Michelle J Alfa1, Iram Fatima, Nancy Olson. 1. Clinical Microbiology Diagnostic Services of Manitoba, University of Manitoba, Winnipeg, Manitoba, Canada. malfa@dsmanitoba.ca
Abstract
BACKGROUND: The study objective was to verify that the adenosine triphosphate (ATP) benchmark of <200 relative light units (RLUs) was achievable in a busy endoscopy clinic that followed the manufacturer's manual cleaning instructions. METHODS: All channels from patient-used colonoscopes (20) and duodenoscopes (20) in a tertiary care hospital endoscopy clinic were sampled after manual cleaning and tested for residual ATP. The ATP test benchmark for adequate manual cleaning was set at <200 RLUs. The benchmark for protein was <6.4 μg/cm(2), and, for bioburden, it was <4-log10 colony-forming units/cm(2). RESULTS: Our data demonstrated that 96% (115/120) of channels from 20 colonoscopes and 20 duodenoscopes evaluated met the ATP benchmark of <200 RLUs. The 5 channels that exceeded 200 RLUs were all elevator guide-wire channels. All 120 of the manually cleaned endoscopes tested had protein and bioburden levels that were compliant with accepted benchmarks for manual cleaning for suction-biopsy, air-water, and auxiliary water channels. CONCLUSION: Our data confirmed that, by following the endoscope manufacturer's manual cleaning recommendations, 96% of channels in gastrointestinal endoscopes would have <200 RLUs for the ATP test kit evaluated and would meet the accepted clean benchmarks for protein and bioburden.
BACKGROUND: The study objective was to verify that the adenosine triphosphate (ATP) benchmark of <200 relative light units (RLUs) was achievable in a busy endoscopy clinic that followed the manufacturer's manual cleaning instructions. METHODS: All channels from patient-used colonoscopes (20) and duodenoscopes (20) in a tertiary care hospital endoscopy clinic were sampled after manual cleaning and tested for residual ATP. The ATP test benchmark for adequate manual cleaning was set at <200 RLUs. The benchmark for protein was <6.4 μg/cm(2), and, for bioburden, it was <4-log10 colony-forming units/cm(2). RESULTS: Our data demonstrated that 96% (115/120) of channels from 20 colonoscopes and 20 duodenoscopes evaluated met the ATP benchmark of <200 RLUs. The 5 channels that exceeded 200 RLUs were all elevator guide-wire channels. All 120 of the manually cleaned endoscopes tested had protein and bioburden levels that were compliant with accepted benchmarks for manual cleaning for suction-biopsy, air-water, and auxiliary water channels. CONCLUSION: Our data confirmed that, by following the endoscope manufacturer's manual cleaning recommendations, 96% of channels in gastrointestinal endoscopes would have <200 RLUs for the ATP test kit evaluated and would meet the accepted clean benchmarks for protein and bioburden.
Authors: Nina Parohl; Doris Stiefenhöfer; Sabine Heiligtag; Henning Reuter; Dana Dopadlik; Frank Mosel; Guido Gerken; Alexander Dechêne; Evelyn Heintschel von Heinegg; Christoph Jochum; Jan Buer; Walter Popp Journal: GMS Hyg Infect Control Date: 2017-03-27
Authors: Harminder Singh; Donald R Duerksen; Gale Schultz; Carol Reidy; Pat DeGagne; Nancy Olson; Zoann Nugent; Michelle J Alfa Journal: Endosc Int Open Date: 2019-01-30