| Literature DB >> 27660042 |
Gabriella Captur1,2,3,4, Peter Gatehouse5, Kathryn E Keenan6, Friso G Heslinga7,8, Ruediger Bruehl9, Marcel Prothmann10, Martin J Graves11, Richard J Eames12, Camilla Torlasco13, Giulia Benedetti14, Jacqueline Donovan15, Bernd Ittermann9, Redha Boubertakh16, Andrew Bathgate17, Celine Royet17, Wenjie Pang17, Reza Nezafat18, Michael Salerno19, Peter Kellman20, James C Moon21,22,23.
Abstract
BACKGROUND: T1 mapping and extracellular volume (ECV) have the potential to guide patient care and serve as surrogate end-points in clinical trials, but measurements differ between cardiovascular magnetic resonance (CMR) scanners and pulse sequences. To help deliver T1 mapping to global clinical care, we developed a phantom-based quality assurance (QA) system for verification of measurement stability over time at individual sites, with further aims of generalization of results across sites, vendor systems, software versions and imaging sequences. We thus created T1MES: The T1 Mapping and ECV Standardization Program.Entities:
Keywords: Phantom; Standardization; T1 mapping
Year: 2016 PMID: 27660042 PMCID: PMC5034411 DOI: 10.1186/s12968-016-0280-z
Source DB: PubMed Journal: J Cardiovasc Magn Reson ISSN: 1097-6647 Impact factor: 5.364
Fig. 1Internal and external phantom structure. Internal (3 T, looking at the front—a) and external (1.5 T, front and back—b) T1MES phantom structure. The nine tubes are supported on a translucent resin base composed of unsaturated polyester/styrene. A careful hardening and curing process ensured a smooth surface finish for the resin base. The front of the phantom (b left) contains an isocenter cross label to aid positioning as well as an LCD thermometer. Careful positioning of the bottle on the scanner table (c) with the cap towards its head end is needed to ensure it is scanned at isocenter each time. HDPE = high-density polyethylene; LCD = liquid crystal display; NiCl2 = Nickel Chloride; PE = polyethylene; PVC = poly vinyl chloride
Fig. 2Artifact examples in earlier prototypes (a-g) and final T1MES phantom (i, j). Four earlier prototypes (models A—D) were rejected before the final model. a Coronal image of the earlier A-model (aqueous fill) showing bright artifacts around the tubes resulting from bSSFP going off-resonance that would have led to variations in T1 values by MOLLI and similar sequences. b Transverse image of A-model showing the characteristic ‘cat’s head’ artifact of air-bubbles trapped in the paramagnetically doped aqueous tubes. Significant off-resonance artifact is also noticeable in the central tubes. c Another coronal image through A-model but with larger gaps between tubes showing the combined effect of motion artifact (due to the aqueous fill) and B 0 distortion. d Transverse image of C-model attempting to use narrower tubes to pack 12 instead of 9, but significant Gibbs artifact can be seen in each tube. e Transverse image of C-model showing three small dark circular artifacts (12, 3 and 9 o’clock positions) caused by glue used to stabilize the tube arrangement. We subsequently switched to silicone-based glues that were less likely to trap air bubbles and were artifact-free. f Severe stabilisation artifact appearing as a thick dark band around the border of a D-model—here the phantom was scanned immediately after being received from the courier company and the bottle was still very cold from the transportation. Additionally susceptibility artifacts can be seen as thin linear bands spoiling some of the tubes (9 and 3 o’clock). g Significant image intensity inhomogeneity during a D-model test session on a GE scanner caused by accidental omission of the folded blanket, intended to separate the phantom bottle from the anterior chest coil. h Curved tube artifact and dark rings arising from ink printed onto the sides of digestive tubes (images courtesy of K. E. Keenan and NIST). i Coronal bSSFP localiser image and (j) typical T1 map of a final 3 T T1MES phantom obtained by MOLLI using a bSSFP readout on a Siemens 3 T Skyra scanner. bSSFP = balanced steady-state free precession; MOLLI = modified Look-Locker inversion recovery. Other abbreviation as in Fig. 1
Fig. 3Prototype models and T1MES project timeline. CE = Conformité Européene; FDA = Food and Drug Administration; GE = General Electric; NIST = US National Institute of Standards and Technology; PTB = German Physikalisch-Technische Bundesanstalt; QA = quality assurance; RH = Resonance Health
Design factors when developing a T1 mapping phantom
| Design factor | Explanation | Our proposed solution |
|---|---|---|
| Bottle magnetostatics and | The ideal phantom would be uniform and ellipsoidal to avoid susceptibility-induced magnetostatic field perturbation. Such a phantom would permit sphere of Lorentz uniformity but this is not easily mass produced. Many phantoms are cylindrical with the long axis along the static field, | An outer phantom body with a smooth surface and soft rounded-edges, placed inside |
| Long term gel stability and risk of moulding | Phantoms with long-term stability could assure the stability of methods applied to patients against scanner alternations and across multiple centers. | Moulding was prevented by aseptic manufacturing, the toxicity of Ni2+ ions, and the absence of nutrients in the type of agarose used. Tap water might contain microbial contamination and metal ions so high purity water was used. The main risk is from contraction of gel on loss of water leading to gaps and water condensation but NiCl2-doped agarose gel phantoms can be stable over a 1-year period [ |
| Seal, leakages, air trapping for aqueous fill | Air pockets in the agarose gel phantom will give rise to susceptibility artifacts on account of the large mismatch in static magnetic susceptibility between air and surrounding gel producing a local distortion in magnetic field strength. | The main phantom was sealed by a black polypropylene screw cap fitted with a polyethylene foam insert. Each internal digestive tube was sealed by a tight screw cap. Gel preparation with warm, degassed water reduced air bubble formation. Note the tube “base-upward” setting procedure and subsequent “top-up” of the contracted gel in each tube after setting, described in the text. |
| Adjustments of | Adjustments of | We specified a constant shim volume for each scan. This is manufacturer-dependent—see the T1MES manual [ |
| Gel diamagnetism | In the T1MES model system, because the impact of the paramagnetic ions is so small, we can conceptually treat the main phantom box as if it had no tubes, as if it were just filled with uniform gel throughout | The T1MES system has partly paramagnetic and partly diamagnetic constituents, but the impact of the paramagnetic Ni2+ ions is small, around 10 % (because concentrations are small) so the overall interaction is diamagnetic, considering the ~9 parts per million diamagnetism of most tissues relative to air from Lenz electronic diamagnetism. |
| Gibbs artifact ringing and other inplane effects | Truncating artifacts appear as lines of alternating brightness and darkness in the read-out and phase encode direction. Some effects also from asymmetric readout and | Large diameter digestive tubes to house the 9 agarose doped solutions, so that central regions of each tube are sufficiently distant (a number of pixels away) from regions impacted by artifacts from abrupt signal intensity transitions at the tube edges. |
| 1.4 T, 1.5 T, 3 T performance | Many paramagnetic relaxation modifiers, including Mn2+ and Cu2+, exhibit significant frequency dependence. | We used Ni2+[ |
| T1|T2 ranges: blood/myocardium, pre/post-GBCA | The T1|T2 values were carefully modelled for native and post-gadolinium based contrast agent, blood and myocardium. | 5 common tubes, 4 tubes specific to 1.5 T, 4 tubes specific to 3 T. There was no macromolecular addition (no magnetisation transfer modelling) [ |
| Tube arrangement | The phantom corners are more prone to inhomogeneities of the | Longer T1 tubes were placed nearer the middle of the phantom layout and avoided the corners. |
Cu copper ions, Mn manganese ions, Ni nickel ions, NiCl nickel chloride
List of T1|T2 values for the target 13 tubes and outer matrix gel and the required agarose/NiCl2 concentrations for the final phantom
| Description target ( | T1 (ms at 1.4 Ta) | T2 (ms at 1.4 Ta) | Agarose (%) | NiCl2 (mM) |
|---|---|---|---|---|
| “Short” post-GBCA blood ( | 256 | 172 | 0.244 | 5.547 |
| “Normal” native blood 1.5 T ( | 1490 | 282 | 0.373 | 0.362 |
| “Long” post-GBCA blood ( | 427 | 212 | 0.325 | 2.860 |
| “Short” native myocardium 1.5 T ( | 818 | 54 | 2.214 | 1.231 |
| “Long” native myocardium 1.5 T ( | 1384 | 57 | 2.279 | 0.461 |
| “Medium” native myocardium 1.5 T ( | 1107 | 56 | 2.256 | 0.725 |
| “Short” post-GBCA myocardium ( | 295 | 50 | 2.174 | 4.510 |
| “Long” post-GBCA myocardium ( | 557 | 51 | 2.377 | 2.103 |
| “Medium” post-GBCA myocardium ( | 429 | 50 | 2.306 | 2.942 |
| “Normal” native blood 3 T ( | 1870 | 288 | 0.388 | 0.180 |
| “Short” native myocardium 3 T ( | 1043 | 56 | 2.245 | 0.858 |
| “Long” native myocardium 3 T ( | 1510 | 55 | 2.289 | 0.342 |
| “Medium” native myocardium 3 T ( | 1279 | 56 | 2.273 | 0.531 |
| Outer matrix gel fill | 846 | 141 | 0.780 | 1.155 |
aBy Bruker minispec mq60 relaxometer 1.4 T (22 °C) at Resonance Health laboratory, Australia
GBCA gadolinium-based contrast agents, ID identity number
Fig. 4T1 and T2 values in T1MES. T1 and T2 values in the phantom mimic those of myocardium and blood pre and post-GBCA at 1.5 T (Panel a) and 3 T (Panel b). The 13 relaxometry scopes (refer to Table 2) are explained in the figure. Slow scan reference data for T1|T2 is displayed in green (for T1 by slow IRSE and for T2 by slow SE, RR interval 900 ms at 21 ± 2 °C), T1 values shown in orange represent the mean value per tube derived from tests on five of the E-model phantoms (using a 5(3)3 256-matrix RR = 900 ms at 21 ± 2 °C variant of MOLLI adapted for native T1 mapping; Siemens WIP 448B at 1.5 T and WIP 780B at 3 T), and in blue are T1|T2 values obtained by the manufacturer in Australia using a 1.4 T Bruker minispec relaxometer at 22 °C. Tube arrangement is such that long T1 tubes potentially suffering from more artifacts are kept towards the middle of the phantom and away from the corners. GBCA = gadolinium-based contrast agents; IRSE = inversion recovery spin echo; myo = myocardium; RR = inter-beat interval; SE = spin echo. All T1|T2 values are stated in ms. Other abbreviation as in Fig. 2
Fig. 5T1 and T2 relaxation times versus ingredients at 1.4 T: agarose and NiCl2. Grid represents results of the model. Red points represent single measurements. a Longitudinal relaxation time constant (T1), RMSE in R1 compared to the linear model was 4.8 × 10−5 /ms. b Spin–spin relaxation time (T2), RMSE in R2 compared to the linear model = 5.3 × 10−4 /ms. Since the x and y axes of both fits are comparable, the ingredient that contributes most can be identified. RMSE = root mean square error
Fig. 6Reference T1|T2 values. Variation in the mean T1 (red dots) and T2 (blue dots) reference values and standard deviation (whiskers) of the nine tubes averaged for the ten final batch T1MES phantoms that underwent ‘gold standard’ slow T1 and T2 measurements by IRSE and SE respectively at 1.5 T (a) and 3 T (b). T1 values obtained by MOLLI (5(3)3 [256] (WIP# 448B at 1.5 T and WIP# 780B at 3 T) pre-GBCA sequence (green dots) are also shown. Abbreviations as in Figs. 2 and 4
Fig. 7B 0 and B 1 field homogeneity. a B field homogeneity across the nine phantom compartments as a measure of off-resonance in Hz at 3 T (single E-model phantom results). These are extremely small shifts in frequency (30 Hz = 0.25 ppm) at 3 T and should not be regarded as significantly different between the tubes. b Diagonal profile of the B 1 field (as per green discontinuous line in the inset) comparing relative flip angles on a Siemens 3 T system. Variance of B 1 was smallest across the 9 compartments with CoV 1.54 % for HDPE beads consisting of smooth, semi-translucent, colourless compact discs (as colouring in plastics has the potential to distort the B 0 magnetic field [12], see Fig. 2h) with a melt index >60 °C. We choose pellets that had not been regrinded, reblended or composite for this purpose. Highly monosized microbeads measured 6 μm and were composed of crosslinked PMMA polymer. Neither microbeads, sucrose nor NaCl were comparably effective in flattening the B 1 field. PMMA = poly methyl methacrylate. Other abbreviation as in Fig. 4
Fig. 8Temperature experiments in T1MES. Temperature dependency experiments (Test 1 in methods) performed on a D-model whole phantom (tube nomenclature differed from that used in E-models) comparing the stability of T1 (a) and T2 (b) values between two repeat experiments (2 days apart) at various temperatures between 15 °C and 32 °C on a 3 T Siemens Verio system. Whiskers represent mean ± standard error. (c) Temperature dependency experiment (Test 2 in methods) comparing T1|T2 values in tubes A, B, C, D, E and I (middle right insert) from a final E-model phantom across five temperatures
Fig. 9Short-term reproducibility. Short-term reproducibility (three runs) at the NIST laboratory (Test 1 in methods) for phantom T1values in six loose tubes (top left insert) from a final E-model phantom showing CoV of 1 % or less. Tube B with the longest T1|T2 showed the greatest variability between reads. CoV = coefficient of variation
Short-term reproducibility experiments in a 3 T final phantom (E-model)*
| Tube | Parameter | Sequence | CoV (%) | Mean diff. ± s.d. |
|---|---|---|---|---|
| A | T1 | pre_MOLLI_5(3)3_256_T1 | 0.16 | 255 ± 0.4 |
| post_MOLLI_4(1)3(1)2_256_MOCO_T1 | 0.18 | 255 ± 0.5 | ||
| T2 | T2_4pt_TRUFI_192i_T2 | 0.66 | 194 ± 1.3 | |
| T2_4pt_GRE_192i_T2 | 0.61 | 134 ± 0.8 | ||
| J | T1 | pre_MOLLI_5(3)3_256_T1 | 0.14 | 1860 ± 2.6 |
| post_MOLLI_4(1)3(1)2_256_MOCO_T1 | 0.17 | 1672 ± 2.8 | ||
| T2 | T2_4pt_TRUFI_192i_T2 | 4.06 | 227 ± 9.2 | |
| T2_4pt_GRE_192i_T2 | 1.37 | 203 ± 2.8 | ||
| C | T1 | pre_MOLLI_5(3)3_256_T1 | 0.08 | 460 ± 0.4 |
| post_MOLLI_4(1)3(1)2_256_MOCO_T1 | 0.08 | 461 ± 0.4 | ||
| T2 | T2_4pt_TRUFI_192i_T2 | 0.52 | 195 ± 1.0 | |
| T2_4pt_GRE_192i_T2 | 0.76 | 160 ± 1.2 | ||
| K | T1 | pre_MOLLI_5(3)3_256_T1 | 0.13 | 953 ± 1.2 |
| post_MOLLI_4(1)3(1)2_256_MOCO_T1 | 0.10 | 917 ± 0.9 | ||
| T2 | T2_4pt_TRUFI_192i_T2 | 0.98 | 60 ± 0.6 | |
| T2_4pt_GRE_192i_T2 | 0.67 | 49 ± 0.3 | ||
| L | T1 | pre_MOLLI_5(3)3_256_T1 | 0.08 | 1372 ± 1.1 |
| post_MOLLI_4(1)3(1)2_256_MOCO_T1 | 0.16 | 1252 ± 2.0 | ||
| T2 | T2_4pt_TRUFI_192i_T2 | 0.91 | 56 ± 0.5 | |
| T2_4pt_GRE_192i_T2 | 0.89 | 49 ± 0.4 | ||
| M | T1 | pre_MOLLI_5(3)3_256_T1 | 0.15 | 1178 ± 1.8 |
| post_MOLLI_4(1)3(1)2_256_MOCO_T1 | 0.12 | 1104 ± 1.3 | ||
| T2 | T2_4pt_TRUFI_192i_T2 | 0.91 | 58 ± 0.5 | |
| T2_4pt_GRE_192i_T2 | 0.66 | 49 ± 0.3 | ||
| G | T1 | pre_MOLLI_5(3)3_256_T1 | 0.19 | 285 ± 0.6 |
| post_MOLLI_4(1)3(1)2_256_MOCO_T1 | 0.20 | 285 ± 0.6 | ||
| T2 | T2_4pt_TRUFI_192i_T2 | 0.29 | 86 ± 0.2 | |
| T2_4pt_GRE_192i_T2 | 1.02 | 49 ± 0.5 | ||
| H | T1 | pre_MOLLI_5(3)3_256_T1 | 0.11 | 527 ± 0.6 |
| post_MOLLI_4(1)3(1)2_256_MOCO_T1 | 0.09 | 527 ± 0.5 | ||
| T2 | T2_4pt_TRUFI_192i_T2 | 0.35 | 66 ± 0.2 | |
| T2_4pt_GRE_192i_T2 | 0.72 | 46 ± 0.3 | ||
| I | T1 | pre_MOLLI_5(3)3_256_T1 | 0.06 | 406 ± 0.3 |
| post_MOLLI_4(1)3(1)2_256_MOCO_T1 | 0.05 | 409 ± 0.2 | ||
| T2 | T2_4pt_TRUFI_192i_T2 | 0.21 | 72 ± 0.2 | |
| T2_4pt_GRE_192i_T2 | 0.19 | 47 ± 0.1 |
*All tests performed at 20.3 ± 0.48 °C on Siemens, Skyra 3 T at RBHT, November 2015 with RR interval 900 ms and using two T1 mapping sequences (pre-MOLLI 5(3)3 [256] and post-MOLLI 4(1)3(1)2 [256] with MOCO, WIPs# 780B) and two T2 mapping sequences (TRUFI T2 map and GRE T2 map)
CoV coefficients of variation, diff. difference, GRE gradient echo, MOCO motion correction, MOLLI modified Look-Locker inversion recovery, RR inter-beat interval, s.d. standard deviation, TRUFI true fast imaging with steady-state free precession
Quality assurance of T1 mapping: the initial T1MES CMR centers
| Center | Magnet characteristics | ||||||
|---|---|---|---|---|---|---|---|
| Vendor | Tesla | Name | YOM | Software | Boreb (cm) | Gradient performancec | |
| St Thomas’ Hospital UK | Siemens | 1.5 | Aera | 2015 | VE11 | 70 | 45/200 |
| St Thomas’ Hospital UK | Philips | 1.5 | Ingenia | 2013 | R4.1.3SP2 | 70 | 33/200 |
| Oslo University Hospital Norway | Siemens | 1.5 | Aera | 2014 | VE11 | 70 | 40/200 |
| Bristol Heart Institute UK | Siemens | 1.5 | Avanto | 2009 | VB17A | 60 | 44/180 |
| Diagnostikum Berlin Germany | Siemens | 1.5 | Aera | 2015 | VE11 | 70 | 45/200 |
| GOSH UK | Siemens | 1.5 | Avanto | 2007 | VB17 | 60 | 40/180 |
| NIH Bethesda US | Siemens | 1.5 | Aera | 2014 | VE11 | 70 | 45/200 |
| Pittsburgh Pennsylvania US | Siemens | 1.5 | Espree | 2009 | VB17A | 70 | 40/200 |
| Leiden UMC The Netherlands | Philips | 1.5 | Ingenia | 2014 | R5.1.7SP2 | 70 | 45/200 |
| Leeds General Infirmary UK | Philips | 1.5 | Ingenia | 2014 | R5.1.7SP2 | 70 | 45/200 |
| MUMC The Netherlands | Philips | 1.5 | Ingenia | 2012 | R 5.1.7SP2 | 70 | 45/200 |
| Policlinico San Donato Italy | Siemens | 1.5 | Aera | 2012 | VD13A | 70 | 45/200 |
| Papworth UK | Siemens | 1.5 | Avanto | 2008 | VB17A | 60 | 50/200 |
| Wythenshawe Manchester UK | Siemens | 1.5 | Avanto | 2008 | VB17A | 60 | 45/200 |
| Copenhagen University Hospital Denmark | Siemens | 1.5 | Avanto | 2008 | VD13A | 60 | 45/200 |
| Queen Elizabeth Hospital Birmingham UK | Siemens | 1.5 | Avanto | 2008 | VB17A | 60 | 33/125 |
| Birmingham Children’s Hospital UK | Siemens | 1.5 | Avanto | 2010 | VB17A | 60 | 33/125 |
| University of Kentucky USA | Siemens | 1.5 | Aera | 2012 | VD13A | 70 | 45/200 |
| Charles Perkins Sydney Australia | Siemens | 1.5 | Avanto | 2013 | VE17A | 70 | 45/200 |
| Taichung Veterans Hospital Taiwan | Siemens | 1.5 | Aera | 2005 | VE11 | 60 | 45/200 |
| Monash Heart Australia | Siemens | 1.5 | Avanto | 2010 | VB17 | 55 | 40/200 |
| Niguarda Hospital Milan Italy | Siemens | 1.5 | Avanto | 2005 | VB17A | 60 | 40/200 |
| Golden Jubilee Glasgow UK | Siemens | 1.5 | Avanto | 2008 | VB17A | 60 | 45/200 |
| T-T!ME Multi-center phantoma | |||||||
| INSERM U1044 France | Siemens | Aera | 2012 | VD13A | 70 | 40/200 | |
| King Abdul-Aziz Saudi Arabia | GE | 1.5 | Discovery MR450 | 2012 | DV24 | 60 | 50/200 |
| Prince Charles Hospital Queensland | Siemens | 1.5 | Aera | 2011 | VD13A | 70 | 45/200 |
| Federal Medical Center Moscow | GE | 1.5 | Optima MR450w | 2014 | DV25 | 70 | 44/200 |
| Medical University of Vienna Austria | Siemens | 1.5 | Avanto | 2006 | VD13B | 60 | 40/200 |
| DHZ Berlin Germany | Philips | 1.5 | Achieva | 2008 | R5.1.8 | 60 | 33/180 |
| St George’s University London UK | Siemens | 1.5 | Aera | 2014 | E11 | 70 | 45/200 |
| RBHT London UK | Siemens | 1.5 | Avanto | 2005 | VB17A | 60 | 40/170 |
| University Hospital Southampton UK | Siemens | 1.5 | Avanto | 2006 | VB17A | 60 | 40/200 |
| Barts Heart Center London UK | Siemens | 1.5 | Aera | 2014 | VD13A | 70 | 45/200 |
| Barts Heart Center London UK | Siemens | 1.5 | Aera | 2015 | VE11A | 70 | 45/200 |
| The Heart Hospital London UK | Siemens | 1.5 | Avanto | 2009 | VD13A | 70 | 40/200 |
| Charité Campus Buch Germany | Siemens | 1.5 | Avanto | 2007 | VB13B | 60 | 40/200 |
| University of Virginia US | Siemens | 1.5 | Avanto | 2005 | VB17A | 60 | 45/200 |
| University of Virginia US | Siemens | 1.5 | Avanto | 2015 | VD13A | 60 | 45/200 |
| SIEMENS EU | Siemens | 1.5 | Aera | 2009 | VE11 | 70 | 45/200 |
| UZ Leuven Belgium | Philips | 1.5 | Ingenia | 2007 | R5.1.7 | 60 | 45/ 200 |
| UZ Leuven Belgium | Philips | 1.5 | Achieva XR | 2014 | R5.1.7 | 70 | 33/122 |
| Beth Israel Deaconess Medical Center, US | Philips | 1.5 | Achieva | 2005 | R3.2 | 60 | 33/180 |
| NIH Bethesda US | Siemens | 1.5 | Aera | 2012 | VD13A | 70 | 45/200 |
| St Thomas’ Hospital UK | Philips | 3 | Achieva TX | 2007 | R3.2.3 | 60 | 40/200 |
| St Thomas’ Hospital UK | Siemens | 3 | Biograph mMR | 2013 | VB20P | 60 | 45/200 |
| Fondazione Toscana Monasterio Pisa Italy | Philips | 3 | Ingenia | 2012 | R5.1.8 | 70 | 45/200 |
| Oslo University Hospital Norway | Philips | 3 | Ingenia | 2011 | 5.1.7 | 70 | 45/200 |
| Oslo University Hospital Norway | Siemens | 3 | Skyra | 2014 | VE11 | 70 | 45/120 |
| CRIC Bristol UK | Siemens | 3 | Skyra | 2009 | VD13C | 60 | 44/180 |
| Diagnostikum Berlin Germany | Siemens | 3 | Skyra | 2012 | VE11 | 70 | 45/200 |
| University of Aberdeen Scotland UK | Philips | 3 | Achieva TX | 2015 | R5.1.7 | 60 | 80/100 |
| NIH Bethesda US | Siemens | 3 | Verio | 2009 | VB17 | 70 | 33/125 |
| Leiden UMC The Netherlands | Philips | 3 | Achieva TX | 2012 | R5.1.8.2 | 70 | 45/200 |
| MUMC The Netherlands | Philips | 3 | Achieva TX | 2011 | R 3.2 | 60 | 40/200 |
| Wythenshawe Manchester UK | Siemens | 3 | Skyra | 2014 | VE11 | 70 | 45/200 |
| Copenhagen University Hospital Denmark | Siemens | 3 | Verio | 2010 | VB17 | 70 | 45/200 |
| Charles Perkins Sydney Australia | GE | 3 | Discovery MR750w | 2014 | DV25 | 70 | 44/200 |
| BHF Glasgow Center UK | Siemens | 3 | Prisma | 2015 | VE11 | 60 | 80/200 |
| INSERM U1044 France | Siemens | 3 | Prisma | 2015 | VE11 | 60 | 80/200 |
| DHZ Berlin Germany | Philips | 3 | Ingenia | 2011 | R5.1.8 | 70 | 45/200 |
| St George’s University London UK | Philips | 3 | Achieva TX | 2012 | R5.1 | 60 | 40/150 |
| RBHT London UK | Siemens | 3 | Skyra PTX | 2011 | VD13C | 70 | 43/180 |
| Barts Heart Center London UK | Siemens | 3 | Prisma | 2015 | VE11 | 60 | 80/200 |
| Leeds General Infirmary UK | Philips | 3 | Achieva TX | 2010 | R5.2 | 60 | 40/120 |
| Montreal Heart Institute Canada | Siemens | 3 | Skyra | 2012 | VD13A | 70 | 45/200 |
| PTB Germany | Siemens | 3 | Verio | 2010 | VB17A | 70 | 45/200 |
| University of Virginia US | Siemens | 3 | Skyra | 2011 | VE11A | 70 | 45/200 |
| UZ Leuven Belgium | Philips | 3 | Ingenia | 2010 | R5.1.7 | 70 | 45/200 |
| NIH Bethesda US | Siemens | 3 | Skyra | 2012 | VD13A | 70 | 45/200 |
| University of Queensland Australia | Siemens | 7 | Magnetom 7 | 2013 | VB17B | 60 | 72/200 |
| University of Queensland Australia | Siemens | 3 | Trio TIM | 2008 | VB17A | 60 | 45/200 |
| Glenfield Hospital Leicester UK | Siemens | 3 | Skyra | 2010 | VD13A | 70 | 45/200 |
| Baker IDI Australia | Siemens | 3 | Prisma | 2014 | VD13D | 60 | 80/200 |
| NIST USd | Agilent | 1.5 | Varian | 2013 | VnmrJ 4 | 14 | 300/475 |
| NIST USd | Agilent | 1.5 | Varian | 2013 | VnmrJ 4 | 14 | 300/475 |
aThis phantom is a gift to support the ongoing ’T-T!ME’ study. It will be scanned across multiple UK centers
bInner diameter i.e. around patient
cMaximum gradient performances as returned on the T1MES registration forms by each site. These values are subject to many modifying conditions. More relevant parameters such as TR and TE will be extracted from uploaded Digital Imaging and Communications in Medicine (DICOM) images where this is possible from DICOM
dLoose tubes only for 1.5 T and 3 T
YOM year of manufacture