compensation strategies for (67)Ga SPECT using numerical observers and human LROC studies.

Authors:  Troy H Farncombe; Howard C Gifford; Manoj V Narayanan; P Hendrik Pretorius; Eric C Frey; Michael A King
Journal:  J Nucl Med       Date:  2004-05       Impact factor: 10.057

10.  Efficient visual-search model observers for PET.

Authors:  H C Gifford
Journal:  Br J Radiol       Date:  2014-05-16       Impact factor: 3.039

View more
  3 in total

1.  Lack of agreement between radiologists: implications for image-based model observers.

Authors:  Juhun Lee; Robert M Nishikawa; Ingrid Reiser; Margarita L Zuley; John M Boone
Journal:  J Med Imaging (Bellingham)       Date:  2017-05-03

2.  DeepAMO: a multi-slice, multi-view anthropomorphic model observer for visual detection tasks performed on volume images.

Authors:  Ye Li; Junyu Chen; Justin L Brown; S Ted Treves; Xinhua Cao; Frederic H Fahey; George Sgouros; Wesley E Bolch; Eric C Frey
Journal:  J Med Imaging (Bellingham)       Date:  2021-01-28

Review 3.  Virtual clinical trials in medical imaging: a review.

Authors:  Ehsan Abadi; William P Segars; Benjamin M W Tsui; Paul E Kinahan; Nick Bottenus; Alejandro F Frangi; Andrew Maidment; Joseph Lo; Ehsan Samei
Journal:  J Med Imaging (Bellingham)       Date:  2020-04-11
  3 in total