A Thomas1, H Yan1, M Oldham1, T Juang1, J Adamovics2, F F Yin1.
Abstract
Six base of skull IMRT treatment plans were delivered to 3D dosimeters within the RPC Head and Neck Phantom for QA verification. Isotropic 2mm 3D data was obtained using the DLOS-PRESAGE system and compared to an Eclipse (Varian) treatment plan. Normalized Dose Distribution pass rates were obtained for a number of criteria. High quality 3D dosimetry data was observed from the DLOS system, illustrated here through colormaps, isodose lines, profiles, and NDD 3D maps. Excellent agreement with the planned dose distributions was also observed with NDD analysis revealing > 90% NDD pass rates [3%, 2mm], noise < 0.5%. This paper focuses on a detailed exploration of the quality and use of 3D dosimetry data obtained with the DLOS-PRESAGE system.Entities:
Year: 2013 PMID: 26877756 PMCID: PMC4752441 DOI: 10.1088/1742-6596/444/1/012049
Source DB: PubMed Journal: J Phys Conf Ser ISSN: 1742-6588
Figure 1CIRS motion phantom (left) and dosimeter insert (middle) with CT of dosimeter and GTV/CTV and ITV/PTV volumes contoured (right).
Figure 2DVHs of (A) The RapidArc plan and (B) The IMRT plan. In both instances dashed lines indicate GTV/CTV coverage and solid lines represent ITV/PTV coverage.
Figure 3Static (Top Row) and Motion (Bottom Row) delivery overlaying the planned Rapid Arc and IMRT deliveries. The 1st column is the coronal view and the 2nd is the axial view. The colorbar indicates dose delivered in % of maximum planned dose. In all plots above the solid lines equal the planned doses and the dashed lines equal the measured doses.