Literature DB >> 19095473

A comparative study of two noninvasive techniques to evaluate implant stability: Periotest and Osstell Mentor.

Ji-Su Oh1, Su-Gwan Kim, Sung-Chul Lim, Joo L Ong.   

Abstract

OBJECTIVES: The purpose of this study was to examine the usefulness of Periotest and Osstell Mentor (a resonance frequency analysis) as nondestructive analytical tools for determining implant stability in clinics, to evaluate the precision of both instruments, and to determine the applicability of these measured values as clinically relevant indices. STUDY
DESIGN: Four adult mongrel dogs weighing about 12 to 15 kg were used in this study. Bilateral extractions of the first through the fourth mandibular and maxillary premolars were performed and a total of 48 commercially pure titanium screw implants (USII Plus; OSTEM Implant, Seoul, Republic of Korea) were placed at 4 weeks after extraction. All implants (10 mm length and 3.3 mm diameter) were self-tapping and surface-treated with resorbable blast media (RBM). Periotest values (PTVs) obtained from Periotest, and Implant Stability Quotient (ISQ) obtained from Osstell Mentor, were measured at the time of implantation, and 3 and 6 weeks after implantation. At 3 and 6 weeks after implantation, 4 dogs were humanely killed and histomorphometric analysis was performed. The new peri-implant bone formation rate (NBFR) was measured.
RESULTS: The PTV value was lower and ISQ value was higher at 6 weeks when compared with data collected at 3 weeks after implantation. The PTVs of the maxilla were higher than the mandible and the ISQ values of the maxilla were lower than the mandible. Based on the NBFR, the 6-week group showed higher bone formation when compared to the 3-week group, correlating to the observed PTV and ISQ values. Additionally, the NBFR was higher in the maxilla than the mandible. No significant difference between PTV and ISQ was also observed when PTV and ISQ were compared to NBFR.
CONCLUSION: The results indicated that the Periotest and Osstell Mentor, both noninvasive diagnostic devices, were useful and comparably reliable, showing a strong association with each other in assessing implant stability.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 19095473     DOI: 10.1016/j.tripleo.2008.08.026

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod        ISSN: 1079-2104


  13 in total

1.  Assessment of damping capacity as an index of root proximity in self-drilling orthodontic mini-implants.

Authors:  Mitsuru Motoyoshi; Yasuki Uchida; Miho Matsuoka; Mizuki Inaba; Hiroaki Iwai; Yoko Karasawa; Noriyoshi Shimizu
Journal:  Clin Oral Investig       Date:  2013-03-22       Impact factor: 3.573

2.  Highly Nonlinear Solitary Waves for the Assessment of Dental Implant Mobility.

Authors:  Bruk Berhanu; Piervincenzo Rizzo; Mark Ochs
Journal:  J Appl Mech       Date:  2012-11-20       Impact factor: 2.168

3.  Low-level laser therapy with 940 nm diode laser on stability of dental implants: a randomized controlled clinical trial.

Authors:  Parviz Torkzaban; Shahin Kasraei; Sara Torabi; Maryam Farhadian
Journal:  Lasers Med Sci       Date:  2017-10-29       Impact factor: 3.161

4.  Vertical Bone Implant Contact Around Anterior Immediate Implants and Their Stability After Using Either Alloplast or L-PRF or Both in Peri-Implant Gap: A Prospective Randomized Trial.

Authors:  Md Shahbaz Alam; Arjun Dhiman; Veena Jain; Ongkila Bhutia; Gunjan Pruthi
Journal:  J Maxillofac Oral Surg       Date:  2020-07-28

5.  Effect of decreased implant healing time on bone (re)modeling adjacent to plateaued implants under functional loading in a dog model.

Authors:  Jing Dai; Ning Cheng; Richard J Miron; Bin Shi; Xiangrong Cheng; Yufeng Zhang
Journal:  Clin Oral Investig       Date:  2013-02-03       Impact factor: 3.573

6.  In vivo evaluation of quantitative percussion diagnostics for determining implant stability.

Authors:  Lindsey R VanSchoiack; Veronica I Shubayev; Robert R Myers; Cherilyn G Sheets; James C Earthman
Journal:  Int J Oral Maxillofac Implants       Date:  2013 Sep-Oct       Impact factor: 2.804

Review 7.  Does the Implant Surgical Technique Affect the Primary and/or Secondary Stability of Dental Implants? A Systematic Review.

Authors:  Rola Muhammed Shadid; Nasrin Rushdi Sadaqah; Sahar Abdo Othman
Journal:  Int J Dent       Date:  2014-07-07

8.  Monitoring the Changes of Material Properties at Bone-Implant Interface during the Healing Process In Vivo: A Viscoelastic Investigation.

Authors:  Hsiang-Ho Chen; Wei-Yi Lai; Tze-Jian Chee; Ya-Hui Chan; Sheng-Wei Feng
Journal:  Biomed Res Int       Date:  2017-03-08       Impact factor: 3.411

9.  Narrow-diameter implants with conical connection for restoring the posterior edentulous region.

Authors:  In-Hee Woo; Ju-Won Kim; So-Young Kang; Young-Hee Kim; Byoung-Eun Yang
Journal:  Maxillofac Plast Reconstr Surg       Date:  2016-08-05

Review 10.  Clinical Assessment of Dental Implant Stability During Follow-Up: What Is Actually Measured, and Perspectives.

Authors:  Elisabetta M Zanetti; Giulia Pascoletti; Michele Calì; Cristina Bignardi; Giordano Franceschini
Journal:  Biosensors (Basel)       Date:  2018-07-13
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.