Peter Serano1,2,3, Leonardo M Angelone1,3, Husam Katnani4, Emad Eskandar4, Giorgio Bonmassar1. 1. Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, MA, U.S.A. 2. Department of Electrical and Computer Engineering, University of Maryland, College Park, MD, U.S.A. 3. Division of Biomedical Physics, Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, U.S. Food and Drug Administration, Silver Spring, MD, U.S.A. 4. Department of Neurosurgery, Massachusetts General Hospital, Harvard Medical School, Boston, MA.
Abstract
Clinical electrical stimulation systems--such as pacemakers and deep brain stimulators (DBS)--are an increasingly common therapeutic option to treat a large range of medical conditions. Despite their remarkable success, one of the significant limitations of these medical devices is the limited compatibility with magnetic resonance imaging (MRI), a standard diagnostic tool in medicine. During an MRI exam, the leads used with these devices, implanted in the body of the patient, act as an electric antenna potentially causing a large amount of energy to be absorbed in the tissue, which can lead to serious heat-related injury. This study presents a novel lead design that reduces the antenna effect and allows for decreased tissue heating during MRI. The optimal parameters of the wire design were determined by a combination of computational modeling and experimental measurements. The results of these simulations were used to build a prototype, which was tested in a gel phantom during an MRI scan. Measurement results showed a three-fold decrease in heating when compared to a commercially available DBS lead. Accordingly, the proposed design may allow a significantly increased number of patients with medical implants to have safe access to the diagnostic benefits of MRI.
Clinical electrical stimulation systems--such as pacemakers and deep brain stimulators (DBS)--are an increasingly common therapeutic option to treat a large range of medical conditions. Despite their remarkable success, one of the significant limitations of these medical devices is the limited compatibility with magnetic resonance imaging (MRI), a standard diagnostic tool in medicine. During an MRI exam, the leads used with these devices, implanted in the body of the patient, act as an electric antenna potentially causing a large amount of energy to be absorbed in the tissue, which can lead to serious heat-related injury. This study presents a novel lead design that reduces the antenna effect and allows for decreased tissue heating during MRI. The optimal parameters of the wire design were determined by a combination of computational modeling and experimental measurements. The results of these simulations were used to build a prototype, which was tested in a gel phantom during an MRI scan. Measurement results showed a three-fold decrease in heating when compared to a commercially available DBS lead. Accordingly, the proposed design may allow a significantly increased number of patients with medical implants to have safe access to the diagnostic benefits of MRI.
Magnetic resonance imaging (MRI) for patients with neurostimulators: advantages and
current limitations. Implanted medical devices such as
cardioverter-defibrillators, pacemakers, spinal cord stimulators, and deep brain
stimulation (DBS) have become well-accepted therapeutic options to treat a wide range of
medical conditions and contribute to improved quality of life1. Many
patients with implanted devices can benefit from MRI, which is the diagnostic tool of
choice for monitoring structural changes in the body, as well as diagnosing many common
illnesses including cancer, cardiovascular disease, and trauma. Additionally, functional
MRI is becoming more prevalent in assessing brain function and cognitive disorders23. However, approximately 300,000 patients with implanted or partially
implanted medical devices are denied MRI each year because of safety concerns4. A major concern when performing MRI examinations in patients with
electrically conductive implants is the increase in induced currents
(“antenna effect”) along conductive leads in the body that are
exposed to the radiofrequency (RF) waves of the MRI. The increase in current flow into
the tissue at the point of contact with the lead (i.e. the electrodes) causes a large
amount of RF energy to be absorbed in the tissue, which in turn causes surges in
temperatures that can lead to serious injury56789101112.
Temperature increases of up to 25°C were measured near a DBS 3389 lead
(Medtronic, Inc., Minneapolis, MN) in an in-vitro gel phantom at 1.5 T
MRI13. Additionally, increases of up to 30°C were measured
with the Medtronic 3389 lead in a swine head at 9.4 T14. More
importantly, two cases of serious, permanent neurological injury, after MRI exposure at
1.0 T in patients with DBS implants, have been reported1516.
In both cases the manufacturer guidelines were not followed and in one case the patient
developed paralysis following MRI examination16. The lack of access to
MRI is expensive to society because patients are denied the benefits of screening and
accurate diagnosis. A class of implantable devices — defined as
“MR Conditional”17 — have been shown to
pose no known hazards in the MRI environment when operated with specified conditions.
For example, the Activa® DBS system (Medtronic, Inc., Minneapolis, MN) is
approved for use in MRI18 with several conditions19,
including limited static and gradient magnetic fields, use of low power sequences, and
specific RF coils. These conditions, however, are restrictive. The limit for power
absorbed by the patient's head is over 30-fold less than typical values
allowed, which restricts the number, the type, and quality of MRI scans that can be
performed in a given session. The most commonly used transmit body coils are not
allowed, excluding the possibility of using MRI to diagnose morbidities in the human
torso (e.g., breast cancer, back pain). Additionally, the conditions exclude the use of
3.0 T MRI systems, which are commonly used in clinical20 and
research applications2122.
A novel technology to allow the safe use of MRI in patients with DBS
implants
The safety evaluation of the RF-induced heating injury risks in patients with
implanted medical devices undergoing MRI is based on several testing strategies and
tools, including pre-clinical (experimental, computational, and animal testing) as
well as clinical testing. Experimental testing includes measuring temperature
changes near the device while it is implanted in a gel that simulates the electrical
and thermal characteristics of the human body23. Additionally,
computational modeling has been increasingly used to complement experimental
testing, as it allows for extensive, cost-effective and systematic analysis of
several variables that can influence the amount of current flow into an implant and
the amount of energy absorbed by surrounding tissue.Several proposals have been made to modify the design of the implant to solve the
issue of RF-induced heating without interfering with device performance, such as
introducing RF chokes24, modifying the materials of the lead (e.g.,
carbon-loaded leads)252627282930, or coiling the wire31. A new type of lead based on “resistive tapered
stripline” (RTS) technology32 is herein proposed. The RTS
design can be best understood by recalling oceanic science, where an area of study
is the prevention of destructive standing waves (clapotis)33. Special
constructions reinforced with wide, rubble-mound beams break up wave energy over
some distance, preventing the formation of clapotis. Similarly, tapered dielectric
structures can break up or scatter RF energy due to their unique frequency response
characteristics. This characteristic has been studied for many applications
including microwave, millimeter-wave and optical-wave engineering343536, as well as stealth aircraft technology37.
This study presents a two-section stripline-based design (Fig.
1a) with an abrupt transition of electrical conductivity along its
length. Contrary to a common metallic wire, this design can break up the induced RF
current along the lead (Fig. 1b) caused by the MRI RF coil.
Consequently, RF-induced current along the RTS lead is more heterogeneously
distributed and significantly reduced at the electrode (Fig.
1c). Please refer to the Supplementary Information
for the theoretical background on RTS design.
Figure 1
RTS design and simulation setup.
(a) Schematic of the PtIr wire (diameter d = 100 μm,
electrical conductivity σ = 4.0·106 S/m) and the two-layer RTS design (electrical conductivity
σ1 and σ2, permittivity
ε1 and ε2, length
L1 and L2) used for the study. (b) Equivalent
circuit used to model the RTS implant with four sections: stimulator, two
layer transmission line, and electrode/tissue interface. The incident RF
field induces currents along the implants, which are reflected depending on
neighboring sections mismatched impedance (Z0, Z1,
Z2, and ZL). The resulting voltage amplitude at
each interface (V0, V1, and V2) was
generated by the induced current. (c) RF-induced currents along the two
leads. The current in the metallic conductor forms a standing wave with high
peaks in amplitude (Iw); conversely, the effect of RTS design is
two-fold: a) reduces the average induced currents (IRTS) along
the implant by worsening the antenna performance, and b) reduces the induced
current at the electrode (ΔI) by introducing scattering within
the implant. (d) CAD Model used in the numerical simulations, including a
16-leg high-pass birdcage body coil with RF shield, coil former and ASTM
phantom. (e) 3 D plot of electric field magnitude at the Larmor
frequency (f0 = 128 MHz) in the ASTM phantom model
used in the simulations. Results were normalized to a power level yielding a
whole-body SAR = 2 W/kg (i.e., Normal Operating Mode41). (f) Placement of lead inside the phantom. The location was
chosen because of the high magnitude of incident electric field.
Overall, two different RTS designs were used for the study: (a) an initial design
constructed with conductive ink deposited on a polymer substrate
(“flat-design”), and (b) a second wire-based design
(“wire-design”). The flat-design was used for the simulations
in phantom (Figs. 1 and 2), the
manufacturing of the first prototype, and the bench testing experiments (Fig. 3). The wire-design was used for simulations with human
body models (Fig. 4) and manufacturing of a second prototype
(Fig. 5). Both simulations and measurements confirmed that
the RTS design “cloaks” the incident RF-field38, so that the lead is “RF-transparent” (i.e., the presence
of the lead does not significantly affect the RF fields present in a phantom).
Figure 2
Optimal RTS design in phantom.
(a) 10 g-avg. SAR inside the phantom at a distance of
0.1 mm from the electrode obtained by varying the length
(L2) of the second section (see Methods). Plots include
different conductivity ratios for the two layers. In all cases the total
resistance of the lead was R = 400 Ω. (b)
10 g-avg. SAR at the same point obtained by varying the total
resistance of the lead. Plots include four combinations of conductivity
ratios of the two layers and length L2 of the second section. (c)
Maximum inductance of the RTS varying the total resistance of the lead.
Plots include five combinations of conductivity ratios of the two layers and
length L2 of the second section. (d) Amplitude of induced current
inside the lead with the PtIr wire, with the RTS lead selected for prototype
manufacturing (right) and in the corresponding volume of the ASTM phantom
without lead. The RTS lead allowed for a 37-fold decrease in induced current
at the electrode (Length along Lead = 0 mm). In all cases the
total length of the leads was 40 cm. (e) Numerical simulation
results at 128 MHz calculated with finite element method using
the geometry shown in Fig. 1d and 1f and with either a
single-electrode PtIr wire or an RTS lead. 10 g-avg. SAR in the
ASTM phantom without lead (left), with the PtIr wire (middle), and with the
RTS design that was selected for prototype manufacturing (right). Values
were normalized to whole-body SAR of 2 W/kg. (f) Temperature maps
for 15 minutes of continuous SAR exposure for the same three cases described
in (e). Simulations showed that the RTS lead is transparent to the incident
RF field and generated similar temperature increase (up to 0.9°C)
compared with the ASTM phantom without lead. By contrast, the PtIr wire
generated a temperature increase up to 12°C near the electrode
(please note that the color bar threshold was set to 2°C to
improve visualization.)
Figure 3
Temperature measurements.
(a) Schematic of the two-layer RTS design (electrical conductivity
σ1 and σ2, permittivity
ε1 and ε2, length
L1 and L2 used for the study. (b) ASTM phantom in
the 3 T system used for the temperature measurements. The lead
was placed laterally on the right side of the phantom, on a white plastic
support. (c) Detail showing the temperature sensors placed near the
manufactured RTS prototype. The four PtIr electrodes are visible, with the
probe located on top of one of them. (d) Commercial lead used for the
comparison and placement of the temperature probes near the electrodes. The
probes were placed perpendicular to the lead to minimize error accuracy53. (e) Results of temperature measurements at three different
positions within the phantom without lead, with a Medtronic 3389 lead, and
with the RTS lead. (f) Configuration of battery testing with RTS lead. (g)
Configuration of testing with the Medtronic 3389 lead. Each of the two leads
was connected to a commercial DBS IPG system via an extension. The full
system (i.e., IPG, extension, and leads) was immersed in physiologic
solution for both the RTS and the commercial lead. Battery consumption was
tested over a four-week period for both the Medtronic 3389 and the RTS
leads. Both leads showed a 0.005 V initial drop in battery
voltage, followed by a constant level over the time evaluated.
Figure 4
RTS design in human body model.
(a) X-Ray image of a patient with implanted bilateral DBS system. A head
holder, the implantable pulse generator (IPG) in the thorax, and the two
leads are visible. (b) Anatomical model of a human body with implanted DBS
lead used for numerical simulations. (c) Model of the human body inside the
RF body coil. (d) Model of the lead, including the lumen (A), the four RTS
wires (B), the four electrodes numbered as in the Medtronic 3389 (i.e., 0
– 3) (C), and the insulation (D). (e) Graph showing the results
of analysis of 10 g-avg. SAR reduction with respect to different
RTS configurations (i.e., from two to six sections.) (f) Numerical
simulation results showing a coronal view of power absorption in the human
body model without implant (left), with PtIr lead (middle), and with RTS
lead (right). The increase of power near the electrode for the PtIr lead is
clearly visible. By contrast, the RTS wire is
“RF-transparent” to the RF field (i.e., is the map is
similar to the case without the lead). (g) Sagittal view of the same
results.
Figure 5
PVD-based manufacturing of wire-design RTS prototype.
(a) Two-layer RTS microwire. The RTS (80 μm diameter)
was built with an Ethilon nylon 6–0 suture used as a substrate
and coated with a Ti/Au layer. (b) Optical microscope (OM) image of the
Ethilon nylon 6–0 suture used as a substrate to build the RTS
microwire. The steep transition between the two RTS layers —
necessary for the maximizing the mismatched impedance and the scattering
within the RTS fiber (see Fig. 1b) — is
clearly visible. (c) OM image of the interface between the two RTS layers.
(d) OM image of microwire fully coated with 100/150 nm of Ti/Au.
(e) Scanning Electron Microscope (SEM) image of the RTS microwire. (f) SEM
view, with increased magnification, of the Au coating of the RTS fiber
showing the characteristic fibrous surface of the Ethilon nylon
substrate.
Methods
Theoretical background on RTS design
The RTS implant exposed to an RF field can be represented with a hybrid model
composed of an antenna attached to a transmission line, which consists of
resistive traces with sharp changes in conductivity to maximize reflections,
followed by a load such as an electrode connected to the tissue (Fig. 1a). As described in ref. 32, the
equivalent antenna (i.e., the entire RTS lead) receives the electromagnetic (EM)
field and injects it into the first port (i.e., layer) with impedance
Z1 of such a network (Fig. 1b). A portion
of the power transmitted to the first port of the RTS is reflected back as a
result of an impedance mismatch between the first port and the antenna, while a
remaining portion is supplied to the second layer of the RTS. The impedance of
this second port is intentionally mismatched to reflect the greatest amount of
power back to the implantable pulse generator (IPG) and away from the electrode
that is in contact with the tissue. The fractional power reflected away and
delivered to the tissue can be computed from the reflection and transmission coefficients:One must
consider the superposition of two steady state sine waves in the RTS traveling
in opposite directions (Fig. 1b): one forward towards the
tissue/electrode (blue) and one backward (red) reflected by the mismatched
boundary towards the IPG. The first and second layer of the RTS act both as an
antenna and transmission line for the signal that is reflected back away from
the tissue/electrode.The following equation is the typical shape of the ideal current distribution in
an RTS wire as sketched in Fig. 1c:where L1 and L2 are the
lengths of the first and second layer. A more precise current distribution
estimated using EM numerical simulations is shown in Fig.
2d. The RTS design reduces the overall inductance of the lead (Fig. 2c). Additionally, the current density has a minimum
value along the lead in proximity to the electrode (Fig.
2d), thereby reducing the risk for energy absorption in the
surrounding tissue. Please see the Supplementary Information
for details on the theoretical background on RTS design.
Computational modeling and simulations
A computational model was used to evaluate several possible electrical and
geometrical configurations of the RTS lead to minimize the absorption of energy
and the temperature increase at the electrode. The model included a clinical
3 T MRI RF transmit coil, which operates at 128 MHz39, loaded with a gel-filled phantom and an implanted lead (Fig. 1d and 1f). The design contained discrete sections of
variable conductivity and length, connected in series, with a fixed length
(i.e., to yield a total length of 40 cm to match common lead lengths
for implantable devices13) and a fixed resistance at
low-frequency (i.e., 400 Ω, i.e., less than the typical
impedance in patients40) (Fig. 2a).
Simulations were performed to determine the values of electrical conductivity
(i.e., σ1 and σ2) and length (i.e.,
L1 and L2) for a two-section RTS design (Fig. 3a and Fig. 4e) in order to build
a prototype for experimental testing. The parameter used in the simulations to
evaluate the power absorbed inside the phantom was the specific absorption rate
(SAR) averaged over 10 g of tissue (10 g-avg. SAR). SAR
(W/kg) is a measure of the energy rate absorbed by the human body when exposed
to an RF field and it is the dosimetric parameter used in RF safety
guidelines41. SAR is averaged either over the entire body, or
over 1 g or 10 g of tissue. Temperature simulations on the
final optimized lead design were also performed. Please see the Supplementary Information for details on all models and simulations,
as well as for additional results.
Uncertainty analysis
A simulation study to assess the uncertainty of design and simulation parameters
was performed (Table 1) following the approach used in
Neufeld et al.42. Please see the Supplementary Information for details on the methodology used for the
analysis.
Table 1
Uncertainty analysis. The methods used were based on the work of Neufeld
et al.42. To evaluate the uncertainty of the
quantities of interest derived by the simulations (i.e., 10 g-avg.
SAR or the magnitude of incident electric field
“ERMS”) Two simulations were run for each
parameter by assigning two different values (“Val 1” and
“Val 2”) to each parameter studied. The first value
(“Val 1”) was the one used for the simulations shown in
Fig. 2, whereas the modified value (“Val
2”) was set to a realistic value that could occur due to either
design choice or manufacturing tolerance. The results obtained for each value
(“Result 1” and “Result 2”,
respectively) were used to evaluate sensitivity factor of the quantity evaluated
(10 g-avg. SAR or magnitude of incident electric field
“ERMS”). The standard deviation
(“Std. Dev. ”) was derived from literature
Parameter
Quantity
Val 1
Val 2
Result 1
Result 2
Sensitivity Factor
[%/mm]
Std. Dev
Uncertainty [%]
Contact Width [mm]
10 g-avg. SAR at electrode [W/kg]
0.381
0.762
6.98
7.22
9.12%
0.1
0.9%
Contact Length [mm]
1.5
3.0
6.98
7.28
2.85%
0.1
0.3%
Contact Thickness [mm]
0.0098
0.0196
6.98
7.00
32.6%
0.1
3.3%
Substrate Thickness [mm]
0.0254
0.0508
6.98
6.9
43.9%
0.1
4.4%
Insulation Thickness [mm]
0.0254
0.0508
6.98
7.07
52.0%
0.1
5.2%
εr (Substrate)
3.4
6.8
6.98
7.11
0.56%
2.00
1.1%
εr (Insulation)
2.5
5.0
6.98
7.00
0.13%
2.00
0.3%
εr (Ink Lead)
5.0
2.5
6.98
6.99
0.04%
2.00
0.1%
σ (Contact) [S/m]
9.3·106
4.0·106
6.98
6.99
0.00%
0.04
0.0%
εr (Phantom)
80
60
6.98
7.40
0.30%
2.00
0.6%
σ (Phantom) [S/m]
0.47
0.60
6.98
7.18
22.3%
0.04
0.9%
Phantom Position X [mm]
ERMS, incident [V/m]
0.0
10.0
300.9
302.2
0.04%
1.15
0.1%
Phantom Position Y [mm]
0.0
10.0
300.9
306.9
0.20%
1.15
0.2%
Phantom Position Z [mm]
0.0
10.0
300.9
307.0
0.20%
1.15
0.2%
Lead Position X [mm]
0.0
1.0
300.9
307.4
2.14%
0.58
1.2%
Lead Position Y [mm]
0.0
1.0
300.9
299.2
0.57%
0.58
0.3%
Lead Position Z [mm]
0.0
1.0
300.9
301.3
0.12%
0.58
0.1%
Manufacturing of flat-design RTS prototype
Based on the optimal parameters of the RTS design's conductivity and
length derived from simulations (see Supplementary
Table S1), a flat-design lead prototype was built using
polymer thick-film (PTF) technology to experimentally test the proposed concept
(see Fig. 3a). The lead was built by printing thin
(10 μm) layers of two different commercially available
conductive inks on a polymer substrate for the length of each of the two
resistive layers. The dimensions of the RTS lead were chosen to obtain the same
volume of the wire in the Medtronic 3389 lead (see Supplementary
Table S1). Please see the Supplementary
Information for additional details on the manufacturing of the
two-section flat-design RTS lead prototype.
Temperature measurements in MRI
The RTS prototype was built by stacking four of the flat-design leads connected
to four electrodes (Fig. 3c) and by insulating the
proximal end (opposite to the electrodes). The RTS prototype was then implanted
in a standard ASTM phantom filled with a polyacrylic acid (PAA) mixed in an
aqueous solution23 and tested in a 3 T MRI system
(Skyra, Siemens, Erlangen, Germany) (Fig. 3b). Three fiber
optic temperature probes (Neoptix Inc., Quebec, Canada) were used to record the
temperature in the phantom and along the lead under several conditions: no lead,
RTS lead, and Medtronic 3389 lead (Fig. 3d). RF energy was
delivered to the phantom at First Level Controlled Operating Mode for
15 minutes. Please see the Supplementary
Information for additional details on the temperature
measurements.
Implantable pulse generator (IPG) battery testing
Battery testing was also performed with both leads (flat-design RTS and Medtronic
3389) (Fig. 3f and 3g) connected to an IPG (i.e.,
Medtronic Activa® PC) through an extension. The IPG, extension, and
lead were then placed in a quart of deionized water mixed with saline solution
to simulate in-body tissue impedance. The IPG was turned on for a total of four
weeks. Please see the Supplementary Information for
additional details on the IPG battery testing.
Manufacturing of wire-design RTS prototype
A second RTS wire-design prototype was manufactured using thin-film physical
vapor deposition (PVD) of titanium and gold over a rotating Ethilon®
6–0 suture substrate, which was selected for its biocompatibility.
Variation in the impedance of each segment is achieved by control of the
thickness of the gold layer of each segment. Please see the Supplementary Information for additional details regarding the
manufacturing of the wire-design RTS prototype.
Results
Electromagnetic simulations
The length of each individual section of the RTS lead affected the
10 g-avg. SAR near the electrode non-linearly (Fig.
2a). Numerical simulations were repeated to observe the correlation
between RTS design and 10 g-avg. SAR by fixing the length of each
section and varying the ratio in conductivity
(σ1/σ2) between the two sections
of the lead, while maintaining a total length of 40 cm. Figure 2b shows the 10 g-avg. SAR in the phantom
near the electrode with a resistance varying from 0 Ω to
1 kΩ for several RTS designs. The RTS lead reduced the
10 g-avg. SAR across the entire range of resistances. The simulations
showed an increase in conductivity ratio between the two sections that
corresponded to a decrease in 10 g-avg. SAR at the electrode. For
example, the optimal RTS (i.e.,
σ1/σ2 = 200) plateaued at
400 Ω with a value of 4.02 W/kg, whereas the
design with σ1/σ2 = 2 showed a
10 g-avg. SAR of 5.75 W/kg at 400 Ω.
For comparison, the peak 10 g-avg. SAR with a 40 cm PtIr
wire was 222 W/kg. Please see Supplementary
Fig. S1 for additional maps of electric and magnetic field
magnitude with RTS and PtIr wire.The SAR reduction was due to a lower inductance of the RTS design (Fig. 2c) (see also “theoretical background on RTS
design” in Methods and Supplementary Information)
that corresponded to a shorter equivalent antenna length and lower induced
currents. Simulations with a single-section platinum-iridium (PtIr) wire of the
same length (40 cm) were also performed for reference. As confirmed
by the simulations, the RTS design was characterized by a reduced current at the
electrode of over two orders of magnitude compared with the PtIr wire (Fig. 2d). The high electrical conductivity of the ink used
for manufacturing allowed a prototype to be built with the following
characteristics: σ1 = 1.968·106 S/m, σ2 = 25.61·103 S/m (i.e., σ1/σ2 =
76.86), L1 = 0.367 m, and L2 =
0.033 m. The total resistance for the RTS design was chosen to be R =
400 Ω, five times less than the maximum electrode/tissue
impedance of 2 kΩ allowed by even older IPG models40. As shown in Figs. 2a and 2b, the
10 g-avg. SAR of this configuration was expected to be very similar
to the best performance of the RTS lead with ratio
σ1/σ2 = 200 (i.e.,
4.1 W/kg vs. 4.02 W/kg, respectively) (for discussion of
the case σ1/σ2 = 1 as well as other
additional cases, please see Supplementary
Fig. S2).
Temperature simulations
Figure 2 shows the 10 g-avg. SAR (Fig. 2e) and temperature maps (Fig. 2f)
recorded in the phantom model under three conditions: without implant, with the
RTS design selected for prototype manufacturing, and with the PtIr wire. The SAR
and temperature maps, which are plotted throughout the plane containing the
lead, show similar results between the phantom with the RTS lead vs. the phantom
without implant. The peak 10 g-avg. SAR was less than
7 W/kg, and temperature changes were below 1°C in both
cases for a 15-minute exposure at a whole-body SAR of 2 W/kg. By
contrast, the simulations with the PtIr wire model predicted a peak
10 g-avg. SAR of 230 W/kg and temperature change of
64°C for the same exposure. As a reference, the value of
2 W/kg is the limit in Normal Operating Mode for SAR averaged over
the entire body, as established by the current guidelines of the International
Electrotechnical Commission (IEC)41.
Temperature measurements
The temperature increase near the electrode of the Medtronic 3389 lead was about
9°C higher than the baseline level of the phantom without a lead and
2°C near the middle of the lead. Conversely, the temperature increase
of the RTS lead was 3°C around the electrode and less than
4°C near the middle of the lead (Fig. 3e).
These temperature values were consistent with the energy distributions predicted
by the simulations (see Supplementary Fig. S3)
suggesting a decrease of current at the electrode and an increase of current
along the lead. For comparison, the baseline temperature increase of the phantom
without an implant was 1.5°C at the location corresponding to the
electrode and 1°C at the location corresponding to the middle of the
lead (Fig. 3e). Given the linear relationship between SAR
and temperature, the corresponding maximum temperature increases at
2 W/kg would be less than 4.5°C with the Medtronic 3389
lead and less than 2°C with the RTS. For reference, the level of
temperature increase suggested by an international safety standard for patients
with implantable neurostimulators is 2°C43, which the
RTS lead met in the experimental setup used in this study.
Battery measurements
The longevity of the Activa PC Neurostimulator battery can last for months to
years depending on the following factors: programmed stimulation parameters, the
total system impedance and the hours per day the battery is in use. The
Medtronic Battery Longevity Manual44 provides a formula that
estimates the approximate period of time that an Activa PC battery can last. The
formula utilizes the aforementioned factors to calculate an estimated energy use
of the battery in a 24 hour period, which can then be used with a
look-up chart (see Fig. 2 in ref. 44) to predict battery longevity in years. For the battery
testing conducted in this investigation, the programmed stimulation parameters,
system impedance and hours per day of stimulation were fixed at the same values
for both the RTS prototype and the Medtronic 3389 lead. Accordingly, longevity
estimates for both leads will be the same. To assess whether actual battery
consumption would correspond with such a prediction, a preliminary comparative
test was performed by connecting the RTS prototype (Fig.
3f) and the Medtronic 3389 lead (Fig. 3g) to
the Activa PC Neurostimulator (Medtronic, Inc., Minneapolis, MN). Over a 30 day
testing period, both leads showed only a 0.005 V drop in battery
voltage, a result that correlates with Medtronic estimation methods. This
indicates that the RTS design affects only the behavior of the lead with respect
to RF (i.e., the reactance) and not with respect to the operational frequencies
of the stimulator (see also theoretical background on RTS design in Supplementary Information).
Discussion
This study presents a novel metamaterial38 lead that reduces the
antenna effect and allows for decreased tissue heating during MRI45.
The optimal parameters of the design were determined by computational modeling and
simulations, validated against in-vitro temperature measurements in a gel-filled
phantom (Supplementary Fig. S3). The numerical
simulations confirmed that a PtIr wire acts as an antenna during the RF transmit
period of the MRI scan, picking up the induced electric field and transferring a
high amount of RF energy into the volume surrounding the exposed electrode tip. In
both simulations and in-vitro testing the proposed RTS design successfully reduced
the amount of energy absorbed and the related temperature increases inside the
gel-phantom in proximity to the electrode. Numerical simulations and experimental
testing confirmed that the RTS design allows for
“RF-cloaking”38 while maintaining proper
low-frequency conductivity that does not affect battery performance.The experimental bench testing confirmed also the practical feasibility of the RTS
design. The primary feature of the RTS is the abrupt change of conductivity between
the two sections. While this discontinuity can be easily modeled computationally,
issues can arise in a prototype, because the RTS needs to be built using different
inks with different electrical properties. In practice, the two ink traces of the
two different layers cannot be perfectly contiguous along the RTS; an overlap is
always present which reduces the transition between the layers and, therefore, the
ideal step discontinuity in electrical conductivity. The experimental testing
confirmed that the prototype contained an adequate discontinuity between the two
layers with a physical overlap between the two layers that was only about
50 µm along the RTS (see Fig. 5e).
Additionally, the proposed RTS design does not require any external physical device
such as an RF choke. RF chokes are difficult to attach to an implant wire because
the dimensions of a choke are larger than the typical dimension of the wire. Chokes
also disrupt the mechanical characteristics of an implant, which should be
flexible46. Although there are extremely miniaturized RF chokes,
these devices can be more prone to burning because of the microscopic physical
dimensions of their components.In order to enhance the signal-to-noise ratio of the measurements, the testing was
performed with high levels of RF power, namely one corresponding to a whole-body SAR
of 4 W/kg as estimated by the MRI system. Most sequences used in MRI
systems are characterized by a whole-body SAR of less than 2 W/kg. SAR
estimation varies for each MRI manufacturer and across systems and depends on
several variables, including coil specifications, landmarks, and patient
registration information, e.g., weight, height, age. Baker and colleagues47 compared the RF-induced heating per unit of SAR due to the presence
of a DBS lead between two 1.5 T MRI systems and observed values ranging
from 3.5 to 5.5 times higher on one MRI system as compared to the other.
As such, the absolute temperature values found in this study are specific to the MRI
system used47. Additionally, the measurements were performed in a
gel-filled phantom12133147. This approach implies lack of
perfusion and does not take into consideration possible changes due to the
thermoregulatory response in a patient48. Such a scenario is
typically considered a worst-case, because perfusion can reduce significantly, e.g.,
two-fold, the heating of tissue in proximity of the lead49.In this study, heating in the proximity of DBS implants, induced by the RF excitation
pulses during the MRI, was measured by fluoroptic temperature sensors, which are the
“state-of-the-art” in this field314750515253. Another common approach for measuring
temperature is MRI thermometry, which allows controlled heating while simultaneously
measuring the spatial and temporal temperature distribution near the DBS implant.
The most common MRI-thermometry method is based on proton resonance frequency shift
(PRFS), which has been used to monitor temperature near a wire54.
However, the susceptibility artifact from the DBS implant may extend up to
5 mm from the electrode surface, and at this distance the temperature
changes are significantly lower than the peak temperature change49.
Additional approaches were proposed to alleviate the susceptibility artifacts around
a wire, but they did not provide real-time measurements at the desired high spatial
resolution5556, underestimating the peak temperature change.
Conversely, fluoroptic thermometers can provide accurate and real time temperature
measurements with a spatial resolution of typically
300 μm53.The configuration of the implanted DBS components relative to the incident RF field
and its orientation can also have a dramatic effect on the induced heating. In this
study, the lead orientation was limited to a single case of overall lead length and
path within the phantom42, i.e., lead placed parallel to the magnet
bore axis (Fig. 1f). This allowed for the evaluation of the
PtIr wire and the proposed RTS lead under the same conditions of high incident
electric field (Fig. 1e) inside the homogenous phantom used in
the study. However, the layout used does not necessarily model the exposure
conditions of a lead implanted in a patient57, nor does it take into
consideration differences between a single vs. bilateral lead. For example, a change
in orientation of the lead with respect to a 1.5 T RF coil can generate
changes in temperature of 20°C or more in a phantom58. A
full systematic analysis of safety of the proposed RTS lead is still required and
would need to include several configurations that would mimic clinically significant
pathways, in line with the technical specifications proposed for safety analysis of
patients with active implanted devices undergoing MRI1259.
Additional numerical simulations were performed with an electrically heterogeneous,
anatomically precise human head and torso model6061 containing a
DBS lead reproducing a clinical case (Fig. 4b) — as
shown by CT imaging (Fig. 4a) — for testing under
different exposure and geometrical conditions. The model was placed with the head in
the isocenter of an MRI coil (Fig. 4c). Figures
4f and 4g show a coronal and sagittal view the SAR in the head and body
without an implanted lead, with a lead made of platinum iridium wire, and with an
optimized wire-design RTS lead. These results also confirmed the advantage of the
RTS design which significantly reduces absorbed power in the brain parenchymal near
the electrode.The benefits of the electrically thin design with its scattering behavior (see
theoretical background on RTS design in Supplementary
Information) can be used to replace any wire currently used in
commercially available implant leads by coating a suture with biocompatible metals.
Hence, a second more realistic and biocompatible RTS wire prototype (Thin-Films
Research Inc., Westford MA) was created (Fig. 5a). Optical
microscope (OM) images show the raw suture (Fig. 5b) before
thin film deposition and after deposition (Fig. 5d). The
abrupt transition of electrical conductivity between the two RTS layers (Fig. 5c) maximizes the mismatched impedance and the scattering
within the fibers (Fig. 1 and Supplementary
Information, theoretical background on RTS design). The two different
layers of the RTS fiber and the surface characteristics were also studied with a
Scanning Electron Microscope (SEM) (Fig. 5e and Fig. 5f) in order to better characterize the transition between the two
layers.The final RTS lead configuration is assembled similarly to the configuration of the
Medtronic 3389 DBS lead (Fig. 4d) and thus it may be used
(i.e., interchangeable) with the Medtronic Activa stimulator. Each lead consists of
an implantable grade polyurethane inner shaft, into which the stylet is advanced,
that terminates at the hermetically sealed distal tip of the lead. The RTS fibers
are positioned around the inner shaft and contained within a protective sheath that
has four platinum/iridium electrodes near the tip for delivery of stimulation to the
target site. The proximal end of the lead also has four electrodes that interface
with the implanted stimulation device after implantation. The leads are
stereotactically introduced into the target and fixed at the skull with a burr hole
cap and ring, as for the Medtronic DBS leads.
Conclusions
This study presents a novel resistive-tapered stripline (RTS) lead design that
“cloaks” the radiofrequency fields induced by magnetic
resonance imaging (MRI) to reduce tissue heating, yet maintains the conduction
of low-frequency stimulation from implanted medical devices. Computational
modeling and simulations were used to find the optimal design parameters of the
RTS lead. Polymer thick-film (PTF) technology was used to manufacture an initial
prototype, which was tested in a 3 T MRI system showing a significant
reduction of heating when compared to a Medtronic 3389 lead. Finally,
state-of-the-art physical vapor deposition (PVD) technology was used to
manufacture a biocompatible RTS wire prototype, which may easily replace any
wire currently used in commercially available implant leads. The results shown
suggest the proposed design may allow a significant increase in the number of
patients with medical implants having safe access to the diagnostic benefits of
MRI.
Authors: L Lemieux; A Salek-Haddadi; O Josephs; P Allen; N Toms; C Scott; K Krakow; R Turner; D R Fish Journal: Neuroimage Date: 2001-09 Impact factor: 6.556
Authors: Ali R Rezai; Daniel Finelli; John A Nyenhuis; Greg Hrdlicka; Jean Tkach; Ashwini Sharan; Paul Rugieri; Paul H Stypulkowski; Frank G Shellock Journal: J Magn Reson Imaging Date: 2002-03 Impact factor: 4.813
Authors: A Sharan; A R Rezai; J A Nyenhuis; G Hrdlicka; J Tkach; K Baker; M Turbay; P Rugieri; M Phillips; F G Shellock Journal: Acta Neurochir Suppl Date: 2003
Authors: Leonardo M Angelone; Andreas Potthast; Florent Segonne; Sunao Iwaki; John W Belliveau; Giorgio Bonmassar Journal: Bioelectromagnetics Date: 2004-05 Impact factor: 2.010
Authors: Jörg Spiegel; Gerhard Fuss; Martin Backens; Wolfgang Reith; Tim Magnus; Georg Becker; Jean-Richard Moringlane; Ulrich Dillmann Journal: J Neurosurg Date: 2003-10 Impact factor: 5.115
Authors: Laleh Golestanirad; Maria Ida Iacono; Boris Keil; Leonardo M Angelone; Giorgio Bonmassar; Michael D Fox; Todd Herrington; Elfar Adalsteinsson; Cristen LaPierre; Azma Mareyam; Lawrence L Wald Journal: Neuroimage Date: 2016-12-21 Impact factor: 6.556
Authors: Laleh Golestanirad; Boris Keil; Leonardo M Angelone; Giorgio Bonmassar; Azma Mareyam; Lawrence L Wald Journal: Magn Reson Med Date: 2016-04-05 Impact factor: 4.668
Authors: Laleh Golestanirad; Leonardo M Angelone; Maria Ida Iacono; Husam Katnani; Lawrence L Wald; Giorgio Bonmassar Journal: Magn Reson Med Date: 2016-10-31 Impact factor: 4.668
Authors: Laleh Golestanirad; Leonardo M Angelone; John Kirsch; Sean Downs; Boris Keil; Giorgio Bonmassar; Lawrence L Wald Journal: IEEE Trans Microw Theory Tech Date: 2019-01-01 Impact factor: 3.599
Authors: Alexandre Boutet; Clement T Chow; Keshav Narang; Gavin J B Elias; Clemens Neudorfer; Jürgen Germann; Manish Ranjan; Aaron Loh; Alastair J Martin; Walter Kucharczyk; Christopher J Steele; Ileana Hancu; Ali R Rezai; Andres M Lozano Journal: Radiology Date: 2020-06-23 Impact factor: 11.105