Correlation between resonance frequency, insertion torque and bone-implant contact in self-cutting threaded implants.

Yahya Açil1, Jan Sievers2, Aydin Gülses3, Mustafa Ayna4, Jörg Wiltfang1, Hendrik Terheyden5.   

Abstract

The aim of this study was to evaluate the correlation between resonance frequency analysis (RFA) values and the histomorphometric bone-implant contact (BIC) immediately after insertion of the implant. Additionally, it was examined to define which extent peak insertion torque (PIT) was correlated with the latter measurements. 15 Xive® S plus root from dental implants were inserted in fresh porcine frontal bones. The insertion torque was measured with the Kavo Intrasurg 300 surgical unit. RFA connector was coupled to the implants and the mean value of 20 consecutive RFA measurements was calculated via Osstell® ISQ device. The implants were removed with the adjacent bone with a band saw. The blocks were processed for undecalcified histology. Two perpendicular longitudinal middle sections of the implant were made and stained with toluidine blue and the BIC was assessed by histomorphometry. The correlation coefficient (Spearman) between RFA and BIC was R = 0.579 (p = 0.026, F test). The correlation between PIT and BIC (0.33, p > 0.05) and PIT and RFA (0.153, p > 0.05) was not statistically significant. The present data confirmed a moderate and statistically positive correlation between RFA and BIC. No correlation between BIC and PIT and PIT and RFA was observed. Further studies considering different bone qualities would be beneficial in understanding the relation between RFA and BIC.

Entities:  

Keywords:  Bone-implant contact; Correlation; Dental implants; Histomorphometric measurements; Immediate loading; Insertion torque; Primary stability; Resonance frequency analysis

Mesh:

Substances:

Year:  2016        PMID: 27538921     DOI: 10.1007/s10266-016-0265-2

Source DB:  PubMed          Journal:  Odontology        ISSN: 1618-1247            Impact factor:   2.634


  30 in total

1.  [Peri-implant bone formation around cylindrical and conical implant systems].

Authors:  A Büchter; J Kleinheinz; H-P Wiesmann; L Seper; U Joos; U Meyer
Journal:  Mund Kiefer Gesichtschir       Date:  2004-07-23

2.  Resonance frequency analysis of one-stage dental implant stability during the osseointegration period.

Authors:  S Ersanli; C Karabuda; F Beck; B Leblebicioglu
Journal:  J Periodontol       Date:  2005-07       Impact factor: 6.993

3.  Influence of factors related to implant stability detected by wireless resonance frequency analysis device.

Authors:  K Ohta; M Takechi; M Minami; H Shigeishi; M Hiraoka; M Nishimura; N Kamata
Journal:  J Oral Rehabil       Date:  2009-11-30       Impact factor: 3.837

4.  Influence of surgical technique and surface roughness on the primary stability of an implant in artificial bone with different cortical thickness: a laboratory study.

Authors:  Afsheen Tabassum; Gert J Meijer; Johannes G C Wolke; John A Jansen
Journal:  Clin Oral Implants Res       Date:  2010-02       Impact factor: 5.977

5.  Implant stability and bone density: assessment of correlation in fresh cadavers using conventional and osteotome implant sockets.

Authors:  Murat Cavit Cehreli; Ali Murat Kökat; Ayhan Comert; Murat Akkocaoğlu; Ibrahim Tekdemir; Kivanç Akça
Journal:  Clin Oral Implants Res       Date:  2009-07-20       Impact factor: 5.977

6.  The effect of static bone strain on implant stability and bone remodeling.

Authors:  Anders Halldin; Ryo Jimbo; Carina B Johansson; Ann Wennerberg; Magnus Jacobsson; Tomas Albrektsson; Stig Hansson
Journal:  Bone       Date:  2011-07-14       Impact factor: 4.398

7.  Resonance frequency analysis, insertion torque, and bone to implant contact of 4 implant surfaces: comparison and correlation study in sheep.

Authors:  Maroun Dagher; Nadim Mokbel; Gabriel Jabbour; Nada Naaman
Journal:  Implant Dent       Date:  2014-12       Impact factor: 2.454

8.  Resonance frequency measurements of implant stability in the dog mandible: experimental comparison with histomorphometric data.

Authors:  H Schliephake; A Sewing; A Aref
Journal:  Int J Oral Maxillofac Surg       Date:  2006-10-05       Impact factor: 2.789

9.  Implant stability and histomorphometry: a correlation study in human cadavers using stepped cylinder implants.

Authors:  Emeka Nkenke; Michael Hahn; Konstanze Weinzierl; Martin Radespiel-Tröger; Friedrich Wilhelm Neukam; Klaus Engelke
Journal:  Clin Oral Implants Res       Date:  2003-10       Impact factor: 5.977

10.  Torque-fitting and resonance frequency analyses of implants in conventional sockets versus controlled bone defects in vitro.

Authors:  K Akça; A M Kökat; A Cömert; M Akkocaoğlu; I Tekdemir; M C Cehreli
Journal:  Int J Oral Maxillofac Surg       Date:  2009-12-24       Impact factor: 2.789

View more
  7 in total

1.  Comparative evaluation of the stability of two different dental implant designs and surgical protocols-a pilot study.

Authors:  David E Simmons; Pooja Maney; Austin G Teitelbaum; Susan Billiot; Lomesh J Popat; A Archontia Palaiologou
Journal:  Int J Implant Dent       Date:  2017-05-02

2.  Influence of a Novel Surface of Bioactive Implants on Osseointegration: A Comparative and Histomorfometric Correlation and Implant Stability Study in Minipigs.

Authors:  Manuel M Romero-Ruiz; Francisco Javier Gil-Mur; José Vicente Ríos-Santos; Pedro Lázaro-Calvo; Blanca Ríos-Carrasco; Mariano Herrero-Climent
Journal:  Int J Mol Sci       Date:  2019-05-09       Impact factor: 5.923

3.  Comparison of implant stability measurements between a resonance frequency analysis device and a modified damping capacity analysis device: an in vitro study.

Authors:  Jungwon Lee; Se-Wook Pyo; Hyun-Jae Cho; Jung-Sub An; Jae-Hyun Lee; Ki-Tae Koo; Yong-Moo Lee
Journal:  J Periodontal Implant Sci       Date:  2020-02-18       Impact factor: 2.614

4.  Biomechanical Effects of a New Macrogeometry Design of Dental Implants: An In Vitro Experimental Analysis.

Authors:  Sergio Alexandre Gehrke; Leticia Pérez-Díaz; Patricia Mazón; Piedad N De Aza
Journal:  J Funct Biomater       Date:  2019-10-25

5.  Effects of the Healing Chambers in Implant Macrogeometry Design in a Low-Density Bone Using Conventional and Undersized Drilling.

Authors:  Sergio A Gehrke; Antônio Scarano; José H C de Lima; Marco A Bianchini; Berenice A Dedavid; Piedad N De Aza
Journal:  J Int Soc Prev Community Dent       Date:  2021-07-30

6.  Correlation between Insertion Torque and Implant Stability Quotient in Tapered Implants with Knife-Edge Thread Design.

Authors:  Domenico Baldi; Teresa Lombardi; Jacopo Colombo; Gabriele Cervino; Giuseppe Perinetti; Roberto Di Lenarda; Claudio Stacchi
Journal:  Biomed Res Int       Date:  2018-05-15       Impact factor: 3.411

7.  Relation Between Insertion Torque and Implant Stability Quotient: A Clinical Study.

Authors:  João Paulo do Vale Souza; Clóvis Lamartine de Moraes Melo Neto; Lucas Tavares Piacenza; Emily Vivianne Freitas da Silva; André Luiz de Melo Moreno; Paulo Augusto Penitente; Juliana Lujan Brunetto; Daniela Micheline Dos Santos; Marcelo Coelho Goiato
Journal:  Eur J Dent       Date:  2021-07-07
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.