Fatigue resistance of two implant/abutment joint designs.

Ameen Khraisat1, Roxana Stegaroiu, Shuichi Nomura, Osamu Miyakawa.   

Abstract

STATEMENT OF PROBLEM: Because of reported mechanical failures, alternative implant system components with suggested optimized strength have been manufactured. However, the endurance of these products has not been well investigated.
PURPOSE: This study was designed to assess the effect of joint design on the fatigue strength and failure mode of 2 single-tooth implant systems: Brånemark and ITI, in which a hex mediated-butt joint and 8-degree internal conical implant/abutment interface are used, respectively.
MATERIALS AND METHODS: Seven 10-mm implants from each implant system were embedded to a depth of 7 mm in cylindrical acrylic resin blocks. CeraOne and Solid abutments with cement-retained castings were assembled to the Brånemark and ITI implants, respectively. The assembled units were mounted in a lever-type-testing machine that was equipped with an automatic counting device and shutoff sensors, enabling the recording of the number of cycles till failure. A cyclic load of 100 N was applied perpendicular to the long axis of the assemblies at a rate of 75 cycles/min. To investigate specimen resistance to fatigue during 6 years of simulated function, a target of 1,800,000 cycles was defined. Specimen preparation and testing was performed by the same operator. The association of the joint design with the occurrence of failure was verified by Fisher's exact probability test (P<.05).
RESULTS: For the Brånemark group, the gold alloy abutment screw in all specimens fractured between 1,178,023 and 1,733,526 cycles with a standard deviation of 224,477 cycles. For the ITI group, all specimens had no failure until 1,800,000 cycles. Statistical analysis showed a highly significant difference between the 2 groups (P=.000582).
CONCLUSIONS: Within the limitations of this in vitro study, the effect of joint design on the fatigue strength and failure mode of the ITI single tooth implant system was significantly better (P>.001) than the Brånemark single-tooth implant system tested.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12488853     DOI: 10.1067/mpr.2002.129384

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  17 in total

1.  External versus internal abutment connection implants: a survey of opinions and decision making among experienced implant dentists in Japan.

Authors:  Yoshiyuki Hagiwara; Alan B Carr
Journal:  Odontology       Date:  2013-12-17       Impact factor: 2.634

2.  Mechanical analysis of conventional and small diameter conical implant abutments.

Authors:  Izabela Cristina Maurício Moris; Adriana Cláudia Lapria Faria; Maria da Gloria Chiarello de Mattos; Ricardo Faria Ribeiro; Renata Cristina Silveira Rodrigues
Journal:  J Adv Prosthodont       Date:  2012-08-28       Impact factor: 1.904

3.  Implant-abutment connections: influence of the design on the microgap and their fatigue and fracture behavior of dental implants.

Authors:  F J Gil; M Herrero-Climent; P Lázaro; J V Rios
Journal:  J Mater Sci Mater Med       Date:  2014-04-10       Impact factor: 3.896

4.  The change of rotational freedom following different insertion torques in three implant systems with implant driver.

Authors:  Joo-Hyun Kwon; Chong-Hyun Han; Sun-Jai Kim; Jae-Seung Chang
Journal:  J Adv Prosthodont       Date:  2009-03-31       Impact factor: 1.904

5.  Reliability and failure modes of narrow implant systems.

Authors:  Ronaldo Hirata; Estevam A Bonfante; Rodolfo B Anchieta; Lucas S Machado; Gileade Freitas; Vinicius P Fardin; Nick Tovar; Paulo G Coelho
Journal:  Clin Oral Investig       Date:  2015-11-03       Impact factor: 3.573

6.  Sealing Capability and SEM Observation of the Implant-Abutment Interface.

Authors:  Fabio C Lorenzoni; Paulo G Coelho; Gerson Bonfante; Ricardo M Carvalho; Nelson R F A Silva; Marcelo Suzuki; Thelma Lopes Silva; Estevam A Bonfante
Journal:  Int J Dent       Date:  2011-07-02

7.  Leakage evaluation of original and compatible implant-abutment connections: In vitro study using Rhodamine B.

Authors:  Antoine Berberi; George Tehini; Khaldoun Rifai; Farah Bou Nasser Eddine; Bassam Badran; Haidar Akl
Journal:  J Dent Biomech       Date:  2014-08-11

8.  Effect of tightening torque on the marginal adaptation of cement-retained implant-supported fixed dental prostheses.

Authors:  Jalil Ghanbarzadeh; Hossin Dashti; Reza Karamad; Marzieh Alikhasi; Mohammadreza Nakhaei
Journal:  Dent Res J (Isfahan)       Date:  2015 Jul-Aug

9.  In vitro evaluation of leakage at implant-abutment connection of three implant systems having the same prosthetic interface using rhodamine B.

Authors:  Antoine Berberi; Georges Tehini; Khaldoun Rifai; Farah Bou Nasser Eddine; Nabil El Zein; Bassam Badran; Haidar Akl
Journal:  Int J Dent       Date:  2014-05-11

10.  Narrow-diameter implants with conical connection for restoring the posterior edentulous region.

Authors:  In-Hee Woo; Ju-Won Kim; So-Young Kang; Young-Hee Kim; Byoung-Eun Yang
Journal:  Maxillofac Plast Reconstr Surg       Date:  2016-08-05
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.