| Literature DB >> 27547749 |
In-Hee Woo1, Ju-Won Kim1, So-Young Kang2, Young-Hee Kim3, Byoung-Eun Yang1.
Abstract
BACKGROUND: The objective of this retrospective study was to show results from platform-switched narrow-diameter implants in the posterior edentulous region, which we followed up for more than 1 year after functional loading.Entities:
Keywords: Conical dental implant-abutment connection; Dental implant platform switching; Narrow-diameter implants; Posterior edentulous
Year: 2016 PMID: 27547749 PMCID: PMC4974261 DOI: 10.1186/s40902-016-0077-x
Source DB: PubMed Journal: Maxillofac Plast Reconstr Surg ISSN: 2288-8101
Fig. 1Crestal bone level after implant placement (a), 3 months after placement (b), after prosthetic restoration (c), and after 4-year follow-up (d)
Length distribution
| Length (mm) | Number |
|---|---|
| 8.0 | 10 |
| 9.5 | 29 |
| 11.0 | 59 |
Distribution of implant positions
| Sum | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| Number of implants | 1 | 7 | 12 | 7 | 8 | 5 | 2 | 42 | |
| Tooth position | 17 | 16 | 15 | 14 | 24 | 25 | 26 | 27 | |
| 47 | 46 | 45 | 44 | 34 | 35 | 36 | 37 | ||
| Number of implants | 3 | 12 | 9 | 6 | 8 | 8 | 8 | 2 | 56 |
Comparison of bone loss and PTV by variable
| Number of implants | Bone loss mean (SD) |
| PTV mean (SD) |
| |
|---|---|---|---|---|---|
| Bone quality | 0.967 | <0.001 | |||
| Soft | 13 | 0.15 mm (0.376) | −2.323 (0.208)a | ||
| Normal | 37 | 0.15 mm (0.423) | −3.273 (0.266)b | ||
| Hard | 48 | 0.13 mm (0.375) | −3.562 (0.335)c | ||
| Location | 0.402 | <0.001 | |||
| Maxilla | 42 | 0.18 mm (0.439) | −2.979 (0.514) | ||
| Mandible | 56 | 0.11 mm (0.351) | −3.522 (0.335) | ||
| Loading period | 0.715 | 0.014 | |||
| Over 1 year | 22 | 0.08 mm (0.282) | −3.125 (0.518)a | ||
| Over 2 years | 19 | 0.21 mm (0.535) | −3.095 (0.424)a | ||
| Over 3 years | 34 | 0.16 mm (0.439) | −3.409 (0.452)b | ||
| Over 4 years | 23 | 0.11 mm (0.259) | −3.465 (0.508)b | ||
| Splinting status | 0.407 | 0.968 | |||
| Yes | 68 | 0.16 mm (0.418) | −3.295 (0.519) | ||
| No | 30 | 0.08 mm (0.245) | −3.290 (0.406) | ||
| Reason for tooth loss | 0.314 | 0.559 | |||
| Periodontitis | 89 | 0.13 mm (0.384) | −3.285 (0.514) | ||
| Caries | 9 | 0.29 mm (0.488) | −3.400 (0.224) |
T test or general linear model
SD standard deviation, PTV Periotest value
a, b, cResults from multiple comparisons using the general linear model
Association between variables and PTV using the regression model
| Variable | Univariate model | Multivariate model | ||||||
|---|---|---|---|---|---|---|---|---|
| Parameter | Standard |
| Pr > | | Parameter | Standard |
| Pr > | | |
| Estimate | Error | Estimate | Error | |||||
| Bone quality | −0.528 | 0.047 | −11.213 | 0.000 | −0.669 | 0.084 | −7.969 | 0.000 |
| Jaw location | −0.544 | 0.085 | −6.401 | 0.000 | 0.089 | 0.158 | 0.565 | 0.574 |
| Loading period | −0.133 | 0.044 | −3.046 | 0.003 | −0.085 | 0.030 | −2.835 | 0.006 |
| Splinting status | −0.005 | 0.125 | −0.040 | 0.968 | 0.036 | 0.087 | 0.417 | 0.678 |
| Reason for tooth loss | −0.094 | 0.156 | −0.600 | 0.550 | 0.078 | 0.140 | 0.554 | 0.581 |
| Sex | −0.101 | 0.100 | −1.016 | 0.312 | −0.080 | 0.073 | −1.106 | 0.272 |
| Age | −0.001 | 0.004 | −0.348 | 0.729 | 0.002 | 0.003 | 0.679 | 0.499 |
| Implant length | 0.078 | 0.049 | 1.601 | 0.113 | −0.012 | 0.036 | −0.348 | 0.729 |
In the regression model and multivariate model, dependent variables were corrected for multiple comparisons
Association between variables and bone loss using regression model
| Variable | Univariate model | Multivariate model | ||||||
|---|---|---|---|---|---|---|---|---|
| Parameter | Standard |
| Pr > | | Parameter | Standard |
| Pr > | | |
| Estimate | Error | Estimate | Error | |||||
| Bone quality | 0.014 | 0.056 | 0.247 | 0.805 | −0.027 | 0.102 | −0.269 | 0.789 |
| Jaw location | 0.067 | 0.079 | 0.841 | 0.402 | 0.432 | 0.192 | 2.257 | 0.026 |
| Loading period | −0.006 | 0.036 | −0.155 | 0.877 | −0.01 | 0.036 | −0.278 | 0.782 |
| Splinting status | −0.081 | 0.098 | −0.833 | 0.407 | −0.172 | 0.106 | −1.631 | 0.106 |
| Reason for tooth loss | −0.073 | 0.123 | −0.593 | 0.555 | 0.017 | 0.17 | 0.1 | 0.921 |
| Sex | 0.077 | 0.078 | 0.987 | 0.326 | 0.082 | 0.088 | 0.933 | 0.353 |
| Age | 0.005 | 0.003 | 1.676 | 0.097 | 0.005 | 0.004 | 1.42 | 0.159 |
| Implant length | 0.009 | 0.039 | 0.244 | 0.808 | 0.004 | 0.043 | 0.089 | 0.93 |
Fig. 2Illustration of an Ankylos® dental implant with a 3.5-mm diameter which was installed in mandibular premolar (a) and molar area (b). The integrated platform-switching design provides more space for soft tissue retention. The illustration was a courtesy of the Graphic Designer, Ernesto Pacheco (pachecojake@mac.com)