The Role Of Contact Force In Atrial Fibrillation Ablation.

Hiroshi Nakagawa1, Warren M Jackman1.   

Abstract

During radiofrequency (RF) ablation, low electrode-tissue contact force (CF) is associated with ineffective RF lesion formation, whereas excessive CF may increase the risk of steam pop and perforation. Recently, ablation catheters using two technologies have been developed to measure real-time catheter-tissue CF. One catheter uses three optical fibers to measure microdeformation of a deformable body in the catheter tip. The other catheter uses a small spring connecting the ablation tip electrode to the catheter shaft with a magnetic transmitter and sensors to measure microdeflection of the spring. Pre-clinical experimental studies have shown that 1) at constant RF power and application time, RF lesion size significantly increases with increasing CF; 2) the incidence of steam pop and thrombus also increase with increasing CF; 3) modulating RF power based on CF (i.e, high RF power at low CF and lower RF power at high CF) results in a similar and predictable RF lesion size. In clinical studies in patients undergoing pulmonary vein (PV) isolation, CF during mapping in the left atrium and PVs showed a wide range of CF and transient high CF. The most common high CF site was located at the anterior/rightward left atrial roof, directly beneath the ascending aorta. There was a poor relationship between CF and previously used surrogate parameters for CF (unipolar or bipolar atrial potential amplitude and impedance). Patients who underwent PV isolation with an average CF of <10 g experienced higher AF recurrence, whereas patients with ablation using an average CF of > 20g had lower AF recurrence. AF recurred within 12 months in 6 of 8 patients (75%) who had a mean Force-Time Integral (FTI, area under the curve for contact force vs. time) < 500 gs. In contrast, AF recurred in only 4 of 13 patients (21%) with ablation using a mean FTI >1000 gs. In another study, controlling RF power based on CF prevented steam pop and impedance rise without loss of lesion effectiveness. These studies confirm that CF is a major determinant of RF lesion size and future systems combining CF, RF power and application time may provide real-time assessment of lesion formation.

Entities:  

Keywords:  Atrial Fibrillation; Catheter Ablation; Contact Force; Electrophysiology; Mapping; Radiofrequency

Year:  2014        PMID: 27957075      PMCID: PMC5135144          DOI: 10.4022/jafib.1027

Source DB:  PubMed          Journal:  J Atr Fibrillation        ISSN: 1941-6911


  19 in total

1.  Electrode impedance: an indicator of electrode-tissue contact and lesion dimensions during linear ablation.

Authors:  X Zheng; G P Walcott; J A Hall; D L Rollins; W M Smith; G N Kay; R E Ideker
Journal:  J Interv Card Electrophysiol       Date:  2000-12       Impact factor: 1.900

2.  Area under the real-time contact force curve (force-time integral) predicts radiofrequency lesion size in an in vitro contractile model.

Authors:  Dipen C Shah; Hendrik Lambert; Hiroshi Nakagawa; Arne Langenkamp; Nicolas Aeby; Giovanni Leo
Journal:  J Cardiovasc Electrophysiol       Date:  2010-09

3.  Initial impedance decrease as an indicator of good catheter contact: insights from radiofrequency ablation with force sensing catheters.

Authors:  Tobias Reichlin; Sven Knecht; Christopher Lane; Michael Kühne; Eyal Nof; Nagesh Chopra; Thomas M Tadros; Vivek Y Reddy; Beat Schaer; Roy M John; Stefan Osswald; William G Stevenson; Christian Sticherling; Gregory F Michaud
Journal:  Heart Rhythm       Date:  2013-10-28       Impact factor: 6.343

4.  The effects of electrode-tissue contact on radiofrequency lesion generation.

Authors:  B Avitall; K Mughal; J Hare; R Helms; D Krum
Journal:  Pacing Clin Electrophysiol       Date:  1997-12       Impact factor: 1.976

5.  Electrical reconnection after pulmonary vein isolation is contingent on contact force during initial treatment: results from the EFFICAS I study.

Authors:  Petr Neuzil; Vivek Y Reddy; Josef Kautzner; Jan Petru; Dan Wichterle; Dipen Shah; Hendrik Lambert; Aude Yulzari; Erik Wissner; Karl-Heinz Kuck
Journal:  Circ Arrhythm Electrophysiol       Date:  2013-03-20

6.  Relation between impedance and endocardial contact during radiofrequency catheter ablation.

Authors:  S A Strickberger; V R Vorperian; K C Man; B D Williamson; S J Kalbfleisch; C Hasse; F Morady; J J Langberg
Journal:  Am Heart J       Date:  1994-08       Impact factor: 4.749

7.  The relationship between contact force and clinical outcome during radiofrequency catheter ablation of atrial fibrillation in the TOCCATA study.

Authors:  Vivek Y Reddy; Dipen Shah; Josef Kautzner; Boris Schmidt; Nadir Saoudi; Claudia Herrera; Pierre Jaïs; Gerhard Hindricks; Petr Peichl; Aude Yulzari; Hendrik Lambert; Petr Neuzil; Andrea Natale; Karl-Heinz Kuck
Journal:  Heart Rhythm       Date:  2012-07-20       Impact factor: 6.343

8.  Catheter tip force required for mechanical perforation of porcine cardiac chambers.

Authors:  Dipen Shah; Hendrik Lambert; Arne Langenkamp; Yuri Vanenkov; Giovanni Leo; Pascale Gentil-Baron; Beat Walpoth
Journal:  Europace       Date:  2010-11-17       Impact factor: 5.214

9.  Importance of catheter contact force during irrigated radiofrequency ablation: evaluation in a porcine ex vivo model using a force-sensing catheter.

Authors:  Aravinda Thiagalingam; Andre D'Avila; Lori Foley; J Luis Guerrero; Hendrik Lambert; Giovanni Leo; Jeremy N Ruskin; Vivek Y Reddy
Journal:  J Cardiovasc Electrophysiol       Date:  2010-02-01

10.  Predictive value of impedance changes for real-time contact force measurements during catheter ablation of atrial arrhythmias in humans.

Authors:  Saurabh Kumar; H M Haqqani; Martin Chan; Justin Lee; Matias Yudi; Michael C G Wong; Joseph B Morton; Liang-Han Ling; Timothy Robinson; Patrick M Heck; Nicholas F Kelland; Karen Halloran; Steven J Spence; Peter M Kistler; Jonathan M Kalman
Journal:  Heart Rhythm       Date:  2013-03-22       Impact factor: 6.343

View more
  9 in total

1.  Earliest pulmonary vein potential-guided cryoballoon ablation for atrial fibrillation.

Authors:  Yoshiaki Mizutani; Masaaki Kanashiro; Yuichiro Makino; Akinori Satake; Wataru Suzuki; Masanari Kurobe; Kouji Mizutani; Hitoshi Ichimiya; Yasuhiro Uchida; Junji Watanabe; Satoshi Ichimiya; Yasuya Inden; Toyoaki Murohara
Journal:  Heart Vessels       Date:  2019-07-11       Impact factor: 2.037

2.  Efficacy of advanced pace-mapping technology for idiopathic premature ventricular complexes ablation.

Authors:  J Fedida; T Strisciuglio; M Sohal; M Wolf; K Van Beeumen; A Neyrinck; P Taghji; C Lepiece; A Almorad; Y Vandekerckhove; R Tavernier; M Duytschaever; S Knecht
Journal:  J Interv Card Electrophysiol       Date:  2018-02-24       Impact factor: 1.900

3.  An ablation index operator-independent approach to improve efficacy in atrial fibrillation ablation at 24-month follow-up: a single center experience.

Authors:  Michela Casella; Antonio Dello Russo; Stefania Riva; Valentina Catto; Gabriele Negro; Rita Sicuso; Selene Cellucci; Alessio Gasperetti; Martina Zucchetti; Valentina Ribatti; Viviana Biagioli; Gaetano Fassini; Luigi Di Biase; Andrea Natale; Claudio Tondo
Journal:  J Interv Card Electrophysiol       Date:  2019-07-16       Impact factor: 1.900

4.  Pulmonary vein isolation using Ablation Index improves outcome in patients with atrial fibrillation.

Authors:  Jonas Elmer Pedersen; Kim Frost Lauritsen; Jens Brock Johansen; Niels Christian Foldager Sandgaard; Jorgen Dalhoj; Stig Djurhuus; Jacob Pontoppidan
Journal:  J Atr Fibrillation       Date:  2020-04-30

5.  Predictors of Lesions Contiguity and Transmurality in Canine Ventricular Models After Catheter Ablation.

Authors:  Abdel Hadi El Hajjar; Mario Mekhael; Chao Huang; Charbel Noujaim; Yichi Zhang; Eugene Kholmovski; Tarek Ayoub; Chan Ho Lim; Nassir Marrouche
Journal:  Front Cardiovasc Med       Date:  2022-06-23

6.  Oesophageal Injury During AF Ablation: Techniques for Prevention.

Authors:  Jorge Romero; Ricardo Avendano; Michael Grushko; Juan Carlos Diaz; Xianfeng Du; Carola Gianni; Andrea Natale; Luigi Di Biase
Journal:  Arrhythm Electrophysiol Rev       Date:  2018-03

7.  Tailored Target Ablation Index Guided Pulmonary Vein Isolation in Treating Paroxysmal Atrial Fibrillation: A Single Center Randomized Study in Asian Population (AI-Asian-I).

Authors:  Qingsong Xiong; Jia Liao; Weijie Chen; Peilin Xiao; Huaan Du; Qushuai He; Yuehui Yin; Zhiyu Ling; Shaojie Chen
Journal:  Front Cardiovasc Med       Date:  2022-07-07

8.  A computational comparison of radiofrequency and pulsed field ablation in terms of lesion morphology in the cardiac chamber.

Authors:  Mario Gómez-Barea; Tomás García-Sánchez; Antoni Ivorra
Journal:  Sci Rep       Date:  2022-09-27       Impact factor: 4.996

9.  Simple Maneuver for Estimating the Depth of the Focal Activation Source in Myocardium.

Authors:  Seil Oh
Journal:  Korean Circ J       Date:  2020-01-15       Impact factor: 3.243

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.