Developing a database of 3-D scattered radiation distributions for a c-arm fluoroscope as a function of exposure parameters and phantom.

Chao Guo1,2, Zhenyu Xiong1,2, Sarath Vijayan1,2, Stephen Rudin1,2,3, Daniel R Bednarek1,2,3.   

Abstract

The purpose of this work is to develop a database of 3D scattered radiation dose-rate distributions to estimate the staff dose by location around a C-Arm fluoroscopic system in an interventional procedure room. The primary x-ray beam of a Toshiba Infinix fluoroscopy machine was modeled using EGSnrc Monte Carlo code and the scattered radiation distributions were calculated using 5 × 109 photons per simulation. These 3D distributions were determined over the volume of the room as a function of various parameters such as the beam kVp and beam filter, the size and shape of the field, the angulation of the C-arm, and the phantom size and shape. Two phantom shapes were used in this study: cylindrical and super-ellipses. The results show that shape of the phantom will affect the dose-rate distribution at distances less than 100 cm, with a higher intensity for the super-ellipse. The scatter intensity per entrance air kerma is seen to be approximately proportional to field area and to increase with increasing kVp. The scatter changes proportionally with increases in primary entrance air kerma for factors such as pulse rate, mA and pulse width. This database will allow estimation of the scatter distribution in the procedure room and, when displayed to the staff during a procedure, may facilitate a reduction of occupational dose.

Entities:  

Year:  2018        PMID: 29899589      PMCID: PMC5994753          DOI: 10.1117/12.2293788

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  2 in total

1.  A tracking system to calculate patient skin dose in real-time during neurointerventional procedures using a biplane x-ray imaging system.

Authors:  V K Rana; S Rudin; D R Bednarek
Journal:  Med Phys       Date:  2016-09       Impact factor: 4.071

2.  Patient dose map indications on interventional X-ray systems and validation with Gafchromic XR-RV3 film.

Authors:  C Bordier; R Klausz; L Desponds
Journal:  Radiat Prot Dosimetry       Date:  2014-06-17       Impact factor: 0.972

  2 in total
  1 in total

1.  Evaluation of Methods of Displaying the Real-Time Scattered Radiation Distribution during Fluoroscopically-Guided Interventions for Staff Dose Reduction.

Authors:  J Kilian-Meneghin; Z Xiong; C Guo; S Rudin; D R Bednarek
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2018-03-09
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.