Jung-Chang Sun1,2, Bor-Tsung Hsieh2, Chih-Ming Hsieh3, Yuk-Wah Tsang1,4, Kai-Yuan Cheng2.
Abstract
BACKGROUND: The gel dosimeter is a chemical as well as a relative dosimeter.Entities:
Keywords: MRI; NIPAM gel dosimeter; dynamic dose effects
Mesh:
Substances:
Year: 2022 PMID: 35124616 PMCID: PMC9028750 DOI: 10.3233/THC-THC228038
Source DB: PubMed Journal: Technol Health Care ISSN: 0928-7329 Impact factor: 1.205
The cylindrical flask irradiation settings in this study
| Parameter | Set value |
|---|---|
| Gantry angle (degrees) | 0 |
| Photon energy (MV) | 6 |
| Irradiation dose (Gy) | 5 |
| Dose rate (cGy/min) | 500 |
| SSD (cm) | 95 |
| SAD (cm) | 100 |
Figure 1.Placing positions of the gel dosimeter and reciprocating phantom.
Figure 2.Prognostic image of the reciprocating dynamic simulation. The simulated prognostic results showed that the gray area had the most doses.
MRI scan parameters in this study
| Parameter | Set value |
|---|---|
| Repetition time (ms) | 3,000 |
| Echo time (ms) | 22, 44, 66, 88, 110, 132, 154, 176, 198, 220, 242, 264, 286, 308, 330, 352 |
| FOV (mm | 240 |
| Matrix size | 512 |
| Slice thickness (mm) | 5 |
| Gap (mm) | 0.5 |
Figure 3.Comparison between the reading results and TPS dose curve. From top to bottom: comparison results of (a) 4 5 cm, (b) 4 4 cm, (c) 4 3 cm.
Figure 4.Gamma map of reading results and TPS. From top to bottom: 3%/3 mm, gamma maps of (a) 4 5 cm, (b) 4 4 cm, (c) 4 3 cm.
Figure 5.Comparison between reading results and TPS 4 4 cm 60% dose curve.
Figure 6.3%/3 mm gamma map of reading results and TPS 4 4 cm 60% dose curve.