Literature DB >> 15877014

Real-time monitoring in managing the decontamination of flexible gastrointestinal endoscopes.

Peter C Obee1, Chris J Griffith, Rose A Cooper, Richard P Cooke, Nerys E Bennion, Michael Lewis.   

Abstract

OBJECTIVES: To evaluate the efficacy of endoscope reprocessing using bacteriologic methods and adenosine triphosphate (ATP) bioluminescence and to evaluate the potential benefits of the latter in managing the process.
SETTING: An endoscopy unit in each of 2 acute district general hospitals in the United Kingdom.
METHODS: Following visual observation, 8 locations important in the decontamination of gastrointestinal (GI) endoscopes were monitored during reprocessing by bacteriologic sampling (paddle method) and ATP bioluminescence.
RESULTS: Endoscope reprocessing was conducted in accordance with British Society of Gastroenterology guidelines. Bacteriologic tests required 24 hours for results to be obtained; ATP tests were completed within 2 minutes. Of the 504 sites tested throughout the decontamination process, bacterial growth was recorded on 32 occasions, 3 from the end product. Throughout the decontamination process, ATP tests obtained 95 results above benchmark values and 1 from the end product.
CONCLUSIONS: The bacteriologic results after terminal disinfection indicated that the reprocessing was carried out effectively overall. ATP results prior to disinfection indicated that manual cleaning might not have always been performed adequately and that this could have impaired the disinfection process. ATP provided a rapid means of assessing the efficacy of cleaning steps prior to terminal disinfection.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15877014     DOI: 10.1016/j.ajic.2004.07.008

Source DB:  PubMed          Journal:  Am J Infect Control        ISSN: 0196-6553            Impact factor:   2.918


  7 in total

Review 1.  Duodenoscope-Associated Infections: Update on an Emerging Problem.

Authors:  M Rubayat Rahman; Abhilash Perisetti; Roxana Coman; Pardeep Bansal; Rajiv Chhabra; Hemant Goyal
Journal:  Dig Dis Sci       Date:  2018-12-19       Impact factor: 3.199

Review 2.  Is peracetic acid suitable for the cleaning step of reprocessing flexible endoscopes?

Authors:  Günter Kampf; Patricia M Fliss; Heike Martiny
Journal:  World J Gastrointest Endosc       Date:  2014-09-16

3.  Adenosine triphosphate bioluminescence for bacteriologic surveillance and reprocessing strategies for minimizing risk of infection transmission by duodenoscopes.

Authors:  Saurabh Sethi; Robert J Huang; Monique T Barakat; Niaz Banaei; Shai Friedland; Subhas Banerjee
Journal:  Gastrointest Endosc       Date:  2016-11-03       Impact factor: 9.427

Review 4.  Transmission of infection by flexible gastrointestinal endoscopy and bronchoscopy.

Authors:  Julia Kovaleva; Frans T M Peters; Henny C van der Mei; John E Degener
Journal:  Clin Microbiol Rev       Date:  2013-04       Impact factor: 26.132

Review 5.  Recent Update on Microbiological Monitoring of Gastrointestinal Endoscopes after High-Level Disinfection.

Authors:  Suk Pyo Shin; Won Hee Kim
Journal:  Clin Endosc       Date:  2015-09-30

6.  Simulated-use validation of a sponge ATP method for determining the adequacy of manual cleaning of endoscope channels.

Authors:  Michelle J Alfa; Nancy Olson
Journal:  BMC Res Notes       Date:  2016-05-04

7.  Microbiological Surveillance of Endoscopes in a Southern Italian Transplantation Hospital: A Retrospective Study from 2016 to 2019.

Authors:  Valentina Marchese; Daniele Di Carlo; Gaetano Fazio; Santi Mauro Gioè; Angelo Luca; Rossella Alduino; Monica Rizzo; Fabio Tuzzolino; Francesco Monaco; Pier Giulio Conaldi; Bruno Douradinha; Giuseppina Di Martino
Journal:  Int J Environ Res Public Health       Date:  2021-03-16       Impact factor: 3.390

  7 in total