Commissioning and benchmarking a 3D dosimetry system for clinical use.

Andrew Thomas1, Joseph Newton, John Adamovics, Mark Oldham.   

Abstract

PURPOSE: A 3D dosimetry system is described which consists of two parts: a radiochromic plastic dosimeter PRESAGE (which responds to absorbed dose with a linear change in optical-density) and the Duke large-field-of-view optical-CT scanner (DLOS). The DLOS/PRESAGE system has recently been commissioned and benchmarked for clinical use and, in particular, for verification and commissioning of complex radiation treatments.
METHODS: DLOS commissioning involved determining the dynamic range, spatial resolution, noise, temporal, and other characteristics of the light source and imaging components. Benchmarking tests were performed on the combined DLOS/PRESAGE system to establish baseline dosimetric performance. The tests consisted of delivering simple radiation treatments to PRESAGE dosimeters, and comparing the measured 3D relative dose distributions with the known gold standard. The gold standard distribution was obtained from machine beam-data or the treatment planning system (TPS). All studies used standardized procedures to ensure consistency.
RESULTS: For commissioning, isotropic spatial resolution was submillimeter (MTF > 0.5 for frequencies of 1.5 lp/mm) and the dynamic range was -60 dB. Flood field uniformity was within 10% and stable after 45 min of warm-up. Stray-light is small, due to telecentricity, but even the residual can be removed through deconvolution by a point-spread-function. For benchmarking, the mean 3D passing NDD (normalized dose distribution) rate (3%, 3mm, 5% dose threshold) over the benchmark data sets was 97.3% +/- 0.6% (range 96%-98%), which is on par with other planar dosimeters used in external beam radiation therapy indicating excellent agreement. Noise was low at < 2% of maximum dose (4-12 Gy) for 2 mm reconstructions. The telecentric design was critical to enabling fast imaging with minimal stray-light artifacts.
CONCLUSIONS: This work presents the first comprehensive benchmarking of a 3D dosimetry system for clinical use. The DLOS/PRESAGE benchmark tests show consistently good agreement to simple known distributions. The system produces accurate isotropic 2 mm dose data over clinical volumes (e.g., 16 cm diameter phantoms, 12 cm height), in under 15 min. It represents a uniquely useful and versatile new tool for commissioning and verification of complex therapy treatments.

Entities:  

Mesh:

Year:  2011        PMID: 21928656      PMCID: PMC3165434          DOI: 10.1118/1.3611042

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  40 in total

1.  A CCD-based optical CT scanner for high-resolution 3D imaging of radiation dose distributions: equipment specifications, optical simulations and preliminary results.

Authors:  S J Doran; K K Koerkamp; M A Bero; P Jenneson; E J Morton; W B Gilboy
Journal:  Phys Med Biol       Date:  2001-12       Impact factor: 3.609

2.  An Investigation into the Robustness of Optical-CT Dosimetry of a Radiochromic Dosimeter Compatible with the RPC Head-and-Neck Phantom.

Authors:  H S Sakhalkar; J Adamovics; G Ibbott; M Oldham
Journal:  J Phys Conf Ser       Date:  2009-01-01

3.  A practical three-dimensional dosimetry system for radiation therapy.

Authors:  Pengyi Guo; John Adamovics; Mark Oldham
Journal:  Med Phys       Date:  2006-10       Impact factor: 4.071

4.  On dose distribution comparison.

Authors:  Steve B Jiang; Greg C Sharp; Toni Neicu; Ross I Berbeco; Stella Flampouri; Thomas Bortfeld
Journal:  Phys Med Biol       Date:  2006-01-25       Impact factor: 3.609

5.  IMRT verification using a radiochromic/optical-CT dosimetry system.

Authors:  Mark Oldham; Pengyi Guo; Gary Gluckman; John Adamovics
Journal:  J Phys Conf Ser       Date:  2006

6.  Investigation of the feasibility of relative 3D dosimetry in the Radiologic Physics Center Head and Neck IMRT phantom using presage/optical-CT.

Authors:  Harshad Sakhalkar; David Sterling; John Adamovics; Geoffrey Ibbott; Mark Oldham
Journal:  Med Phys       Date:  2009-07       Impact factor: 4.071

7.  Investigation of radiological properties and water equivalency of PRESAGE dosimeters.

Authors:  Tina Gorjiara; Robin Hill; Zdenka Kuncic; John Adamovics; Stephen Bosi; Jung-Ha Kim; Clive Baldock
Journal:  Med Phys       Date:  2011-04       Impact factor: 4.071

8.  Evaluation of polymer gels and MRI as a 3-D dosimeter for intensity-modulated radiation therapy.

Authors:  D A Low; J F Dempsey; R Venkatesan; S Mutic; J Markman; E Mark Haacke; J A Purdy
Journal:  Med Phys       Date:  1999-08       Impact factor: 4.071

9.  A method to correct for stray light in telecentric optical-CT imaging of radiochromic dosimeters.

Authors:  Andrew Thomas; Joseph Newton; Mark Oldham
Journal:  Phys Med Biol       Date:  2011-06-30       Impact factor: 3.609

Review 10.  Polymer gel dosimetry.

Authors:  C Baldock; Y De Deene; S Doran; G Ibbott; A Jirasek; M Lepage; K B McAuley; M Oldham; L J Schreiner
Journal:  Phys Med Biol       Date:  2010-02-11       Impact factor: 3.609

View more
  29 in total

1.  PRESAGE 3D dosimetry accurately measures Gamma Knife output factors.

Authors:  Slade J Klawikowski; James N Yang; John Adamovics; Geoffrey S Ibbott
Journal:  Phys Med Biol       Date:  2014-11-04       Impact factor: 3.609

2.  Commissioning a small-field biological irradiator using point, 2D, and 3D dosimetry techniques.

Authors:  Joseph Newton; Mark Oldham; Andrew Thomas; Yifan Li; John Adamovics; David G Kirsch; Shiva Das
Journal:  Med Phys       Date:  2011-12       Impact factor: 4.071

3.  A comprehensive investigation of the accuracy and reproducibility of a multitarget single isocenter VMAT radiosurgery technique.

Authors:  Andrew Thomas; Michael Niebanck; Titania Juang; Zhiheng Wang; Mark Oldham
Journal:  Med Phys       Date:  2013-12       Impact factor: 4.071

4.  An investigation of a PRESAGE® in vivo dosimeter for brachytherapy.

Authors:  A K Vidovic; T Juang; S Meltsner; J Adamovics; J Chino; B Steffey; O Craciunescu; M Oldham
Journal:  Phys Med Biol       Date:  2014-06-24       Impact factor: 3.609

5.  Comprehensive quality assurance for base of skull IMRT.

Authors:  A Thomas; J O'Daniel; J Adamovics; G Ibbott; M Oldham
Journal:  J Phys Conf Ser       Date:  2013

6.  On the feasibility of polyurethane based 3D dosimeters with optical CT for dosimetric verification of low energy photon brachytherapy seeds.

Authors:  Justus Adamson; Yun Yang; Titania Juang; Kelsey Chisholm; Leith Rankine; John Adamovics; Fang Fang Yin; Mark Oldham
Journal:  Med Phys       Date:  2014-07       Impact factor: 4.071

7.  Investigations into the feasibility of optical-CT 3D dosimetry with minimal use of refractively matched fluids.

Authors:  Kelsey Chisholm; Devin Miles; Leith Rankine; Mark Oldham
Journal:  Med Phys       Date:  2015-05       Impact factor: 4.071

8.  Optical cone beam tomography of Cherenkov-mediated signals for fast 3D dosimetry of x-ray photon beams in water.

Authors:  Adam K Glaser; Jacqueline M Andreozzi; Rongxiao Zhang; Brian W Pogue; David J Gladstone
Journal:  Med Phys       Date:  2015-07       Impact factor: 4.071

9.  A quality assurance method that utilizes 3D dosimetry and facilitates clinical interpretation.

Authors:  Mark Oldham; Andrew Thomas; Jennifer O'Daniel; Titania Juang; Geoffrey Ibbott; John Adamovics; John P Kirkpatrick
Journal:  Int J Radiat Oncol Biol Phys       Date:  2012-02-22       Impact factor: 7.038

10.  Commissioning a CT-compatible LDR tandem and ovoid applicator using Monte Carlo calculation and 3D dosimetry.

Authors:  Justus Adamson; Joseph Newton; Yun Yang; Beverly Steffey; Jing Cai; John Adamovics; Mark Oldham; Junzo Chino; Oana Craciunescu
Journal:  Med Phys       Date:  2012-07       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.