Analysis of intensity variability in multislice and cone beam computed tomography.

Olivia Nackaerts1, Frederik Maes, Hua Yan, Paulo Couto Souza, Ruben Pauwels, Reinhilde Jacobs.   

Abstract

OBJECTIVES: The aim of this study was to evaluate the variability of intensity values in cone beam computed tomography (CBCT) imaging compared with multislice computed tomography Hounsfield units (MSCT HU) in order to assess the reliability of density assessments using CBCT images.
MATERIAL AND METHODS: A quality control phantom was scanned with an MSCT scanner and five CBCT scanners. In one CBCT scanner, the phantom was scanned repeatedly in the same and in different positions. Images were analyzed using registration to a mathematical model. MSCT images were used as a reference.
RESULTS: Density profiles of MSCT showed stable HU values, whereas in CBCT imaging the intensity values were variable over the profile. Repositioning of the phantom resulted in large fluctuations in intensity values.
CONCLUSIONS: The use of intensity values in CBCT images is not reliable, because the values are influenced by device, imaging parameters and positioning.
© 2011 John Wiley & Sons A/S.

Mesh:

Year:  2011        PMID: 21244502     DOI: 10.1111/j.1600-0501.2010.02076.x

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  48 in total

1.  Ultralow dose dentomaxillofacial CT imaging and iterative reconstruction techniques: variability of Hounsfield units and contrast-to-noise ratio.

Authors:  Gerlig Widmann; Alexander Bischel; Andreas Stratis; Apoorv Kakar; Hilde Bosmans; Reinhilde Jacobs; Eva-Maria Gassner; Wolfgang Puelacher; Ruben Pauwels
Journal:  Br J Radiol       Date:  2016-01-21       Impact factor: 3.039

2.  Image quality assessment of three cone beam CT machines using the SEDENTEXCT CT phantom.

Authors:  J Bamba; K Araki; A Endo; T Okano
Journal:  Dentomaxillofac Radiol       Date:  2013       Impact factor: 2.419

3.  Influence of object location in different FOVs on trabecular bone microstructure measurements of human mandible: a cone beam CT study.

Authors:  N Ibrahim; A Parsa; B Hassan; P van der Stelt; I H A Aartman; P Nambiar
Journal:  Dentomaxillofac Radiol       Date:  2013-11-21       Impact factor: 2.419

Review 4.  CBCT-based bone quality assessment: are Hounsfield units applicable?

Authors:  R Pauwels; R Jacobs; S R Singer; M Mupparapu
Journal:  Dentomaxillofac Radiol       Date:  2015       Impact factor: 2.419

5.  Evaluation of a metal artefact reduction tool on different positions of a metal object in the FOV.

Authors:  Polyane M Queiroz; Gustavo M Santaella; Thais D J da Paz; Deborah Q Freitas
Journal:  Dentomaxillofac Radiol       Date:  2017-02-17       Impact factor: 2.419

Review 6.  Quality assurance phantoms for cone beam computed tomography: a systematic literature review.

Authors:  Marcus V L de Oliveira; Ann Wenzel; Paulo S F Campos; Rubens Spin-Neto
Journal:  Dentomaxillofac Radiol       Date:  2017-02-17       Impact factor: 2.419

7.  Assessment of cortical bone microdamage following insertion of microimplants using optical coherence tomography: a preliminary study.

Authors:  Hemanth Tumkur Lakshmikantha; Naresh Kumar Ravichandran; Mansik Jeon; Jeehyun Kim; Hyo-Sang Park
Journal:  J Zhejiang Univ Sci B       Date:  2018 Nov.       Impact factor: 3.066

8.  Computer-based automatic classification of trabecular bone pattern can assist radiographic bone quality assessment at dental implant site.

Authors:  Laura Ferreira Pinheiro Nicolielo; Jeroen Van Dessel; G Harry van Lenthe; Ivo Lambrichts; Reinhilde Jacobs
Journal:  Br J Radiol       Date:  2018-09-17       Impact factor: 3.039

9.  Influence of cone beam CT scanning parameters on grey value measurements at an implant site.

Authors:  A Parsa; N Ibrahim; B Hassan; A Motroni; P van der Stelt; D Wismeijer
Journal:  Dentomaxillofac Radiol       Date:  2012-08-29       Impact factor: 2.419

10.  Variability of dental cone beam CT grey values for density estimations.

Authors:  R Pauwels; O Nackaerts; N Bellaiche; H Stamatakis; K Tsiklakis; A Walker; H Bosmans; R Bogaerts; R Jacobs; K Horner
Journal:  Br J Radiol       Date:  2013-01       Impact factor: 3.039

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.