Literature DB >> 16461498

Radiation dose vs. image quality for low-dose CT protocols of the head for maxillofacial surgery and oral implant planning.

M Loubele1, R Jacobs, F Maes, F Schutyser, D Debaveye, R Bogaerts, W Coudyzer, D Vandermeulen, J van Cleynenbreugel, G Marchal, P Suetens.   

Abstract

The goal of this study was to determine the acquisition parameters for a low-dose multi-slice CT protocol and to compare the effective dose and the image quality of this low-dose protocol with the image quality of a clinical multi-slice CT protocol, routinely used for visualisation of the head. The low-dose protocol was derived from a clinical multi-slice CT protocol by lowering mA s and kV and increasing the pitch. The low-dose protocol yielded a dose reduction from 1.5 to 0.18 mSv for a multi-slice CT scan of the whole head, whereas noise in the low-dose CT images was increased. For bone segmentation, noise could be reduced by use of a non-linear edge preserving smoothing filter. Tests on ESP and skull phantom indicated that the accuracy of the measurements on low-dose CT is acceptable for image-based planning of maxillofacial and oral implant surgery, reducing the dose by a factor of 8.

Entities:  

Mesh:

Year:  2006        PMID: 16461498     DOI: 10.1093/rpd/nci749

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  14 in total

1.  Effect of electric potential and current on mandibular linear measurements in cone beam CT.

Authors:  S Panmekiate; W Apinhasmit; A Petersson
Journal:  Dentomaxillofac Radiol       Date:  2012-04-12       Impact factor: 2.419

2.  Accuracy of computer-aided design models of the jaws produced using ultra-low MDCT doses and ASIR and MBIR.

Authors:  Asma'a A Al-Ekrish; Sara A Alfadda; Wadea Ameen; Romed Hörmann; Wolfgang Puelacher; Gerlig Widmann
Journal:  Int J Comput Assist Radiol Surg       Date:  2018-06-16       Impact factor: 2.924

3.  Effective radiation dose and eye lens dose in dental cone beam CT: effect of field of view and angle of rotation.

Authors:  R Pauwels; G Zhang; C Theodorakou; A Walker; H Bosmans; R Jacobs; R Bogaerts; K Horner
Journal:  Br J Radiol       Date:  2014-10       Impact factor: 3.039

4.  Validity of linear measurements of the jaws using ultralow-dose MDCT and the iterative techniques of ASIR and MBIR.

Authors:  Asma'a A Al-Ekrish; Reema Al-Shawaf; Peter Schullian; Ra'ed Al-Sadhan; Romed Hörmann; Gerlig Widmann
Journal:  Int J Comput Assist Radiol Surg       Date:  2016-06-02       Impact factor: 2.924

5.  The topography of displaced canines: a 3D-CT study.

Authors:  Elisabeth Hofmann; Melanie Rodich; Ursula Hirschfelder
Journal:  J Orofac Orthop       Date:  2011-08       Impact factor: 1.938

6.  The growing concern of radiation dose in paediatric dental and maxillofacial CBCT: an easy guide for daily practice.

Authors:  Andreas Stratis; Guozhi Zhang; Reinhilde Jacobs; Ria Bogaerts; Hilde Bosmans
Journal:  Eur Radiol       Date:  2019-07-01       Impact factor: 5.315

7.  Standardization and optimization of CT protocols to achieve low dose.

Authors:  Sigal Trattner; Gregory D N Pearson; Cynthia Chin; Dianna D Cody; Rajiv Gupta; Christopher P Hess; Mannudeep K Kalra; James M Kofler; Mayil S Krishnam; Andrew J Einstein
Journal:  J Am Coll Radiol       Date:  2014-03       Impact factor: 5.532

8.  Imaging of mandible invasion by oral squamous cell carcinoma using computed tomography, cone-beam computed tomography and bone scintigraphy with SPECT.

Authors:  Samer G Hakim; Henning Wieker; Thomas Trenkle; Peter Sieg; Jens Konitzer; Konstanze Holl-Ulrich; Hans-Christian Jacobsen
Journal:  Clin Oral Investig       Date:  2013-07-20       Impact factor: 3.573

Review 9.  Effective dose of dental CBCT-a meta analysis of published data and additional data for nine CBCT units.

Authors:  J B Ludlow; R Timothy; C Walker; R Hunter; E Benavides; D B Samuelson; M J Scheske
Journal:  Dentomaxillofac Radiol       Date:  2015       Impact factor: 2.419

10.  Imaging Factors Impacting on Accuracy and Radiation Dose in 3D Printing.

Authors:  Jorge Javier de Lima Moreno; Gabriela Salatino Liedke; Roberto Soler; Heloisa Emília Dias da Silveira; Heraldo Luis Dias da Silveira
Journal:  J Maxillofac Oral Surg       Date:  2018-02-28