| Literature DB >> 29286616 |
Sangaiah Ashokkumar1,2, K M Ganesh, K Ramalingam, K Karthikeyan, N Jagadheeskumar.
Abstract
Background: To avoid inconvenience to patients due to linear accelerator down time in busy radio-therapy departments, treatment plans can be switched between linear accelerators provided that all exhibit the same same dosimetric characteristics. In other words linear accelerators should be beam-matched. Aim: The aim of this study was to evaluate the clinical significance of beam-matching using VMAT plans. Materials andEntities:
Keywords: VMAT; AAA; TPS; beam matching; 2300CD; 2100CD; unique performance
Mesh:
Year: 2017 PMID: 29286616 PMCID: PMC5980907 DOI: 10.22034/APJCP.2017.18.12.3439
Source DB: PubMed Journal: Asian Pac J Cancer Prev ISSN: 1513-7368
Energy Match Analysis
| Data | Field size (cm) | 2100 CD | 2300 CD | Unique | Average | SD |
|---|---|---|---|---|---|---|
| dmax (cm) | 4x4 | 1.5 | 1.66 | 1.37 | 1.51 | 0.15 |
| 10x10 | 1.4 | 1.58 | 1.39 | 1.46 | 0.11 | |
| 40x40 | 1.14 | 1.32 | 1.16 | 1.21 | 0.1 | |
| PDD10cm | 4X4 | 62 | 62.1 | 61.4 | 61.83 | 0.38 |
| 10X10 | 66.5 | 67.1 | 66.4 | 66.67 | 0.38 | |
| 40X40 | 71.6 | 71.6 | 71.5 | 71.57 | 0.06 | |
| PDD20cm | 4X4 | 33.7 | 33.9 | 33 | 33.53 | 0.47 |
| 10X10 | 38.2 | 38.6 | 38 | 38.27 | 0.31 | |
| 40X40 | 45.8 | 45.5 | 45.8 | 45.7 | 0.17 | |
| TPR20/10 | 4X4 | 0.63 | 0.63 | 0.62 | 0.63 | 0.01 |
| 10X10 | 0.67 | 0.67 | 0.67 | 0.67 | 0 | |
| 40X40 | 0.75 | 0.75 | 0.75 | 0.75 | 0 |
SD, Standard Deviation
Beam Profile Analysis
| Depth | Linear Accelerator | 10cmX10cm Field size | 40cmx40cm Field size | ||
|---|---|---|---|---|---|
| Penumbra(mm) | Field width(cm) | Penumbra(mm) | Field width(cm) | ||
| 1.5 cm | 2100CD | 5.4 | 10.14 | 5.6 | 40.2 |
| 2300CD | 5.6 | 10.2 | 5.5 | 40.9 | |
| UNIQUE | 5.5 | 10.18 | 5.6 | 40.93 | |
| 10 cm | 2100CD | 6.8 | 11.1 | 9.9 | 43.6 |
| 2300CD | 6.9 | 11.1 | 9.8 | 44.2 | |
| UNIQUE | 7 | 11.05 | 9.5 | 44.3 | |
Relative Output Factor
| Percentage of Deviation | |||||
|---|---|---|---|---|---|
| Field size | 2100CD | 2300CD | Unique | 2100 vs. 2300 | 2100 vs. unique |
| 4X4 | 0.8993 | 0.872 | 0.918 | 3.04 | -2.08 |
| 8X8 | 0.9763 | 0.954 | 0.98 | 2.28 | -0.38 |
| 10X10 | 1 | 1 | 1 | 0 | 0 |
| 15X15 | 1.0406 | 1.0316 | 1.035 | 0.86 | 0.54 |
| 25X25 | 1.0875 | 1.0776 | 1.076 | 0.91 | 1.06 |
| 30X30 | 1.1062 | 1.0979 | 1.088 | 0.75 | 1.65 |
| 35X35 | 1.119 | 1.1156 | 1.098 | 0.3 | 1.88 |
| 40X40 | 1.1297 | 1.1316 | 1.102 | -0.17 | 2.45 |
MLC Parameters
| Machine | MLC Transmission | Dosimetric Leaf Gap |
|---|---|---|
| 2100CD | 1.45% | 0.2cm |
| 2300CD | 1.40% | 0.2cm |
| UNIQUE | 1.15% | 0.17cm |
MLC Stripping Field Dose Deviation
| DLS | TPS Dose 2100CD | Measured Dose (cGy) | Dose Deviation (cGy) | ||||
|---|---|---|---|---|---|---|---|
| (cGy) | 2100CD | 2300CD | Unique | 2100CD | 2300CD | Unique | |
| 0.5 mm | 15.4 | 15.96 | 17.47 | 15.3 | 0.56 | 2.07 | -0.1 |
| 1 mm | 17.6 | 18.16 | 19.51 | 17.79 | 0.56 | 1.91 | 0.19 |
| 2 mm | 22 | 22.25 | 23.87 | 22.45 | 0.25 | 1.87 | 0.45 |
| 5 mm | 34.6 | 34.85 | 36.68 | 34.53 | 0.25 | 2.08 | -0.07 |
| 10 mm | 54.1 | 54.47 | 55.9 | 54.08 | 0.37 | 1.8 | -0.02 |
| 15 mm | 88.2 | 88.43 | 90.54 | 88.06 | 0.23 | 2.34 | -0.14 |
p -values for Head and Neck Cases
| p-Value | |||||||
|---|---|---|---|---|---|---|---|
| Spine Max | PRV Spine | Brain Stem | Lt. Parotid | Rt. Parotid | Body Mean | PTV Mean | |
| 2100CD Vs 2300CD | 0.311 | 0.743 | 0.953 | 0.948 | 0.917 | 0.901 | 0.426 |
| 2100CD Vs Unique | 0.472 | 0.362 | 0.873 | 0.348 | 0.323 | 0.828 | 0.264 |
| 2300CD Vs Unique | 0.022 | 0.223 | 0.828 | 0.384 | 0.38 | 0.794 | 0.072 |
p -values for Thorax Cases
| p-Value | ||||||
|---|---|---|---|---|---|---|
| Spine | PRV Spine | Left Lung Mean Dose | Right Lung Mean | Mean Body | PTV Mean | |
| 2100CD Vs 2300CD | 0.8745 | 0.8519 | 0.9729 | 0.9822 | 0.9599 | 0.864 |
| 2100CD Vs Unique | 0.8068 | 0.8022 | 0.8557 | 0.9238 | 0.8032 | 0.8374 |
| 2300CD Vs Unique | 0.6885 | 0.6626 | 0.8294 | 0.9062 | 0.7647 | 0.7066 |
p -values for Pelvic Cases
| p-Value | |||||||
|---|---|---|---|---|---|---|---|
| Rectum | Bladder | Bowel | LT.Hof | RT.Hof | Mean Body | PTV Mean | |
| 2100CD Vs 2300CD | 0.576 | 0.707 | 0.904 | 0.255 | 0.916 | 0.958 | 0.042 |
| 2100CD Vs Unique | 0.673 | 0.687 | 0.898 | 0.407 | 0.384 | 0.995 | 0.094 |
| 2300CD Vs Unique | 0.341 | 0.437 | 0.799 | 0.569 | 0.332 | 0.96 | 0.001 |
Figure 1Point Dose Analysis of Three Machines
Figure 2Gamma Analysis for 3% Dose Difference and 3mm Distance to Agreement Criteria