Literature DB >> 29896882

Organ doses in pediatric patients undergoing cardiac-centered fluoroscopically guided interventions: Comparison of three methods for computational phantom alignment.

Emily L Marshall1, David Borrego2, James C Fudge3, Dhanashree Rajderkar4, Wesley E Bolch1.   

Abstract

PURPOSE: To assess various computational phantom alignment techniques within Monte Carlo radiation transport models of pediatric fluoroscopically guided cardiac interventional studies.
METHODS: Logfiles, including all procedure radiation and machine data, were extracted from a Toshiba infinix-I unit in the University of Florida Pediatric Catheterization Laboratory for a cohort of 10 patients. Two different alignment methods were then tested against a ground truth standard based upon identification of a unique anatomic reference point within images co-registered to specific irradiation events within each procedure. The first alignment method required measurement of the distance from the edge of the exam table to the top of the patient's head (table alignment method). The second alignment method fixed the anatomic reference point to be the geometric center of the heart muscle, as all 10 studies were cardiac in nature. Monte Carlo radiation transport simulations were performed for each patient and intervention using morphometry-matched hybrid computational phantoms for the reference and two tested alignment methods. For each combination, absorbed doses were computed for 28 organs and root mean square organ doses were assessed and compared across the alignment methods.
RESULTS: The percent error in root mean square organ dose ranged from -57% to +41% for the table alignment method, and from -27% to +22% for the heart geometric centroid alignment method. Absorbed doses to specific organs, such as the heart and lungs, demonstrated higher accuracy in the heart geometric centroid alignment method, with average percent errors of 10% and 1.4%, respectively, compared to average percent errors of -32% and 24%, respectively, using the table alignment method.
CONCLUSIONS: Of the two phantom alignment methods investigated in this study, the use of an anatomical reference point - in this case the geometric centroid of the heart - provided a reliable method for radiation transport simulations of organ dose in pediatric interventional cardiac studies. This alignment method provides the added benefit of requiring no physician input, making retrospective calculations possible. Moving forward, additional anatomical reference methods can be tested to assess the reliability of anatomical reference points beyond cardiac centered procedures.
© 2018 American Association of Physicists in Medicine.

Entities:  

Keywords:  cardiac catheterization; computational hybrid phantoms; fluoroscopically guided interventions; organ doses; pediatric patients

Year:  2018        PMID: 29896882      PMCID: PMC6292782          DOI: 10.1002/mp.13044

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  26 in total

1.  Skin dose mapping for fluoroscopically guided interventions.

Authors:  Perry B Johnson; David Borrego; Stephen Balter; Kevin Johnson; Daniel Siragusa; Wesley E Bolch
Journal:  Med Phys       Date:  2011-10       Impact factor: 4.071

2.  Simultaneous determination of equivalent dose to organs and tissues of the patient and of the physician in interventional radiology using the Monte Carlo method.

Authors:  A Bozkurt; D Bor
Journal:  Phys Med Biol       Date:  2006-12-21       Impact factor: 3.609

3.  The UF/NCI family of hybrid computational phantoms representing the current US population of male and female children, adolescents, and adults--application to CT dosimetry.

Authors:  Amy M Geyer; Shannon O'Reilly; Choonsik Lee; Daniel J Long; Wesley E Bolch
Journal:  Phys Med Biol       Date:  2014-08-21       Impact factor: 3.609

4.  A tracking system to calculate patient skin dose in real-time during neurointerventional procedures using a biplane x-ray imaging system.

Authors:  V K Rana; S Rudin; D R Bednarek
Journal:  Med Phys       Date:  2016-09       Impact factor: 4.071

5.  Standing adult human phantoms based on 10th, 50th and 90th mass and height percentiles of male and female Caucasian populations.

Authors:  V F Cassola; F M Milian; R Kramer; C A B de Oliveira Lira; H J Khoury
Journal:  Phys Med Biol       Date:  2011-05-31       Impact factor: 3.609

6.  Automatic monitoring of localized skin dose with fluoroscopic and interventional procedures.

Authors:  Yasaman Khodadadegan; Muhong Zhang; William Pavlicek; Robert G Paden; Brian Chong; Beth A Schueler; Kenneth A Fetterly; Steve G Langer; Teresa Wu
Journal:  J Digit Imaging       Date:  2011-08       Impact factor: 4.056

7.  Extension of RPI-adult male and female computational phantoms to obese patients and a Monte Carlo study of the effect on CT imaging dose.

Authors:  Aiping Ding; Matthew M Mille; Tianyu Liu; Peter F Caracappa; X George Xu
Journal:  Phys Med Biol       Date:  2012-04-05       Impact factor: 3.609

8.  History of right heart catheterization: 100 years of experimentation and methodology development.

Authors:  Bobby D Nossaman; Brittni A Scruggs; Vaughn E Nossaman; Subramanyam N Murthy; Philip J Kadowitz
Journal:  Cardiol Rev       Date:  2010 Mar-Apr       Impact factor: 2.644

9.  Basic anatomical and physiological data for use in radiological protection: reference values. A report of age- and gender-related differences in the anatomical and physiological characteristics of reference individuals. ICRP Publication 89.

Authors: 
Journal:  Ann ICRP       Date:  2002

10.  Local organ dose conversion coefficients for angiographic examinations of coronary arteries.

Authors:  H Schlattl; M Zankl; J Hausleiter; C Hoeschen
Journal:  Phys Med Biol       Date:  2007-06-26       Impact factor: 3.609

View more
  1 in total

1.  Pitfalls in interventional X-ray organ dose assessment-combined experimental and computational phantom study: application to prostatic artery embolization.

Authors:  Philipp Roser; Annette Birkhold; Xia Zhong; Philipp Ochs; Elizaveta Stepina; Markus Kowarschik; Rebecca Fahrig; Andreas Maier
Journal:  Int J Comput Assist Radiol Surg       Date:  2019-08-03       Impact factor: 2.924

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.