Literature DB >> 27685106

Is the dose distribution distorted in IMRT and RapidArc treatment when patient plans are swapped across beam-matched machines?

Chockkalingam Krishnappan1, Chandrasekaran Anu Radha, Vendhan Subramani, Madhan Kumar Gunasekaran.   

Abstract

The purpose of this study is to evaluate the degree of dose distribution distortion in advanced treatments like IMRT and RapidArc when patient plans are swapped across dosimetrically equivalent so-called "beam-matched" machines. For this purpose the entire work is divided into two stages. At forefront stage all basic beam properties of 6 MV X-rays like PDD, profiles, output factors, TPR20/10 and MLC transmission of two beam-matched machines - Varian Clinac iX and Varian 600 C/D Unique - are compared and evaluated for differences. At second stage 40 IMRT and RapidArc patient plans from the pool of head and neck (H&N) and pelvis sites are selected for the study. The plans are swapped across the machines for dose recalculation and the DVHs of target and critical organs are evaluated for dose differences. Following this, the accuracy of the beam-matching at the TPS level for treatments like IMRT and RapidArc are compared. On PDD, profile (central 80%) and output factor comparison between the two machines, a maxi-mum percentage disagreement value of -2.39%, -2.0% and -2.78%, respectively, has been observed. The maximum dose difference observed at volumes in IMRT and RapidArc treatments for H&N dose prescription of 69.3 Gy/33 fractions is 0.88 Gy and 0.82 Gy, respectively. Similarly, for pelvis, with a dose prescription of 50 Gy/25 fractions, a maximum dose difference of 0.55 Gy and 0.53 Gy is observed at volumes in IMRT and RapidArc treatments, respectively. Overall results of the swapped plans between two machines' 6 MV X-rays are well within the limits of accepted clinical tolerance.
© 2016 The Authors.

Entities:  

Mesh:

Year:  2016        PMID: 27685106      PMCID: PMC5874098          DOI: 10.1120/jacmp.v17i5.6104

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


I. INTRODUCTION

Beam matching is the concept of altering or tuning the beams of teletherapy machines so that they match with one another. Beam commissioning in conventional Co‐60 machines uses a single set of universal beam profile charts supplied by the manufacturer. These charts resemble the Co‐60 beam of all machines and thus the concept of beam matching was irrelevant. However, with time Co‐60 units have almost been completely replaced by medical linear accelerators. Computer‐controlled linear accelerators (linacs) that generate high‐energy X‐rays havetheir own specific and unique beam characteristics. Hence, each beam from a linac has to be commissioned individually before it is used clinically for patients. Steep increase in cancer incidence has made radiotherapy centers think about having a second linac in their department. Often, a second linac is considered as standby machine if one of them is inevitably down. So, the two linacs are made dosimetrically equivalent through so‐called beam‐matching. The term “beam‐matched linacs” ensures that the X‐ray beams of matched linacs exhibit almost similar dosimetric characteristics. One of the clear advantages of beam‐matching linear accelerators is the improved efficiency and flexibility in patient treatment for institutions with two or more linear accelerators. Effects of beam‐matching results and beam data reproducibility for various accelerators have previously been analyzed and presented by several authors. , , , , , The beam‐matching criteria basically depend on depth dose/ionization curves, as well as beam profiles measured in both inline and crossline directions under vendor‐defined prescribed geometry. Even though during the accelerators' customer acceptance procedure all the vendor‐defined criteria are duly fulfilled, they are inadequate for beam matching. The vendor‐defined criteria consider only some points on profiles and depth dose curves instead of the full portion of the same. Apart from this they do not include the output factors and therefore run a risk of good agreement solely due to normalization. All vendor‐defined measurements are carried only for open static fields with no inclusion of multileaf collimators (MLC), whereas MLC is an integral part of modern‐day radiotherapy. MLC effects on beam‐matched linacs can bring about a severe alteration in patient dose distributions, especially in inverse planning–based advanced treatments like intensity‐modulated radiation therapy (IMRT) and RapidArc (Varian Medical Systems, Palo Alto, CA). Even though the beam matching is done at the factory, its accuracy has to be ensured in clinics before shifting the patients across the machines if the same dose distributions are to be achieved for the patient. Although several studies have been carried out on beam data comparisons of beam‐matched linacs, there are no data available on the effects of beam‐matching at the patients' levels, especially in high‐end treatments like IMRT and RapidArc. The aim of this study is to evaluate the accuracy of beam matching by overcoming the shortfalls of vendor‐defined criteria and to study the effects of beam matching in advanced treatments like IMRT and RapidArc.

II. MATERIALS AND METHODS

Recently at our center, we commissioned a Varian 600 C/D Unique linac. It is a magnetron‐based low‐energy linac capable of generating only one energy of X‐ray photons (6 MV). It has electron gun, standing waveguide, tungsten target and flattening filter all together as one complete central beam‐line unit. There are no bending magnets as the entire waveguide unit is mounted vertically above the ion chamber. This linac is equipped with the 120 Millennium Multi‐Leaf Collimator as tertiary MLC, in addition to beam‐limiting jaws (upper and lower). This linac has been beam‐matched with the already existing Varian Clinac iX Trilogy at our department. Other than collimating diaphragms and MLCs, the linac has a completely different head and waveguide design. Both the linacs are capable of delivering high dose rates (up to 600 MU/min). The accuracy of beam matching between the two linacs is evaluated in two segments.

A. At machine commissioning

For both machines' 6 MV X‐rays, basic beam data like percentage depth doses (PDDs) and cross‐beam profiles are measured in a radiation field analyzer (Blue Phantom, IBA Dosimetry GmbH, Schwarzenbruck, Germany) using CC13 chambers (IBA Dosimetry) for open fields. As per TRS398 protocol the beam quality index (TPR 20/10) is measured for 6 MV X‐rays of both the machines. Open‐beam output factors, along with their individual scatter components (phantom and head) are also measured for both the machines. FC65 chamber (IBA Dosimetry) and 1D Phantom (IBA dosimetry) are used to measure TPR 20/10 and output factors. Dosimetric leaf gap (DLG) and transmission of multi‐leaf collimators (MLCs) of both the machines are measured using CC13 chamber placed in an RW3 phantom (IBA Dosimetry). A comparison is made between the PDDs of 6 MV X‐rays from both the machines at all depths for different field sizes ranging from to . Before analysis, the PDDs of the machines at each field size were normalized to 100% at of the Clinac iX. Disagreement in PDD (%) [(PDD (Clinac iX) – PDD (Unique))] is then calculated at all depths on that particular field size by simple subtraction of PDD values of both the machines. Beam profiles in the cross plane of the both machines for different field sizes (, and ) at five different depths (, 5, 10, 20, and 30 cm) are compared. All profiles at all depths are normalized to 100% at central axis. Profile disagreement analysis between the machines is made by calculating the difference in profile values [(Clinac iX –Unique)] at different regions (central 80% and penumbra). The difference between the TPR 20/10 values of 6 MV X‐rays of both the machines are calculated. With the Clinac iX machine as baseline reference, the deviation in overall output factors, as well as phantom and head scatter factors, for the Unique are calculated and compared at all field sizes from to Percentage difference in output factors and TPR20/10 is calculated as . The difference in MLC transmission and DLG values (Clinac iX –Unique) for both the machines is calculated.

B. At TPS commissioning

Beam data and all other dosimetric properties of both the machines are fed into the Eclipse treatment planning system (TPS) (Varian) version 11.0 in accordance with the vendor's specification and recommendations. Once the machine type is selected in Eclipse, it automatically takes all required basic machine characteristics from the available machine library data for beam modelling. Only measured PDDs, profiles, output factors, and absolute dose calibration factor are fed for the generation of beam model. Measured DLG and MLC transmission values are added as add‐on dosimetric parameter to the machine. These values can be further tweaked if necessary to improve the test results of a series of prerequisite QA performance tests for TPS and machines. In our context only the accuracy of the dose‐calculation method of TPS is evaluated individually for both the machines as part of the TPS QA. , Accordingly, chamber measurements (CC01 chamber) at different specified locations in a water phantom for a set of field sizes (, and ) are compared with the TPS‐calculated value in two different scenarios (jaws only and MLC only). In addition, IMRT commissioning tests are performed in the TPS for both the machines separately. A specific set of test plans as specified in TG119 are then delivered at the machines. These test shapes and plans are representative of common clinical treatments and are used to test the overall accuracy of our IMRT commissioned system. The results are evaluated using both chamber measurements and 2D planar dosimetry. The portal imager is used for measuring 2D planar image and the results are analyzed using Portal Dosimetry (Varian). Gamma criteria used for 2D planar analysis are 3 mm and 3% as distance‐to‐agreement (DTA) and dose difference (DD), respectively. The tolerance limit is area gamma is greater than 95%. It is vital to evaluate the accuracy of beam matching between the two machines, especially in advanced patient treatment techniques like IMRT and RapidArc, beforepronouncing the machines “dosimetrically equivalent.” For this purpose 10 patient plans belonging to each technique (IMRT and RapidArc) and treatment sites (H&N and pelvis) for a total of 40 patient plans which were earlier treated on the Clinac iX machine at our hospital are chosen. The treatment plans are migrated to the Varian 600 C/D (Unique) machine in the Eclipse TPS and the dose recalculated for the same field fluence, field size and MUs of the Clinac iX machine using the option of “calculate with fixed MU.” In both cases, analytical anisotropic algorithm (AAA) is used for dose calculation with a grid resolution kept at 2.5 mm. The plans are compared to find the dosimetric difference between the two machines. The dose‐volume histogram (DVH) of target volumes and critical organs are used for evaluation.

B.1 Head and neck (H&N)

The simultaneous integrated boost (SIB) mode of dose prescription is usually practiced in H&N tumors. Accordingly, the dose prescribed to PTV I (high‐risk), PTV II (intermediate‐risk), and PTV III (low‐risk) volumes are 69.3 Gy, 66 Gy, and 59.4 Gy over 33 fractions, respectively. Spinal cord and parotids are the major organs at risk (OARs). Clinically important dose parameters and of dose‐volume histograms (DVHs) are compared and average variation is calculated for the PTVs and parotids between the two machine plans. , For serial organs like spinal cord and parameters are considered for comparison. Average variation of each volume at mentioned levels is analyzed.

B.1.1 Intensity‐modulated radiation therapy (IMRT)

Normally seven to nine fields of 6 MV X‐rays are used in IMRT planning. For inverse optimization, Dose Volume Optimizer (DVO) version 11.0.31 is employed in the Eclipse TPS. The average volume of PTV I, PTV II, PTV III, spinal cord, and left and right parotid are cc, , and , respectively. The average monitor units for the plans are .

B.1.2 RapidArc

Planning of RapidArc usually involves two full arcs of 6 MV X‐rays. RapidArc plans are generated by the Progressive Resolution Optimizer (PRO) III version 11.0.31 working principle based on direct‐aperture optimization. The average volume of PTV I, PTV II, PTV III, spinal cord, and left and right parotid are , and , respectively. The average monitor units for the plans are .

B.2 Pelvis

In case of pelvic tumors, a dose of 50 Gy in 25 fractions is usually prescribed to the PTV. Normally, organs like rectum, bladder, and femoral heads are the OARs involved in the treatment of pelvic tumors. Dose of , and parameter values are extracted from the PTV and OARs DVH, for the comparison between the two machines. , Average variation of each volume at mentioned levels is analyzed.

B.2.1 Intensity‐modulated radiation therapy (IMRT)

IMRT planning at pelvis involves the use of seven to nine fields of 6 MV X‐rays similar to the planning of H&N IMRT. The average volume of PTV, rectum, bladder, femoral head, and left and right parotid are and , respectively. The average monitor units for the plans are .

B.2.2 RapidArc

Two conventional full arcs of 6 MV X‐rays are used for planning. The average volume of PTV, rectum, bladder, femoral head, and left and right parotid are , and , respectively. The average monitor units for the plans are .

III. RESULTS AND DISCUSSION

The PDD values of both the machines at different depths for different fields are shown in Table 1. For field sizes and above, at all depths (excluding the buildup region), the absolute percent difference of PDD comparisons made between the two machines are well within 1%. In the same region, the degree of absolute percent difference increases for field sizes less than and reaches a maximum deviation of 2.05% at the depth of 16.5 cm for field size. On comparing the buildup region doses between the two machines, the absolute percent difference varies considerably for all field sizes. A maximum of deviation is seen at the field sizes of and . Excluding the buildup region, PDD comparison between the two machine fetches an overall percent difference value less than 2.05%. PDD comparison between the two machines at field sizes and are shown in 1, 2, respectively.
Table 1

PDD at different depths and field sizes for Clinac iX and Unique.

Field Size
Depth Machine 4×4cm2 10×10cm2 20×20cm2 40×40cm2
Dmax (cm)Clinac iX1.661.591.461.34
Unique1.501.501.351.23
PDD5% Clinac iX84.486.687.888.5
Unique83.486.587.687.9
PDD10% Clinac iX62.466.970.172.2
Unique61.266.970.071.9
PDD20% Clinac iX33.838.542.946.1
Unique32.338.442.846.2
Figure 1

PDD comparison of Clinac iX and Unique machines for a field size of .

Figure 2

PDD comparison of Clinac iX and Unique machines for a field size of .

PDD at different depths and field sizes for Clinac iX and Unique. PDD comparison of Clinac iX and Unique machines for a field size of . PDD comparison of Clinac iX and Unique machines for a field size of . Profile penumbra and field width measured at different depths are listed in Table 2 for both the machines. Profile disagreement within central 80% of the plane for all field sizes at all five depths between the two machines varies from to 1.8%. The maximum variation −2% is found at the depth of 30 cm for the field sizes of . The average absolute percent disagreement in the central 80% is . The absolute percent difference in penumbra region for all the field sizes at all five depths between two machines varies from 4.4% to 20.7%. The maximum variation of 20.7% is found at the depth of 30 cm for the field sizes of . The average disagreement in penumbra region for all the field sizes at all the depths is .
Table 2

Profile comparison for Clinac iX and Unique at different field sizes.

3×3cm2 10×10cm2 20×20cm2 30×30cm2
Depth Machine Penumbra (mm) Field Width (cm) Penumbra (mm) Field Width (cm) Penumbra (mm) Field Width (cm) Penumbra (mm) Field Width (cm)
Dmax Clinac iX4.93.065.310.255.520.485.530.68
Unique4.42.985.410.145.620.345.630.54
10 cmClinac iX5.43.336.711.117.922.158.733.17
Unique4.93.276.811.007.822.038.733.04
20 cmClinac iX5.73.648.212.1313.024.1617.336.14
Unique5.33.598.512.0112.624.0416.436.04
TPR 20/10 values of 6 MV X‐rays are 0.667 and 0.670 for Clinac iX and Unique machines, respectively. Variation of TPR 20/10 value between the two machines is 0.45%. The output factors and their individual component values are tabulated in Table 3 for both the machines at different field sizes. For overall output factors, percentage variation between the two machines is in the range of 1.66% to as the field size increases from to . The phantom scatter factor remains almost constant with a maximum difference of 0.57%. The difference in head scatter factor of the two machines varies from 1.55% to as the field size increases from to . Since both the machines have different head characteristics and wave‐guide design the variations observed in output factors, especially the head scatter factors are quite understandable. The MLC transmission measured is 1.47% and 1.35% for Clinac iX and Unique machine respectively. The measured DLG value is same for both the machines and it is 2 mm. The percentage difference in MLC transmission between the two machines is 0.12%.
Table 3

Comparison of output factor, head scatter and phantom scatter factor for Clinac iX and Unique machine at different field sizes.

Field Size
Machine 3×3cm2 10×10cm2 20×20cm2 40×40cm2
Output factorClinac iX0.86661.0001.06511.1263
Unique0.88101.0001.05671.0950
Head scatter factor Clinac iX0.91341.0001.03201.0606
Unique0.92751.0001.02351.0347
Phantom scatter factorClinac iX0.94881.0001.03201.0620
Unique0.94991.0001.03241.0583
Profile comparison for Clinac iX and Unique at different field sizes. Comparison of output factor, head scatter and phantom scatter factor for Clinac iX and Unique machine at different field sizes. The results of the TPS QA tests done as per TG53 for both the machines are shown in 4, 5. The results are within the acceptable criteria for both machines. In IMRT commissioning tests show both machines can easily achieve the planning constraints as stated in TG119. From the results tabulated in 6, 7 it is seen that the machines can successfully deliver the plans within the specified tolerance criteria for all the shapes.
Table 4

Point‐dose results in different regions of a phantom for both machine at different field sizes (jaws only), per TG53.

Point Dose (cGy)
FS 3×3cm2 FS 10×10cm2 FS 25×25cm2
Region of Measurement Machine Measured Planned % Variation Measured Planned % Variation Measured Planned % Variation
InnerClinac iX82.482.7 0.36 95.695.30.31104.6103.60.97
Unique83.783.60.1295.495.00.42103.5102.01.47
OuterClinac iX1.20.80.484.13.90.217.67.9 0.29
Unique1.40.80.723.83.50.327.77.00.69
BuildupClinac iX95.596.8 1.57 109.1107.12.10115.8114.90.87
Unique99.299.20.00109.2107.81.47116.8114.12.65
PenumbraClinac iX36.142.3 7.50 44.150.1 6.30 50.855.5 4.54
Unique34.442.7 9.93 37.849.9 12.74 48.054.5 6.38
Table 5

Point‐dose results in different regions of a phantom for both machine at different field sizes (MLC only) as per TG53.

Point Dose (cGy)
Region of Measurement FS 3×3cm2 FS 10×10cm2 FS 25×25cm2
Machine Measure d Planned % Variation Measured Planned % Variation Measured Planned % Variation
InnerClinac iX82.882.70.1295.795.30.42103.9103.60.29
Unique83.083.5 0.60 94.995.0 0.11 103.0102.00.98
OuterClinac iX1.50.80.854.23.90.318.07.90.10
Unique1.00.70.363.93.50.427.27.00.20
Build UpClinac iX Unique98.496.81.93107.6107.10.52116.5114.91.54
Unique99.499.10.36109.2107.81.47117.0114.12.84
PenumbraClinac iX Unique37.242.3 6.17 41.550.1 9.0 52.655.5 2.80
Unique34.042.6 10.30 37.249.9 13.37 47.854.5 6.57
Table 6

Gamma analysis and point‐dose results of IMRT plans of TG119 test cases in Clinac iX and Unique machines with confidence limit.

Clinac iX Unique
Point Dose (Gy) Planar Dosimetry Point Dose (Gy) Planar Dosimetry
Test Shape Location Measured Planned Deviation Gamma <1(%) Measured Planned Deviation Gamma <1(%)
ProstateIsocenter 2.5 cm posterior207.7204.00.018599.89205.5204.60.004599.83
127.2128.1 0.0045 127.0128.4 0.0070
Head & NeckIsocenter 4.0 cm posterior209.69205.90.019098.98203.5204.6 0.0055 96.97
115.5118.7 0.0160 110.9118.0 0.0355
C‐Shape(E)Isocenter 2.5 cm anterior77.9672.50.027398.2371.369.90.007097.02
221.862180.0193213.2216.6 0.0170
C‐Shape(H)Isocenter 2.5 cm anterior48.2746.10.010896.7442.743.4 0.0035 96.80
221.56218.70.0143216218.5 0.0125
Multi TargetIsocenter 4.0 cm superior214.73211.80.014699.36211.2211.7 0.0025 98.87
136.5135.50.0050136.1136.4 0.0015
4.0 cm inferior65.965.50.002064.265.6 0.0070
Confidence LimitsHigh dose region0.0223.760.0234.79
Low dose region0.0333.760.036
Table 7

Gamma analysis and point‐dose results of VMAT plans of TG119 test cases in Clinac iX and Unique machines with confidence limit.

Clinac iX Unique
Point Dose (Gy) Planar Dosimetry Point Dose (Gy) Planar Dosimetry
Test Shape Location Measured Planned Deviation Gamma <1(%) Measured Planned Deviation Gamma <1(%)
ProstateIsocenter 2.5 cm posterior200.7200.20.002598.73199.8199.10.003599.40
134.1133.60.0025133.0132.80.0010
Head & NeckIsocenter 4.0 cm posterior200.2201.1 0.0045 96.12198.3199.5 0.0060 97.35
139.7138.30.0700136.6136.9 0.0015
C‐Shape(E)Isocenter 2.5 cm anterior60.760.20.002596.2756.957.7 0.0030 96.70
208.3209.1 0.0040 203.8206.9 0.0155
C‐Shape(H)Isocenter 2.5 cm anterior43.742.60.005594.3939.139.9 0.0040 97.61
210.8209.90.0045204.1207.9 0.0190
Multi TargetIsocenter 4.0 cm superior 4.0 cm inferior210.1209.40.003598.84209.7208.00.008596.20
112.81120.0040110.6109.50.0055
60.861.8 0.0050 58.859.4 0.0030
Confidence LimitsHigh dose region0.0096.850.0294.94
Low dose region0.0110.008
Point‐dose results in different regions of a phantom for both machine at different field sizes (jaws only), per TG53. Point‐dose results in different regions of a phantom for both machine at different field sizes (MLC only) as per TG53. Gamma analysis and point‐dose results of IMRT plans of TG119 test cases in Clinac iX and Unique machines with confidence limit. Gamma analysis and point‐dose results of VMAT plans of TG119 test cases in Clinac iX and Unique machines with confidence limit.

B.1 H&N

The average dose for different volumes of all organs for both machines for the same MU and fluence are tabulated in the Table 8, along with the dose difference between the two machines. With respect to PTV, at all dose levels, the average dose difference varies from 0.04 Gy to 0.40 Gy. The average dose difference for spinal cord at different dose levels between the two machines varies from 0.51 Gy to 0.61 Gy. A variation of 0.02 Gy to 0.87 Gy is observed in the dose parameters for both the parotids. The dose variation is found to increase constantly across the parotid region as we move away from the PTVs. A maximum variation of 0.87 Gy is observed between the two machines at level for the right parotid.
Table 8

Average dose obtained from TPS for the same fluence and MU for the case of IMRT H&N (in Gy).

D2 D5 D10 D50 D90 D98 Dmean
PTVIClinac iX 71.2±0.68 71.1±0.59 70.8±0.56 69.8±0.58 68.5±0.57 67.5±0.72 69.7±0.54
Unique 71.0±0.65 70.7±0.61 70.4±0.58 69.4±0.61 68.2±0.55 67.3±0.64 69.3±0.57
Difference 0.23±0.23 0.40±0.16 0.40±0.16 0.38±0.14 0.35±0.15 0.18±0.16 0.38±0.14
PTVIIClinac iX 70.5±0.81 70.2±0.80 69.6±0.87 66.9±1.01 64.8±0.99 62.8±0.87 67.0±0.87
Unique 70.3±0.77 69.8±0.81 69.2±0.88 66.5±1.02 64.5±1.00 62.7±0.84 66.7±0.90
Difference 0.21±0.19 0.38±0.13 0.39±0.12 0.37±0.07 0.27±0.08 0.07±0.08 0.33±0.07
PTVIIIClinac iX 63.7±1.16 62.3±0.96 61.1±0.90 59.2±0.65 57.8±0.85 56.1±1.02 59.4±0.67
Unique 63.6±1.09 62.1±0.98 60.9±0.93 59.1±0.70 57.7±0.92 56.0±1.10 59.3±0.71
Difference 0.04±0.12 0.20±0.1 0.18±0.10 0.12±0.11 0.05±0.11 0.07±0.23 0.12±0.11
SpinalClinac iX 39.7±1.21 35.1±1.47 41.9±1.48
Unique 39.1±1.18 34.5±1.53 41.3±1.47
Difference 0.50±0.15 0.60±0.13 0.61±0.23
Parotid LClinac iX 65.1±2.89 62.9±3.90 59.1±6.26 24.7±10.7 11.2±2.75 9.6±1.92 30.5±6.83
Unique 65.0±2.87 62.6±3.89 58.8±6.34 23.9±10.8 10.3±2.75 8.9±1.91 29.9±6.85
Difference 0.13±0.12 0.28±0.10 0.33±0.14 0.78±0.18 0.86±0.10 0.71±0.31 0.60±0.20
Parotid RClinac iX 64.0±2.47 61.3±2.86 57.5±3.61 23.3±5.00 11.3±1.96 9.7±1.64 29.4±3.12
Unique 64.0±2.46 61.1±2.85 57.2±3.60 22.5±5.04 10.4±1.92 8.9±1.56 28.7±3.13
Difference 0.02±0.15 0.21±0.08 0.26±0.09 0.81±0.09 0.87±0.09 0.82±0.10 0.68±0.05
Average dose obtained from TPS for the same fluence and MU for the case of IMRT H&N (in Gy). In the H&N RapidArc plans, the average dose for different volumes of all organs for both machines for the same MU and fluence are shown in the Table 9, along with the difference in dose between the two machines.
Table 9

Average dose obtained from TPS for the same fluence and MU for the case of Rapid Arc H&N (in Gy).

D2 D5 D10 D50 D90 D98 Dmean
PTVIClinac iX 71.8±0.53 71.5±0.51 71.2±0.49 70.1±0.41 68.8±0.45 67.9±0.52 70.1±0.42
Unique 71.3±0.52 71.0±0.50 70.6±0.49 69.5±0.40 68.0±0.48 67.1±0.54 69.4±0.42
Difference 0.55±0.15 0.56±0.15 0.60±0.15 0.67±0.12 0.78±0.15 0.82±0.19 0.68±0.13
PTVIIClinac iX 71.1±0.52 70.5±0.51 69.9±0.50 67.5±0.56 65.2±0.70 62.9±0.79 67.5±0.54
Unique 70.5±0.51 69.9±0.50 69.3±0.49 67.0±0.54 64.7±0.66 62.5±0.74 67.0±0.52
Difference 0.57±0.13 0.59±0.11 0.58±.09 0.51±0.07 0.51±0.11 0.43±0.10 0.53±0.08
PTVIIIClinac iX 64.5±1.48 62.8±0.94 61.6±0.61 60.1±0.51 58.5±0.57 56.9±0.72 60.1±0.51
Unique 64.1±1.48 62.5±0.98 61.4±0.66 59.8±0.53 58.2±0.55 56.7±0.69 59.9±0.53
Difference 0.32±0.12 0.31±0.13 0.25±0.10 0.26±0.07 0.27±0.08 0.23±0.07 0.26±0.08
SpinalClinac iX 37.5±1.33 33.6±1.40 38.7±3.14
Unique 36.8±1.33 32.9±1.40 38.0±3.08
Difference 0.72±0.06 0.62±0.08 0.67±0.11
Parotid LClinac iX 62.5±4.17 58.7±5.89 52.9±8.49 21.9±4.34 12.1±2.64 10.1±2.16 27.6±4.14
Unique 62.1±4.21 58.2±5.98 52.4±8.61 21.2±4.35 11.4±2.64 9.5±2.18 27.0±4.16
Difference 0.39±0.11 0.45±0.16 0.51±0.15 0.76±0.09 0.65±0.08 0.58±0.10 0.66±0.08
Parotid RClinac iX 63.2±5.13 59.8±7.31 54.5±9.81 21.7±4.11 11.7±2.00 9.7±1.85 28.1±4.04
Unique 62.9±5.25 59.4±7.44 54.4±9.98 20.9±4.17 11.1±1.97 9.1±1.81 27.6±3.86
Difference 0.38±0.17 0.42±0.16 0.44±0.19 0.76±0.09 0.65±0.10 0.58±0.12 0.49±0.44
In all PTVs, the average dose difference at all levels of dose ; varies from 0.23 Gy to 0.82 Gy. The average dose difference between the two machines for spinal cord at different dose levels of , and varies from 0.62 Gy to 0.72 Gy. In both parotids, an average dose variation of 0.38 Gy to 0.76 Gy is observed between the two machines at all levels of dose , and . A maximum variation of 0.82 Gy is found between the two machines at level for PTV I. Average dose obtained from TPS for the same fluence and MU for the case of Rapid Arc H&N (in Gy). Though there is not much difference in the overall dose variation it is clearly observable that variation in RapidArc plans are more than in IMRT plans for all the volumes. Among the volumes, PTV III has the least variation; 0.04 Gy to 0.32 Gy in both techniques. This is due to the presence of less critical organs in its vicinity. Interestingly the parotids variation as observed in IMRT tends to increase from to whereas in RapidArc the maximum variation is observed at . The observed difference in parotid is due to the fact that the structure is overlapping with the PTV and the IMRT fields are delivered at fixed gantry angles. Variation in PTV I, PTV II, and spinal cord do not follow any specific trends. The average dose for different volumes of all organs for both machines for the same MU and fluence are tabulated in the Table 10, along with the difference in dose between the two machines.
Table 10

Average dose obtained from TPS for the same fluence and MU for the case of IMRT pelvis (in Gy).

D2 D5 D10 D50 D98 Dmean
PTVClinac iX 52.0±0.46 51.9±0.43 51.3±0.44 50.3±0.41 49.3±0.42 47.8±0.78 50.2±0.43
Unique 51.6±0.47 51.2±0.48 50.9±0.48 50.0±0.41 48.9±0.43 47.5±0.75 49.9±0.43
Unique 0.36±0.12 0.34±0.14 0.34±0.13 0.31±0.08 0.34±0.11 0.29±0.15 0.31±0.09
RectumClinac iX 50.6±0.63 50.0±0.66 49.2±0.90 37.1±4.19 17.2±4.65 10.5±3.20 35.1±2.46
Unique 50.1±0.62 49.5±0.62 48.7±0.84 36.6±4.22 16.7±4.65 9.98±3.04 34.7±2.47
Difference 0.42±0.14 0.45±0.19 0.45±0.17 0.43±0.13 0.55±0.10 0.50±0.18 0.46±0.11
BladderClinac iX 51.6±0.67 51.2±0.56 50.9±0.54 47.7±3.46 33.6±12.5 28.5±15.8 44.85±4.59
Unique 51.2±0.62 50.8±0.55 50.5±0.54 47.2±3.38 33.2±12.5 28.2±15.7 44.4±4.54
Difference 0.39±0.14 0.38±0.07 0.39±0.07 0.44±0.11 0.40±0.10 0.34±0.16 0.43±0.08
Lt.femurClinac iX 35.3±3.53 32.0±3.30 29.0±2.41 20.0±2.23 13.1±2.90 10.0±2.86 20.7±1.97
Unique 35.0±3.53 31.6±3.32 28.6±2.43 19.7±2.24 12.7±2.91 9.6±2.85 20.3±1.98
Difference 0.35±0.10 0.39±0.08 0.40±0.10 0.36±0.10 0.37±0.07 0.36±0.09 0.37±0.07
Rt.femurClinac iX 35.3±4.22 32.4±4.16 29.5±3.61 20.3±2.75 12.8±2.50 9.8±2.89 21.0±2.58
Unique 34.9±4.24 32.0±4.16 29.1±3.60 19.9±2.74 12.5±2.47 9.5±2.85 20.6±2.58
Difference 0.40±0.09 0.42±0.09 0.41±0.08 0.39±0.09 0.38±0.07 0.33±0.10 0.39±0.07
At different levels of dose the average dose difference for PTV varies from 0.29 Gy to 0.36 Gy. The average dose difference between the two machines for the structures adjacent to the target, such as the bladder and rectum, at different dose levels of varies from 0.33 Gy to 0.55 Gy. For less critical structures like femoral heads, a variation of 0.33 Gy to 0.42 Gy is observed at all levels of dose parameters. The maximum variation of 0.55 Gy was found between the two machines at rectum in levels. Average dose obtained from TPS for the same fluence and MU for the case of IMRT pelvis (in Gy). In pelvis RapidArc plans, the average dose for different volumes of all organs for both machines for the same MU and fluence are shown in the Table 11, along with the dose difference between the two machines. Average dose difference for PTV between the two machines at different levels of dose varies from 0.29 Gy to 0.34 Gy. Looking at the statistics of critical structures such as the bladder and rectum at different dose levels of D2, , the deviation is found to vary from 0.30 Gy to 0.53 Gy. In case of femoral heads the average dose difference between the two machines at different levels of dose varies from 0.21 Gy to 0.42 Gy. The overall maximum variation of 0.53 Gy between the two machines was found for bladder at and levels.
Table 11

Average dose obtained from TPS for the same fluence and MU for the case of Rapid Arc pelvis (in Gy).

D2 D5 D10 D50 D90 D98 Dmean
PTVClinac iX 52.1±0.45 51.9±0.44 51.6±0.43 50.7±0.35 49.6±0.33 48.4±0.46 50.6±0.35
Unique 51.9±0.43 51.6±0.42 51.3±0.42 50.4±0.38 49.3±0.34 48.0±0.45 50.3±0.36
Difference 0.29±0.09 0.29±0.10 0.30±0.11 0.34±0.09 0.33±0.14 0.34±0.11 0.34±0.09
RectumClinac iX 51.0±0.73 50.3±1.07 49.2±1.70 27.8±6.51 11.9±2.60 9.05±2.04 29.5±3.55
Unique 50.6±0.70 50.0±1.04 48.9±1.68 27.3±6.51 11.5±2.56 8.75±2.03 29.1±3.51
Difference 0.32±0.10 0.33±0.09 0.35±0.08 0.46±0.20 0.40±0.17 0.30±0.18 0.41±0.15
BladderClinac iX 51.7±0.53 51.3±0.61 50.7±0.97 44.1±5.90 32.3±7.17 25.3±9.72 42.7±4.75
Unique 51.2±0.65 50.7±0.69 50.2±1.03 43.5±5.90 31.9±7.18 24.9±9.64 42.3±4.75
Difference 0.52±0.18 0.53±0.15 0.52±0.13 0.53±0.19 0.43±0.11 0.37±0.17 0.47±0.11
Lt.femurClinac iX 37.2±2.71 34.1±2.31 31.1±1.91 21.8±2.00 16.6±2.23 14.8±2.42 22.9±2.00
Unique 36.8±2.68 33.7±2.25 30.7±1.86 21.5±1.94 16.3±2.19 14.6±2.34 22.6±1.94
Difference 0.41±0.16 0.41±0.14 0.38±0.11 0.35±0.14 0.26±0.12 0.21±0.14 0.35±0.13
Rt.femurClinac iX 35.8±4.21 33.1±3.53 30.4±2.94 21.9±2.80 16.2±2.52 14.5±2.57 22.7±2.60
Unique 35.4±4.23 32.6±3.48 29.9±2.90 21.5±2.76 16.0±2.46 14.2±2.52 22.3±2.55
Difference 0.40±0.16 0.42±0.15 0.41±0.15 0.34±0.13 0.28±0.13 0.27±0.13 0.41±0.25
Average dose obtained from TPS for the same fluence and MU for the case of Rapid Arc pelvis (in Gy). Unlike H&N, in pelvic plans variations observed in the organs do not seem to be technique‐oriented. They are due to deep‐seated as well body‐centered PTV. The variations in PTV do not follow any trends. Maximum dose variation in rectum is observed at and for IMRT and RapidArc respectively. Surprisingly the variation pattern observed in rectum is similar to that of the parotids in H&N, rectum being an overlapping structure. The bladder shows slightly higher variation in the RapidArc plans. Femoral heads never follow any trends, neither in volumes nor in technique. A closer look at beam‐matching results reveals that notable dose difference is observed in H&N patients, especially in RapidArc techniques. Generally, the complexity of H&N plans is relatively greater than in pelvis plans because more critical organs are involved in the H&N region. In addition, the results show that volumes receive more doses in Clinac iX machine as compared to 600 C/D (Unique) machine for the same monitor units. This is due to the reciprocal effect of output factor variation observed at particular field sizes between the machines, which in turn is due to the different head designs employed between the two linear accelerators.

V. CONCLUSION

The accuracy of “beam matching” between the two machines for 6 MV X‐rays when the linear accelerators are set within the manufacturer's specifications was evaluated in a systematic and detailed manner. The results of comparisons made between the beam‐matched machines for PDDs, profiles, and output factors are within the range of satisfaction. Comparative dosimetric analysis of IMRT and RapicArc patient plans swapped between the two machines at the TPS level in both the H&N and pelvis treatment sites are well within clinically acceptable tolerance. Overall results show that in a busy center, during down times, patients can be shifted across the beam‐matched machines with the assurance of pretreatment verification alone, without the need for replanning.

COPYRIGHT

This work is licensed under a Creative Commons Attribution 3.0 Unported License. Supplementary Material Click here for additional data file. Supplementary Material Click here for additional data file. Supplementary Material Click here for additional data file. Supplementary Material Click here for additional data file. Supplementary Material Click here for additional data file. Supplementary Material Click here for additional data file. Supplementary Material Click here for additional data file. Supplementary Material Click here for additional data file. Supplementary Material Click here for additional data file. Supplementary Material Click here for additional data file. Supplementary Material Click here for additional data file.
  10 in total

1.  Comparative measurements on a series of accelerators by the same vendor.

Authors:  R J Watts
Journal:  Med Phys       Date:  1999-12       Impact factor: 4.071

2.  Reference photon dosimetry data and reference phase space data for the 6 MV photon beam from varian clinac 2100 series linear accelerators.

Authors:  Sang Hyun Cho; Oleg N Vassiliev; Seungsoo Lee; H Helen Liu; Geoffrey S Ibbott; Radhe Mohan
Journal:  Med Phys       Date:  2005-01       Impact factor: 4.071

3.  Quantitative evaluation of a beam-matching procedure using one-dimensional gamma analysis.

Authors:  Jan Hrbacek; Tom Depuydt; An Nulens; Ans Swinnen; Frank Van den Heuvel
Journal:  Med Phys       Date:  2007-07       Impact factor: 4.071

4.  A beam-matching concept for medical linear accelerators.

Authors:  David Sjöström; Ulf Bjelkengren; Wiviann Ottosson; Claus F Behrens
Journal:  Acta Oncol       Date:  2009       Impact factor: 4.089

Review 5.  American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: quality assurance for clinical radiotherapy treatment planning.

Authors:  B Fraass; K Doppke; M Hunt; G Kutcher; G Starkschall; R Stern; J Van Dyke
Journal:  Med Phys       Date:  1998-10       Impact factor: 4.071

6.  IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119.

Authors:  Gary A Ezzell; Jay W Burmeister; Nesrin Dogan; Thomas J LoSasso; James G Mechalakos; Dimitris Mihailidis; Andrea Molineu; Jatinder R Palta; Chester R Ramsey; Bill J Salter; Jie Shi; Ping Xia; Ning J Yue; Ying Xiao
Journal:  Med Phys       Date:  2009-11       Impact factor: 4.071

Review 7.  AAPM code of practice for radiotherapy accelerators: report of AAPM Radiation Therapy Task Group No. 45.

Authors:  R Nath; P J Biggs; F J Bova; C C Ling; J A Purdy; J van de Geijn; M S Weinhous
Journal:  Med Phys       Date:  1994-07       Impact factor: 4.071

8.  Accelerator beam data commissioning equipment and procedures: report of the TG-106 of the Therapy Physics Committee of the AAPM.

Authors:  Indra J Das; Chee-Wai Cheng; Ronald J Watts; Anders Ahnesjö; John Gibbons; X Allen Li; Jessica Lowenstein; Raj K Mitra; William E Simon; Timothy C Zhu
Journal:  Med Phys       Date:  2008-09       Impact factor: 4.071

9.  Dosimetric analysis of beam-matching procedure of two similar linear accelerators.

Authors:  Janhavi R Bhangle; V K Sathiya Narayanan; Namitha K Kumar; R Vaitheeswaran
Journal:  J Med Phys       Date:  2011-07

10.  Commissioning measurements for photon beam data on three TrueBeam linear accelerators, and comparison with Trilogy and Clinac 2100 linear accelerators.

Authors:  Gloria P Beyer
Journal:  J Appl Clin Med Phys       Date:  2013-01-07       Impact factor: 2.102

  10 in total
  3 in total

1.  Evaluation of beam matching accuracy among six linacs from the same vendor.

Authors:  Chockkalingam Krishnappan; Chandrasekaran Anu Radha; Karunakaran Balaji; Prasanna Kumar Mani; Vendhan Subramani; Velmurugan Thanigaimalai; Madhan Kumar Gunasekaran; Velayudham Ramasubramanian
Journal:  Radiol Phys Technol       Date:  2018-09-29

2.  Dosimetric Validation of Volumetric Modulated Arc Therapy with Three 6MV Beam-Matched Linear Accelerators

Authors:  Sangaiah Ashokkumar; K M Ganesh; K Ramalingam; K Karthikeyan; N Jagadheeskumar
Journal:  Asian Pac J Cancer Prev       Date:  2017-12-29

3.  Dosimetric and mechanical equivalency of Varian TrueBeam linear accelerators.

Authors:  Mohammed Ghazal; Lars Södergren; Mathias Westermark; Julia Söderström; Tobias Pommer
Journal:  J Appl Clin Med Phys       Date:  2020-10-18       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.