Evaluation of dental implant insertion torque using a manual ratchet.

M M Goswami1, Mukul Kumar2, Abhinav Vats3, A S Bansal4.   

Abstract

BACKGROUND: Dental implant insertion torque is crucial for the success of the implant and the prosthesis. This in-vivo study was undertaken to determine the average insertion torque being applied to the dental implant while surgically placing it with a non-calibrated manual ratchet.
METHODS: Three dental surgeons placed a total of 45 dental implants (Touareg, ADIN, Afula, Israel) in 42 selected patients. Each surgeon placed 15 implants. Standardised protocols were followed to prepare the site to place the dental implant. Each implant was placed using a manual non-calibrated implant ratchet first. Once the implant was nearly placed, a manual calibrated torque gauge ratchet was used to place the implant in its final position and at that instance, the maximum final torque applied was noted on the torque gauge scale.
RESULTS: The mean dental implant insertion torque applied by three surgeons using a non-calibrated manual ratchet was estimated to be 63.26 Ncm with a standard error of 6.80 i.e. (63.26 + 6.8), which was significantly higher than the baseline of 35 Ncm (p < 0.0001). The mean dental implant torque applied by Surgeon 1, 2 and 3, respectively, was 65.93 Ncm, 62.60 Ncm and 62.13 Ncm and this difference amongst them was found to be statistically insignificant (p > 0.05) and each of them had reached more than the baseline level of 35 Ncm individually and significantly (p < 0.0001).
CONCLUSION: Without the use of torque measuring devices, an average surgeon may achieve an average insertion torque of 63.26 + 6.8 Ncm.

Entities:  

Keywords:  Dental implant; Insertion torque; Primary stability

Year:  2013        PMID: 26843747      PMCID: PMC4705154          DOI: 10.1016/j.mjafi.2013.07.010

Source DB:  PubMed          Journal:  Med J Armed Forces India        ISSN: 0377-1237


  18 in total

1.  Measurements comparing the initial stability of five designs of dental implants: a human cadaver study.

Authors:  D O'Sullivan; L Sennerby; N Meredith
Journal:  Clin Implant Dent Relat Res       Date:  2000       Impact factor: 3.932

2.  Influence of implant taper on the primary and secondary stability of osseointegrated titanium implants.

Authors:  Dominic O'Sullivan; Lars Sennerby; Neil Meredith
Journal:  Clin Oral Implants Res       Date:  2004-08       Impact factor: 5.977

3.  Meta-analysis of methods used to assess implant stability.

Authors:  Murat Cavit Cehreli; Durdu Karasoy; Kivanc Akca; Steven E Eckert
Journal:  Int J Oral Maxillofac Implants       Date:  2009 Nov-Dec       Impact factor: 2.804

4.  Correlation between placement torque and survival of single-tooth implants.

Authors:  Judith Maria Pinheiro Ottoni; Zilda Fagundes Lima Oliveira; Roberto Mansini; Antonio Melo Cabral
Journal:  Int J Oral Maxillofac Implants       Date:  2005 Sep-Oct       Impact factor: 2.804

Review 5.  State of the art of oral implants.

Authors:  Tomas Albrektsson; Lars Sennerby; Ann Wennerberg
Journal:  Periodontol 2000       Date:  2008       Impact factor: 7.589

6.  The effect of implant design on insertion torque and immediate micromotion.

Authors:  Amilcar C Freitas; Estevam A Bonfante; Gabriela Giro; Malvin N Janal; Paulo G Coelho
Journal:  Clin Oral Implants Res       Date:  2011-03-23       Impact factor: 5.977

7.  Influence of high insertion torque on implant placement: an anisotropic bone stress analysis.

Authors:  Bruno Salles Sotto-Maior; Eduardo Passos Rocha; Erika Oliveira de Almeida; Amilcar Chagas Freitas-Júnior; Rodolfo Bruniera Anchieta; Altair Antoninha Del Bel Cury
Journal:  Braz Dent J       Date:  2010

8.  Osseointegrated implants in the treatment of the edentulous jaw. Experience from a 10-year period.

Authors:  P I Brånemark; B O Hansson; R Adell; U Breine; J Lindström; O Hallén; A Ohman
Journal:  Scand J Plast Reconstr Surg Suppl       Date:  1977

9.  Effect of drilling dimension on implant placement torque and early osseointegration stages: an experimental study in dogs.

Authors:  Felipe E Campos; Julio B Gomes; Charles Marin; Hellen S Teixeira; Marcelo Suzuki; Lukasz Witek; Darceny Zanetta-Barbosa; Paulo G Coelho
Journal:  J Oral Maxillofac Surg       Date:  2012-01       Impact factor: 1.895

10.  A comparison between cutting torque and resonance frequency in the assessment of primary stability and final torque capacity of standard and TiUnite single-tooth implants under immediate loading.

Authors:  Hiron Andreaza da Cunha; Carlos Eduardo Francischone; Hugo Nary Filho; Rubelisa Cândido Gomes de Oliveira
Journal:  Int J Oral Maxillofac Implants       Date:  2004 Jul-Aug       Impact factor: 2.804

View more
  2 in total

1.  Reliability and accuracy of the torque applied to osteosynthesis screws by maxillofacial surgeons and residents.

Authors:  Barzi Gareb; Valerie D M van Munster; Pieter U Dijkstra; Ruud R M Bos; Arjan Vissink; Nico B van Bakelen; Baucke van Minnen
Journal:  Sci Rep       Date:  2022-08-24       Impact factor: 4.996

Review 2.  Influence of different implant placement techniques to improve primary implant stability in low-density bone: A systematic review.

Authors:  Jayant N Palaskar; Nikhil Joshi; Pooja M Shah; Poorva Gullapalli; Vineet Vinay
Journal:  J Indian Prosthodont Soc       Date:  2020-01-27
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.