P S Tofts1, G J Barker, T L Dean, H Gallagher, A P Gregory, R N Clarke.
Abstract
The selection of a suitable low dielectric constant liquid for radiofrequency coil nonuniformity measurements is described. Measurements of dielectric constant (relative permittivity) were made on a range of candidate liquids. After excluding liquids that were too inflammable or too viscous, Esso Bayol 82 oil (dielectric constant epsilon' = 2.37) was chosen. At 1.5 T, a 27 cm diameter cylindrical phantom (test object) filled with Bayol 82 has a maximum nonuniformity of 1.9% arising from radiofrequency standing waves. The maximum diameter cylinder that can be used without the nonuniformity exceeding 2% is given for a range of liquids and field strengths. The construction of customized containers that fit closely inside a radiofrequency head coil from glass fiber reinforced resin ("fiberglass") is described. Thermal expansion of the liquid takes place without a rise in the internal pressure in the container.Entities:
Mesh:
Substances:
Year: 1997 PMID: 9084027 DOI: 10.1016/s0730-725x(96)00220-2
Source DB: PubMed Journal: Magn Reson Imaging ISSN: 0730-725X Impact factor: 2.546