Literature DB >> 27059266

Feasibility of using linearly polarized rotating birdcage transmitters and close-fitting receive arrays in MRI to reduce SAR in the vicinity of deep brain simulation implants.

Laleh Golestanirad1,2, Boris Keil1,2,3, Leonardo M Angelone4, Giorgio Bonmassar1,2, Azma Mareyam1, Lawrence L Wald1,2.   

Abstract

PURPOSE: MRI of patients with deep brain stimulation (DBS) implants is strictly limited due to safety concerns, including high levels of local specific absorption rate (SAR) of radiofrequency (RF) fields near the implant and related RF-induced heating. This study demonstrates the feasibility of using a rotating linearly polarized birdcage transmitter and a 32-channel close-fit receive array to significantly reduce local SAR in MRI of DBS patients.
METHODS: Electromagnetic simulations and phantom experiments were performed with generic DBS lead geometries and implantation paths. The technique was based on mechanically rotating a linear birdcage transmitter to align its zero electric-field region with the implant while using a close-fit receive array to significantly increase signal to noise ratio of the images.
RESULTS: It was found that the zero electric-field region of the transmitter is thick enough at 1.5 Tesla to encompass DBS lead trajectories with wire segments that were up to 30 degrees out of plane, as well as leads with looped segments. Moreover, SAR reduction was not sensitive to tissue properties, and insertion of a close-fit 32-channel receive array did not degrade the SAR reduction performance.
CONCLUSION: The ensemble of rotating linear birdcage and 32-channel close-fit receive array introduces a promising technology for future improvement of imaging in patients with DBS implants. Magn Reson Med 77:1701-1712, 2017.
© 2016 International Society for Magnetic Resonance in Medicine. © 2016 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  MRI; computational modeling; deep brain stimulation (DBS); electrode artifact; finite element modeling; medical implants; safety; specific absorption rate

Mesh:

Year:  2016        PMID: 27059266      PMCID: PMC5052088          DOI: 10.1002/mrm.26220

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  45 in total

1.  Implantable spinal fusion stimulator: assessment of MR safety and artifacts.

Authors:  F G Shellock; M Hatfield; B J Simon; S Block; J Wamboldt; P M Starewicz; W F Punchard
Journal:  J Magn Reson Imaging       Date:  2000-08       Impact factor: 4.813

Review 2.  How does deep brain stimulation work? Present understanding and future questions.

Authors:  Cameron C McIntyre; Marc Savasta; Benjamin L Walter; Jerrold L Vitek
Journal:  J Clin Neurophysiol       Date:  2004 Jan-Feb       Impact factor: 2.177

3.  Deep brain stimulation for Parkinson's disease: surgical technique and perioperative management.

Authors:  Andre Machado; Ali R Rezai; Brian H Kopell; Robert E Gross; Ashwini D Sharan; Alim-Louis Benabid
Journal:  Mov Disord       Date:  2006-06       Impact factor: 10.338

4.  Bio-heat transfer model of deep brain stimulation-induced temperature changes.

Authors:  Maged M Elwassif; Qingjun Kong; Maribel Vazquez; Marom Bikson
Journal:  J Neural Eng       Date:  2006-11-06       Impact factor: 5.379

5.  Reduction of implant RF heating through modification of transmit coil electric field.

Authors:  Yigitcan Eryaman; Burak Akin; Ergin Atalar
Journal:  Magn Reson Med       Date:  2010-12-08       Impact factor: 4.668

6.  Changes in cerebral activity pattern due to subthalamic nucleus or internal pallidum stimulation in Parkinson's disease.

Authors:  P Limousin; J Greene; P Pollak; J Rothwell; A L Benabid; R Frackowiak
Journal:  Ann Neurol       Date:  1997-09       Impact factor: 10.422

7.  On the RF heating of coronary stents at 7.0 Tesla MRI.

Authors:  Lukas Winter; Eva Oberacker; Celal Özerdem; Yiyi Ji; Florian von Knobelsdorff-Brenkenhoff; Gerd Weidemann; Bernd Ittermann; Frank Seifert; Thoralf Niendorf
Journal:  Magn Reson Med       Date:  2014-10-07       Impact factor: 4.668

8.  Evaluation of specific absorption rate as a dosimeter of MRI-related implant heating.

Authors:  Kenneth B Baker; Jean A Tkach; John A Nyenhuis; Michael Phillips; Frank G Shellock; Jorge Gonzalez-Martinez; Ali R Rezai
Journal:  J Magn Reson Imaging       Date:  2004-08       Impact factor: 4.813

9.  Effects of receive-only inserts on specific absorption rate, B1 (+) field, and Tx coil performance.

Authors:  Narayanan Krishnamurthy; Tiejun Zhao; Tamer S Ibrahim
Journal:  J Magn Reson Imaging       Date:  2013-08-01       Impact factor: 4.813

10.  A novel brain stimulation technology provides compatibility with MRI.

Authors:  Peter Serano; Leonardo M Angelone; Husam Katnani; Emad Eskandar; Giorgio Bonmassar
Journal:  Sci Rep       Date:  2015-04-29       Impact factor: 4.379

View more
  20 in total

1.  The 'virtual DBS population': five realistic computational models of deep brain stimulation patients for electromagnetic MR safety studies.

Authors:  Bastien Guerin; Maria Ida Iacono; Mathias Davids; Darin Dougherty; Leonardo M Angelone; Lawrence L Wald
Journal:  Phys Med Biol       Date:  2019-02-04       Impact factor: 3.609

2.  Construction and modeling of a reconfigurable MRI coil for lowering SAR in patients with deep brain stimulation implants.

Authors:  Laleh Golestanirad; Maria Ida Iacono; Boris Keil; Leonardo M Angelone; Giorgio Bonmassar; Michael D Fox; Todd Herrington; Elfar Adalsteinsson; Cristen LaPierre; Azma Mareyam; Lawrence L Wald
Journal:  Neuroimage       Date:  2016-12-21       Impact factor: 6.556

3.  Local SAR near deep brain stimulation (DBS) electrodes at 64 and 127 MHz: A simulation study of the effect of extracranial loops.

Authors:  Laleh Golestanirad; Leonardo M Angelone; Maria Ida Iacono; Husam Katnani; Lawrence L Wald; Giorgio Bonmassar
Journal:  Magn Reson Med       Date:  2016-10-31       Impact factor: 4.668

4.  Reducing RF-induced Heating near Implanted Leads through High-Dielectric Capacitive Bleeding of Current (CBLOC).

Authors:  Laleh Golestanirad; Leonardo M Angelone; John Kirsch; Sean Downs; Boris Keil; Giorgio Bonmassar; Lawrence L Wald
Journal:  IEEE Trans Microw Theory Tech       Date:  2019-01-01       Impact factor: 3.599

Review 5.  Improving Safety of MRI in Patients with Deep Brain Stimulation Devices.

Authors:  Alexandre Boutet; Clement T Chow; Keshav Narang; Gavin J B Elias; Clemens Neudorfer; Jürgen Germann; Manish Ranjan; Aaron Loh; Alastair J Martin; Walter Kucharczyk; Christopher J Steele; Ileana Hancu; Ali R Rezai; Andres M Lozano
Journal:  Radiology       Date:  2020-06-23       Impact factor: 11.105

6.  Reconfigurable MRI technology for low-SAR imaging of deep brain stimulation at 3T: Application in bilateral leads, fully-implanted systems, and surgically modified lead trajectories.

Authors:  Ehsan Kazemivalipour; Boris Keil; Alireza Vali; Sunder Rajan; Behzad Elahi; Ergin Atalar; Lawrence L Wald; Joshua Rosenow; Julie Pilitsis; Laleh Golestanirad
Journal:  Neuroimage       Date:  2019-05-13       Impact factor: 6.556

7.  A simple geometric analysis method for measuring and mitigating RF induced currents on Deep Brain Stimulation leads by multichannel transmission/reception.

Authors:  Yigitcan Eryaman; Naoharu Kobayashi; Sean Moen; Joshua Aman; Andrea Grant; J Thomas Vaughan; Gregory Molnar; Michael C Park; Jerrold Vitek; Gregor Adriany; Kamil Ugurbil; Noam Harel
Journal:  Neuroimage       Date:  2018-09-28       Impact factor: 6.556

8.  Evaluation of RF interactions between a 3T birdcage transmit coil and transcranial magnetic stimulation coils using a realistically shaped head phantom.

Authors:  Lucia I Navarro de Lara; Laleh Golestanirad; Sergey N Makarov; Jason P Stockmann; Lawrence L Wald; Aapo Nummenmaa
Journal:  Magn Reson Med       Date:  2020-01-23       Impact factor: 4.668

9.  Realistic modeling of deep brain stimulation implants for electromagnetic MRI safety studies.

Authors:  Bastien Guerin; Peter Serano; Maria Ida Iacono; Todd M Herrington; Alik S Widge; Darin D Dougherty; Giorgio Bonmassar; Leonardo M Angelone; Lawrence L Wald
Journal:  Phys Med Biol       Date:  2018-05-04       Impact factor: 3.609

10.  Changes in the specific absorption rate (SAR) of radiofrequency energy in patients with retained cardiac leads during MRI at 1.5T and 3T.

Authors:  Laleh Golestanirad; Amir Ali Rahsepar; John E Kirsch; Kenichiro Suwa; Jeremy C Collins; Leonardo M Angelone; Boris Keil; Rod S Passman; Giorgio Bonmassar; Peter Serano; Peter Krenz; Jim DeLap; James C Carr; Lawrence L Wald
Journal:  Magn Reson Med       Date:  2018-06-12       Impact factor: 4.668

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.