Validity of linear measurements of the jaws using ultralow-dose MDCT and the iterative techniques of ASIR and MBIR.

Asma'a A Al-Ekrish1, Reema Al-Shawaf2, Peter Schullian3, Ra'ed Al-Sadhan2, Romed Hörmann4, Gerlig Widmann3.   

Abstract

PURPOSE: To assess the comparability of linear measurements of dental implant sites recorded from multidetector computed tomography (MDCT) images obtained using standard-dose filtered backprojection (FBP) technique with those from various ultralow doses combined with FBP, adaptive statistical iterative reconstruction (ASIR), and model-based iterative reconstruction (MBIR) techniques. The results of the study may contribute to MDCT dose optimization for dental implant site imaging.
METHODS: MDCT scans of two cadavers were acquired using a standard reference protocol and four ultralow-dose test protocols (TP). The volume CT dose index of the different dose protocols ranged from a maximum of 30.48-36.71 mGy to a minimum of 0.44-0.53 mGy. All scans were reconstructed using FBP, ASIR-50, ASIR-100, and MBIR, and either a bone or standard reconstruction kernel. Linear measurements were recorded from standardized images of the jaws by two examiners. Intra- and inter-examiner reliability of the measurements were analyzed using Cronbach's alpha and inter-item correlation. Agreement between the measurements obtained with the reference-dose/FBP protocol and each of the test protocols was determined with Bland-Altman plots and linear regression. Statistical significance was set at a P-value of 0.05.
RESULTS: No systematic variation was found between the linear measurements obtained with the reference protocol and the other imaging protocols. The only exceptions were TP3/ASIR-50 (bone kernel) and TP4/ASIR-100 (bone and standard kernels). The mean measurement differences between these three protocols and the reference protocol were within ±0.1 mm, with the 95 % confidence interval limits being within the range of ±1.15 mm.
CONCLUSIONS: A nearly 97.5 % reduction in dose did not significantly affect the height and width measurements of edentulous jaws regardless of the reconstruction algorithm used.

Entities:  

Keywords:  Algorithms; Computer-assisted; Dental implants; Image processing; Multidetector computed tomography; Radiation dosage; Reproducibility of results

Mesh:

Substances:

Year:  2016        PMID: 27255542     DOI: 10.1007/s11548-016-1419-y

Source DB:  PubMed          Journal:  Int J Comput Assist Radiol Surg        ISSN: 1861-6410            Impact factor:   2.924


  30 in total

1.  Achieving routine submillisievert CT scanning: report from the summit on management of radiation dose in CT.

Authors:  Cynthia H McCollough; Guang Hong Chen; Willi Kalender; Shuai Leng; Ehsan Samei; Katsuyuki Taguchi; Ge Wang; Lifeng Yu; Roderic I Pettigrew
Journal:  Radiology       Date:  2012-06-12       Impact factor: 11.105

2.  Substantial dose reduction in modern multi-slice spiral computed tomography (MSCT)-guided craniofacial and skull base surgery.

Authors:  G Widmann; M Fasser; P Schullian; A Zangerl; W Puelacher; F Kral; H Riechelmann; W Jaschke; R Bale
Journal:  Rofo       Date:  2012-01-13

3.  The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103.

Authors: 
Journal:  Ann ICRP       Date:  2007

4.  Radiation dose vs. image quality for low-dose CT protocols of the head for maxillofacial surgery and oral implant planning.

Authors:  M Loubele; R Jacobs; F Maes; F Schutyser; D Debaveye; R Bogaerts; W Coudyzer; D Vandermeulen; J van Cleynenbreugel; G Marchal; P Suetens
Journal:  Radiat Prot Dosimetry       Date:  2006-02-03       Impact factor: 0.972

5.  Accuracy of linear measurements using dental cone beam and conventional multislice computed tomography.

Authors:  A Suomalainen; T Vehmas; M Kortesniemi; S Robinson; J Peltola
Journal:  Dentomaxillofac Radiol       Date:  2008-01       Impact factor: 2.419

6.  Dosimetry and image quality of four dental cone beam computed tomography scanners compared with multislice computed tomography scanners.

Authors:  A Suomalainen; T Kiljunen; Y Käser; J Peltola; M Kortesniemi
Journal:  Dentomaxillofac Radiol       Date:  2009-09       Impact factor: 2.419

Review 7.  Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm.

Authors:  Alvin C Silva; Holly J Lawder; Amy Hara; Jennifer Kujak; William Pavlicek
Journal:  AJR Am J Roentgenol       Date:  2010-01       Impact factor: 3.959

8.  Ultralow-dose CT of the craniofacial bone for navigated surgery using adaptive statistical iterative reconstruction and model-based iterative reconstruction: 2D and 3D image quality.

Authors:  Gerlig Widmann; Peter Schullian; Eva-Maria Gassner; Romed Hoermann; Reto Bale; Wolfgang Puelacher
Journal:  AJR Am J Roentgenol       Date:  2015-03       Impact factor: 3.959

9.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

10.  Low-dose dental computed tomography: significant dose reduction without loss of image quality.

Authors:  P Rustemeyer; U Streubühr; J Suttmoeller
Journal:  Acta Radiol       Date:  2004-12       Impact factor: 1.990

View more
  4 in total

1.  Identification of mandibular canal in cone beam computed tomography plane with different voxel sizes.

Authors:  Ameera Alabdulwahid; Wafa Alfaleh
Journal:  Saudi Dent J       Date:  2019-11-06

Review 2.  Ultralow Dose MSCT Imaging in Dental Implantology.

Authors:  Gerlig Widmann; Asma'a A Al-Ekrish
Journal:  Open Dent J       Date:  2018-01-31

3.  Localization of the inferior alveolar canal using ultralow dose CT with iterative reconstruction techniques.

Authors:  Asmaa A Al-Ekrish; Wafa Alfaleh; Romed Hörmann; Ameera Alabdulwahid; Wolfgang Puelacher; Gerlig Widmann
Journal:  Dentomaxillofac Radiol       Date:  2018-07-23       Impact factor: 2.419

4.  Comparability of dental implant site ridge measurements using ultra-low-dose multidetector row computed tomography combined with filtered back-projection, adaptive statistical iterative reconstruction, and model-based iterative reconstruction.

Authors:  Asma'a Abdurrahman Al-Ekrish; Reema Al-Shawaf; Wafa Alfaleh; Romed Hörmann; Wolfgang Puelacher; Gerlig Widmann
Journal:  Oral Radiol       Date:  2018-10-13       Impact factor: 1.852

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.