Michelle J Alfa1,2, Nancy Olson3.
Abstract
BACKGROUND: The objective of this study was to validate the relative light unit (RLU) cut-off of adequate cleaning of flexible colonoscopes for an ATP (adenosine tri-phosphate) test kit that used a sponge channel collection method.Entities:
Keywords: Cleaning adequacy; Cut-off; Sponge-sample
Mesh:
Substances:
Year: 2016 PMID: 27142441 PMCID: PMC4855436 DOI: 10.1186/s13104-016-2066-7
Source DB: PubMed Journal: BMC Res Notes ISSN: 1756-0500
Tests performed on each colonoscope channel sample type
| Test method | Instrument channel segmenta | ||
|---|---|---|---|
| Samples collected (N = 5) | |||
| Flush | Sponge elution | Sponge | |
| ATP | X | X | X |
| Protein quantitation | X | X | Not done |
| Viable count | X | X | Not done |
aEach sample is from a different colonoscope
Comparison of residual ATP detected in a soiled colonoscope instrument channel evaluated using various sample collection methods
| Channel sample collection methoda | RLUsb (standard deviation) | Protein μg/cm2 (standard deviation) |
|---|---|---|
| Sponge directc | 9304.2 (1028.6) | Not applicable |
| Sponge eluted (20 mL sample) | 2511.4 (1142.1) | 0.54 (0.14) |
| Flush only (20 mL sample) | 2850.4 (1261.6) | 26.66 (5.98) |
aData for each test parameter represents the average of five replicate endoscopes tested (N = 5)
bATP was reported as relative light units (RLUs)
cThe range of RLUS for the sponge direct collection method was 7709–9999 with three of the five replicates having RLU values of 9999 (i.e. maximum RLU value read by the luminometer). This 9999 maximum RLU value was used for the purposes of calculating the average and standard deviation recognizing that it was not an exact RLU value
Comparison of residual ATP detected by the direct sponge method versus level of residual protein and viable count in colonoscope instrument channels
| Sponge sampleb (N = 3) | Flush samplec (N = 3) | |||
|---|---|---|---|---|
| Test parametera | RLUs (Std) | Protein μg/cm2 (Std) |
|
|
| Positive control | 9747.00 (436.48) | 335.4 (65.55) | 6.782 (0.212) | 6.669 (0.192) |
| Partial clean | 6601.00 (1684.00) | 3.99 (3.16) | 5.254 (0.142) | 4.476 (0.253) |
| Full cleaning | 29.00 (36.39) | 0.23 (0.07) | 0.787 (0.364) | 1.608 (0.269) |
| Negative control | 14.00 (21.66) | 0.04 (0.07) | 0.025 any organism (0.090) | |
aData for each test parameter represents the average and standard deviation (Std) of three replicate endoscope tests (N = 3)
bATP was measured as relative light units (RLUs)
cThe protein and viable counts were done on a parallel set of colonoscopes that were soiled in the same manner and were harvested using the flush method of sample collection