Zhuo Liu1, Ye Sun, Zhuolu Zhang, Lei Chen, Nan Hong.
Abstract
Usually, coronary computed tomography angiography (CCTA) is performed during breath-holding to reduce artifact caused by respiration. The objective of this study was to evaluate the feasibility of free-breathing CCTA compared to breath-holding using CT scanner with wide detector. To evaluate the feasibility of CCTA during free-breathing using a 256-MDCT. In 80 patients who underwent CCTA, 40 were performed during breath-holding (group A), and the remaining 40 during free-breathing (group B). The quality scores for coronary arteries were analyzed and defined as: 3 (excellent), 2 (good), and 1 (poor). The image noise, signal-to-noise ratio and effective radiation dose as well as the heart rate variation were compared. The noise, signal-to-noise ratio, and effective radiation dose were not significantly different between the 2 groups. The mean heart rate variation between planning and scanning for group A was 7 ± 7.6 bpm, and larger than 3 ± 2.6 bpm for group B (P = 0.012). Quality scores of the free-breathing group were better than those of the breath-holding group (group A: 2.55 ± 0.64, group B: 2.85 ± 0.36, P = 0.018). Free-breathing CCTA is feasible on wide detector CT scanner to provide acceptable image quality with reduced heart rate variation and better images for certain patients.Entities:
Mesh:
Substances:
Year: 2016 PMID: 27399104 PMCID: PMC5058833 DOI: 10.1097/MD.0000000000004096
Source DB: PubMed Journal: Medicine (Baltimore) ISSN: 0025-7974 Impact factor: 1.889
Automatic scanning phase selection by auto gating based on heart rates.
Figure 1Subjective evaluation criteria (3: excellent; 2: good; and 1 poor).
Comparison results between group A and group B (P < 0.05 indicated statistically significance).
Figure 267-year-old female patient, free-breathing, heart rate during scanning: 52 bpm.
Figure 482-year-old female patient, breath-holding, mean heart rate during planning: 63 bpm, exposure phase 70%–80% R-R interval, heart rate during scanning: 82 bpm, quality score was 1.