Estimation of Patient Eye-Lens Dose During Neuro-Interventional Procedures using a Dense Neural Network (DNN).

J Collins1, S Sun1, C Guo1, A Podgorsak1, S Rudin1, D R Bednarek1.   

Abstract

The patient's eye-lens dose changes for each projection view during fluoroscopically-guided neuro-interventional procedures. Monte-Carlo (MC) simulation can be done to estimate lens dose but MC cannot be done in real-time to give feedback to the interventionalist. Deep learning (DL) models were investigated to estimate patient-lens dose for given exposure conditions to give real-time updates. MC simulations were done using a Zubal computational phantom to create a dataset of eye-lens dose values for training the DL models. Six geometric parameters (entrance-field size, LAO gantry angulation, patient x, y, z head position relative to the beam isocenter, and whether patient's right or left eye) were varied for the simulations. The dose for each combination of parameters was expressed as lens dose per entrance air kerma (mGy/Gy). Geometric parameter combinations associated with high-dose values were sampled more finely to generate more high-dose values for training purposes. Additionally, dose at intermediate parameter values was calculated by MC in order to validate the interpolation capabilities of DL. Data was split into training, validation and testing sets. Stacked models and median algorithms were implemented to create more robust models. Model performance was evaluated using mean absolute percentage error (MAPE). The goal for this DL model is that it be implemented into the Dose Tracking System (DTS) developed by our group. This would allow the DTS to infer the patient's eye-lens dose for real-time feedback and eliminate the need for a large database of pre-calculated values with interpolation capabilities.

Entities:  

Keywords:  DNN; DTS; Eye-Lens Dose; Neuro-Interventional Procedures

Year:  2021        PMID: 34334873      PMCID: PMC8323862          DOI: 10.1117/12.2580723

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  4 in total

1.  Verification of the performance accuracy of a real-time skin-dose tracking system for interventional fluoroscopic procedures.

Authors:  Daniel R Bednarek; Jeffery Barbarits; Vijay K Rana; Srikanta P Nagaraja; Madhur S Josan; Stephen Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2011-02-13

2.  A tracking system to calculate patient skin dose in real-time during neurointerventional procedures using a biplane x-ray imaging system.

Authors:  V K Rana; S Rudin; D R Bednarek
Journal:  Med Phys       Date:  2016-09       Impact factor: 4.071

3.  Deterministic Effects to the Lens of the Eye Following Ionizing Radiation Exposure: is There Evidence to Support a Reduction in Threshold Dose?

Authors:  Christopher Thome; Douglas B Chambers; Antony M Hooker; Jeroen W Thompson; Douglas R Boreham
Journal:  Health Phys       Date:  2018-03       Impact factor: 1.316

4.  Radiation Doses in Patient Eye Lenses during Interventional Neuroradiology Procedures.

Authors:  R M Sánchez; E Vañó; J M Fernández; S Rosati; L López-Ibor
Journal:  AJNR Am J Neuroradiol       Date:  2015-11-05       Impact factor: 3.825

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.