Photon-counting hexagonal pixel array CdTe detector: Spatial resolution characteristics for image-guided interventional applications.

Srinivasan Vedantham1, Suman Shrestha1, Andrew Karellas1, Linxi Shi1, Matthew J Gounis1, Ronaldo Bellazzini2, Gloria Spandre2, Alessandro Brez2, Massimo Minuti2.   

Abstract

PURPOSE: High-resolution, photon-counting, energy-resolved detector with fast-framing capability can facilitate simultaneous acquisition of precontrast and postcontrast images for subtraction angiography without pixel registration artifacts and can facilitate high-resolution real-time imaging during image-guided interventions. Hence, this study was conducted to determine the spatial resolution characteristics of a hexagonal pixel array photon-counting cadmium telluride (CdTe) detector.
METHODS: A 650 μm thick CdTe Schottky photon-counting detector capable of concurrently acquiring up to two energy-windowed images was operated in a single energy-window mode to include photons of 10 keV or higher. The detector had hexagonal pixels with apothem of 30 μm resulting in pixel pitch of 60 and 51.96 μm along the two orthogonal directions. The detector was characterized at IEC-RQA5 spectral conditions. Linear response of the detector was determined over the air kerma rate relevant to image-guided interventional procedures ranging from 1.3 nGy/frame to 91.4 μGy/frame. Presampled modulation transfer was determined using a tungsten edge test device. The edge-spread function and the finely sampled line spread function accounted for hexagonal sampling, from which the presampled modulation transfer function (MTF) was determined. Since detectors with hexagonal pixels require resampling to square pixels for distortion-free display, the optimal square pixel size was determined by minimizing the root-mean-squared-error of the aperture functions for the square and hexagonal pixels up to the Nyquist limit.
RESULTS: At Nyquist frequencies of 8.33 and 9.62 cycles/mm along the apothem and orthogonal to the apothem directions, the modulation factors were 0.397 and 0.228, respectively. For the corresponding axis, the limiting resolution defined as 10% MTF occurred at 13.3 and 12 cycles/mm, respectively. Evaluation of the aperture functions yielded an optimal square pixel size of 54 μm. After resampling to 54 μm square pixels using trilinear interpolation, the presampled MTF at Nyquist frequency of 9.26 cycles/mm was 0.29 and 0.24 along the orthogonal directions and the limiting resolution (10% MTF) occurred at approximately 12 cycles/mm. Visual analysis of a bar pattern image showed the ability to resolve close to 12 line-pairs/mm and qualitative evaluation of a neurovascular nitinol-stent showed the ability to visualize its struts at clinically relevant conditions.
CONCLUSIONS: Hexagonal pixel array photon-counting CdTe detector provides high spatial resolution in single-photon counting mode. After resampling to optimal square pixel size for distortion-free display, the spatial resolution is preserved. The dual-energy capabilities of the detector could allow for artifact-free subtraction angiography and basis material decomposition. The proposed high-resolution photon-counting detector with energy-resolving capability can be of importance for several image-guided interventional procedures as well as for pediatric applications.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27147324      PMCID: PMC4826388          DOI: 10.1118/1.4944868

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  24 in total

1.  Full breast digital mammography with an amorphous silicon-based flat panel detector: physical characteristics of a clinical prototype.

Authors:  S Vedantham; A Karellas; S Suryanarayanan; D Albagli; S Han; E J Tkaczyk; C E Landberg; B Opsahl-Ong; P R Granfors; I Levis; C J D'Orsi; R E Hendrick
Journal:  Med Phys       Date:  2000-03       Impact factor: 4.071

2.  Accuracy of a simple method for deriving the presampled modulation transfer function of a digital radiographic system from an edge image.

Authors:  Egbert Buhr; Susanne Günther-Kohfahl; Ulrich Neitzel
Journal:  Med Phys       Date:  2003-09       Impact factor: 4.071

3.  Performance of a 41 x 41 cm2 amorphous silicon flat panel x-ray detector designed for angiographic and R&F imaging applications.

Authors:  Paul R Granfors; Richard Aufrichtig; George E Possin; Brian W Giambattista; Zhong S Huang; Jianqiang Liu; Bing Ma
Journal:  Med Phys       Date:  2003-10       Impact factor: 4.071

4.  Evaluation of the imaging properties of an amorphous selenium-based flat panel detector for digital fluoroscopy.

Authors:  D C Hunt; O Tousignant; J A Rowlands
Journal:  Med Phys       Date:  2004-05       Impact factor: 4.071

5.  International Subarachnoid Aneurysm Trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised trial.

Authors:  Andrew Molyneux; Richard Kerr; Irene Stratton; Peter Sandercock; Mike Clarke; Julia Shrimpton; Rury Holman
Journal:  Lancet       Date:  2002-10-26       Impact factor: 79.321

6.  Evaluation of the microangiographic fluoroscope (MAF) using generalized system performance metrics.

Authors:  Amit Jain; Daniel R Bednarek; Stephen Rudin
Journal:  Med Phys       Date:  2013-03       Impact factor: 4.071

7.  Investigation of image lag and modulation transfer function in fluoroscopy images obtained with a dynamic flat-panel detector.

Authors:  Hiroki Kawashima; Rie Tanaka; Katsuhiro Ichikawa; Kosuke Matsubara; Hiroji Iida; Shigeru Sanada
Journal:  Radiol Phys Technol       Date:  2013-04-09

8.  High-resolution imager for digital mammography: physical characterization of a prototype sensor.

Authors:  Sankararaman Suryanarayanan; Andrew Karellas; Srinivasan Vedantham; Steven K Onishi
Journal:  Phys Med Biol       Date:  2005-08-11       Impact factor: 3.609

9.  DQE(f) of four generations of computed radiography acquisition devices.

Authors:  J T Dobbins; D L Ergun; L Rutz; D A Hinshaw; H Blume; D C Clark
Journal:  Med Phys       Date:  1995-10       Impact factor: 4.071

10.  Endovascular image-guided interventions (EIGIs).

Authors:  Stephen Rudin; Daniel R Bednarek; Kenneth R Hoffmann
Journal:  Med Phys       Date:  2008-01       Impact factor: 4.071

View more
  3 in total

1.  Emerging Breast Imaging Technologies on the Horizon.

Authors:  Srinivasan Vedantham; Andrew Karellas
Journal:  Semin Ultrasound CT MR       Date:  2017-09-13       Impact factor: 1.875

2.  Photon-counting hexagonal pixel array CdTe detector: Spatial resolution characteristics for image-guided interventional applications.

Authors:  Srinivasan Vedantham; Suman Shrestha; Andrew Karellas; Linxi Shi; Matthew J Gounis; Ronaldo Bellazzini; Gloria Spandre; Alessandro Brez; Massimo Minuti
Journal:  Med Phys       Date:  2016-05       Impact factor: 4.071

3.  Direct Thermal Growth of Large Scale Cl-doped CdTe Film for Low Voltage High Resolution X-ray Image Sensor.

Authors:  Silah Lee; Jin Sung Kim; Kyeong Rok Ko; Gun Hwan Lee; Dong Jin Lee; Dong Wook Kim; Jin Eui Kim; Ho Kyung Kim; Dong Woon Kim; Seongil Im
Journal:  Sci Rep       Date:  2018-10-04       Impact factor: 4.379

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.