Accuracy of the ClearSight™ system in patients undergoing abdominal aortic aneurysm surgery.

Miho Sumiyoshi1, Takuma Maeda2,3, Erika Miyazaki4, Naoshi Hotta4, Hitoshi Sato4, Eisuke Hamaguchi4, Hiroko Kanazawa4, Yoshihiko Ohnishi4, Masataka Kamei1.   

Abstract

PURPOSE: The ClearSight™ device monitors continuous pressure and cardiac output via pulse contour analysis. ClearSight™, however, may not be reliable in patients with reduced peripheral perfusion caused by high peripheral resistance. This study aimed to elucidate the accuracy and trending ability of ClearSight™ in patients undergoing abdominal aortic aneurysm (AAA) surgery by comparing the ClearSightTM-derived cardiac index (CICS) with that measured using three-dimensional echocardiography (CI3D).
METHODS: The study included 20 patients who underwent elective AAA surgery. CICS and CI3D were measured simultaneously at eight time points during the surgery. Trending ability was investigated after aortic clamping and unclamping. We used CI3D as the reference method.
RESULTS: Bland-Altman analysis showed a wide limit of agreement between CICS and CI3D (percentage error 41.3%). Subgroup analysis showed a lower percentage error (33.2%) in patients with CI ≥ 2.5 L/min/m2. The cubic splines related to the CI3D and CI discrepancy were negatively sloped, indicating that CI3D had significant influence on the CI discrepancy (p < 0.001). Four-quadrant plot analysis showed that the tracking ability of ClearSight™ after aortic clamping and declamping were clinically unacceptable (81.3% and 78.6%, respectively). Also, the polar plot analysis showed that the concordance rate of ClearSight™ after aortic clamping and declamping were clinically unacceptable (58.3% and 66.7%, respectively).
CONCLUSIONS: ClearSight™ was not sufficiently accurate in patients undergoing AAA surgery. The tracking ability of ClearSight™ after aortic clamping was below the acceptable limit.

Entities:  

Keywords:  Blood pressure monitor; Cardiac output; ClearSight; Echocardiography; Ultrasound

Mesh:

Year:  2019        PMID: 30904953     DOI: 10.1007/s00540-019-02632-6

Source DB:  PubMed          Journal:  J Anesth        ISSN: 0913-8668            Impact factor:   2.078


  28 in total

1.  Recommendations for evaluation of the severity of native valvular regurgitation with two-dimensional and Doppler echocardiography.

Authors:  William A Zoghbi; Maurice Enriquez-Sarano; Elyse Foster; Paul A Grayburn; Carol D Kraft; Robert A Levine; Petros Nihoyannopoulos; Catherine M Otto; Miguel A Quinones; Harry Rakowski; William J Stewart; Alan Waggoner; Neil J Weissman
Journal:  J Am Soc Echocardiogr       Date:  2003-07       Impact factor: 5.251

Review 2.  A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output.

Authors:  Lester A Critchley; Anna Lee; Anthony M-H Ho
Journal:  Anesth Analg       Date:  2010-08-24       Impact factor: 5.108

3.  Real-time 3-dimensional echocardiographic quantification of left ventricular volumes: multicenter study for validation with magnetic resonance imaging and investigation of sources of error.

Authors:  Victor Mor-Avi; Carly Jenkins; Harald P Kühl; Hans-Joachim Nesser; Thomas Marwick; Andreas Franke; Christian Ebner; Benjamin H Freed; Regina Steringer-Mascherbauer; Heidi Pollard; Lynn Weinert; Johannes Niel; Lissa Sugeng; Roberto M Lang
Journal:  JACC Cardiovasc Imaging       Date:  2008-07

Review 4.  Real-time 3-dimensional echocardiography: an integral component of the routine echocardiographic examination in adult patients?

Authors:  Victor Mor-Avi; Lissa Sugeng; Roberto M Lang
Journal:  Circulation       Date:  2009-01-20       Impact factor: 29.690

5.  Assessment of trending ability of cardiac output monitors by polar plot methodology.

Authors:  Lester A Critchley; Xiao X Yang; Anna Lee
Journal:  J Cardiothorac Vasc Anesth       Date:  2011-03-17       Impact factor: 2.628

6.  Comparison of noninvasive cardiac output measurements using the Nexfin monitoring device and the esophageal Doppler.

Authors:  Guo Chen; Lingzhong Meng; Brenton Alexander; Nam Phuong Tran; Zeev N Kain; Maxime Cannesson
Journal:  J Clin Anesth       Date:  2012-06       Impact factor: 9.452

7.  Clinical comparison of an echocardiograph-derived versus pulse counter-derived cardiac output measurement in abdominal aortic aneurysm surgery.

Authors:  Yusuke Kusaka; Kenji Yoshitani; Tomoya Irie; Yuzuru Inatomi; Masahide Shinzawa; Yoshihiko Ohnishi
Journal:  J Cardiothorac Vasc Anesth       Date:  2011-09-15       Impact factor: 2.628

8.  Acute endothelin A receptor blockade causes selective pulmonary vasodilation in patients with chronic heart failure.

Authors:  M M Givertz; W S Colucci; T H LeJemtel; S S Gottlieb; J M Hare; M T Slawsky; C V Leier; E Loh; J M Nicklas; B E Lewis
Journal:  Circulation       Date:  2000-06-27       Impact factor: 29.690

9.  Comparison of left ventricular stroke volume assessment by two- and three-dimensional echocardiography in a swine model of acute myocardial infarction validated by thermodilution method.

Authors:  Yuichi J Shimada; Kiyotake Ishikawa; Yoshiaki Kawase; Dennis Ladage; Lisa Tilemann; Takahiro Shiota; Roger J Hajjar
Journal:  Echocardiography       Date:  2012-06-05       Impact factor: 1.724

10.  Reproducibility and accuracy of echocardiographic measurements of left ventricular parameters using real-time three-dimensional echocardiography.

Authors:  Carly Jenkins; Kristen Bricknell; Lizelle Hanekom; Thomas H Marwick
Journal:  J Am Coll Cardiol       Date:  2004-08-18       Impact factor: 24.094

View more
  6 in total

1.  Update on the assessment of fluid responsiveness.

Authors:  Koichi Suehiro
Journal:  J Anesth       Date:  2020-01-07       Impact factor: 2.078

2.  Anesthetic management of modified electroconvulsive therapy for a patient with coronary aneurysms: a case report.

Authors:  Riho Nakayama; Takuya Yoshida; Norihiko Obata; Satoshi Mizobuchi
Journal:  JA Clin Rep       Date:  2019-11-22

3.  Continuous Non-Invasive Arterial Pressure Monitoring (ClearSight System) and Ankle Blood Pressure Measurements as Alternatives to Conventional Arm Blood Pressure.

Authors:  Seohee Lee; Jaeyeon Chung; Jinyoung Bae; Youn Joung Cho; Karam Nam; Yunseok Jeon
Journal:  J Clin Med       Date:  2020-11-10       Impact factor: 4.241

4.  Accuracy of noninvasive continuous arterial pressure monitoring using ClearSight during one-lung ventilation.

Authors:  Sang-Wook Lee; Sangho Lee; Hyungtae Kim; Yun-Jong Kim; Mihyeon Kim; Jeong-Hyun Choi
Journal:  Medicine (Baltimore)       Date:  2021-03-19       Impact factor: 1.817

5.  Noninvasive cardiac output measurements are inaccurate in patients with severe aortic valve stenosis undergoing transcatheter aortic valve implantation.

Authors:  Musashi Yahagi; Kyuma Omi; Takuma Maeda
Journal:  Korean J Anesthesiol       Date:  2021-10-20

6.  Non-invasive continuous blood pressure monitoring using the ClearSight system for pregnant women at high risks of post-partum hemorrhage: comparison with invasive blood pressure monitoring during cesarean section.

Authors:  Takuya Misugi; Takashi Juri; Koichi Suehiro; Kohei Kitada; Yasushi Kurihara; Mie Tahara; Akihiro Hamuro; Akemi Nakano; Masayasu Koyama; Takasi Mori; Daisuke Tachibana
Journal:  Obstet Gynecol Sci       Date:  2022-06-27
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.